W4849 - Toshiba AQ One Final Drawings - Integrity VeterinaryTHESE DRAWINGS ARE FORMATTED FOR 11 X 17 PAPER S-1 STRUCTURAL & ANCHORING INFORMATION M-1 ENVIRONMENTAL INFORMATION E-1 ELECTRICAL NOTES E-2 ELECTRICAL NOTES E-3 CT ELECTRICAL LEGEND E-4 CT ELECTRICAL PLAN E-5 ELECTRICAL DETAILS E-6 ELECTRICAL DETAILS E-7 CT SYSTEM CABLES SUBMISSION HISTORY 2-21-24 TEST FIT SKETCH 2-28-24 PRELIMS ISSUED 3-10-24 FINALS ISSUED MEDICAL EQUIPMENT SITE PLANNING INFORMATION PROVIDED BY *** ADDITIONAL REQUIRED REFERENCE MATERIAL, SEPARATE DOCUMENTS *** TOSHIBA SITE PLANNING GUIDE NO. 2A201 536EN*G THESE DRAWINGS ARE PROVIDED FOR COORDINATION PURPOSES AND ARE NOT TO BE USED FOR PERMITTING OR CONSTRUCTION 5335 AVION PARK DR. UNIT A HIGHLAND HEIGHTS, OH 44143 THESE DRAWINGS ARE ISSUED TO THE CUSTOMER FOR THE SOLE PURPOSE OF COORDINATING THE EQUIPMENT REQUIREMENTS WITH THE CUSTOMER'S PROFESSIONAL DESIGN TEAM AND ARE NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES. THIS IS YOUR FINAL SET OF DRAWINGS ISSUED FOR USE OF YOUR DESIGN TEAM, PLEASE SIGN WHERE INDICATED ON SHEET "APRV" AND RETURN TO YOUR PROJECT MANGER WITHIN 5 BUSINESS DAYS TO RELEASE YOUR EQUIPMENT FOR STAGING AND PREP. FINAL DRAWING INDEX COVER SHEET APRV CUSTOMER ACKNOWLEDGEMENT AND APPROVAL EDA EQUIPMENT DELIVERY AUTHORIZATION N-1 GENERAL NOTES N-2 GENERAL NOTES A-1 FLOOR PLAN A-2 CT RIGGING PATH CT-1 CT EQUIPMENT LEGEND CT-2 CT EQUIPMENT PLACEMENT CT-3 CT EQUIPMENT CLEARANCES CT-4 CT RADIATION SCATTER PLOT CT-5 CT EQUIPMENT DETAILS & NOTES CT-6 CT EQUIPMENT DETAILS & NOTES CT-7 CT RIGGING AND DELIVERY INFORMATION PROJECT IDENTIFICATION INFORMATION JOB NAME:INTEGRITY VETERINARY WO NUMBER:W4849 PATTERSON NO:N/A DRAFTING NO.2024-01 CUSTOMER INTEGRITY VETERINARY CENTER 518 PLEASANT ST NORTHAMPTON, MA 01060 LAUREN PAYNE LAUREN@INTEGRITYVETCENTER.COM ARCHITECT:T.B.D CONTRACTOR WAINSCOTT BUILDING MATTHEW WAINSCOTT 37 STAGE ROAD WILLIAMSBURG, MA 01096 413.559.0875 WAINSCOTTBUILDING@GMAIL.COM EQUIPMENT PRE-OWNED TOSHIBA AQUILION ONE CT STANDARD TABLE PROJECT MANAGER:BRIAN MALONEY OFFICE:(888) 505-0319 CELL:(216) 633-3854 JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED APRV CUSTOMER ACKNOWLEDGEMENT AND APPROVALNOTE TO CUSTOMER - FINAL EQUIPMENT PLANS: 1.THIS IS YOUR FINAL EQUIPMENT PLANNING PACKAGE, REVIEW THESE PLANS CAREFULLY, AND IF NECESSARY, MARK-UP WITH ANY DESIRED CHANGES WITHIN FIVE (5) BUSINESS DAYS. 2.RETURN A COPY OF THIS SHEET AND ANY SHEETS WITH MARK-UPS TO THE PROJECT MANAGER NOT LATER THAN FIVE (5) WORKING DAYS OF ISSUANCE TO CONFIRM YOUR RECEIPT AND ACKNOWLEDGEMENT OF THIS DRAWING PACKAGE. YOUR APPROVAL AUTHORIZES THE STAGING OF YOUR EQUIPMENT AND MOVES YOUR EQUIPMENT ORDER STATUS FROM SALES & PLANNING TO STAGING & PRE-DELIVERY. 3.REVISIONS REQUIRED TO THE FINAL EQUIPMENT PLANS AFTER THE INITIAL ISSUANCE IS SUBJECT TO ADDITIONAL DRAFTING FEES. PART 2 - FINAL EQUIPMENT PLAN APPROVAL AND ACKNOWLEDGEMENT THIS IS YOUR PHASE II, "FINAL" DRAWING PACKAGE ISSUED FOR USE OF YOUR DESIGN PROFESSIONALS TO GENERATE CONSTRUCTION DRAWINGS. THE CUSTOMER MUST SIGN BELOW AND RETURN A COPY OF THIS SHEET WITH SIGNATURE, ALONG WITH ANY COMMENTS, TO THE PRIZMED PROJECT MANAGER WITHIN FIVE (5) WORKING DAYS TO MOVE YOUR EQUIPMENT ORDER FROM THE SALES PHASE TO THE STAGING PHASE. IF A SIGNED APPROVAL IS NOT RECEIVED WITHIN THE SPECIFIED TIME THIS PACKAGE WILL BE CONSIDERED "APPROVED WITHOUT COMMENT" AND THE CUSTOMER AGREES AS SET FORTH BELOW. THE DRAWINGS WILL BE DEEMED TO BE ACKNOWLEDGED AND APPROVED BY THE CUSTOMER, AS IF SIGNED BY THE CUSTOMER, IF THIS SHEET IS NOT RETURNED AS STATED ABOVE. HAVING REVIEWED THE FINAL DRAWINGS WE HEREBY ACKNOWLEDGE, CONFIRM AND ACCEPT THAT: 1.THE EQUIPMENT PLANNING REQUIREMENTS INDICATED HEREIN ARE SUBJECT TO CHANGE UNTIL THE ACTUAL EQUIPMENT FOR THE PROJECT HAS BEEN PHYSICALLY RETAINED AND INSPECTED AND WE UNDERSTAND THAT THE EQUIPMENT SUPPLIER IS NOT RESPONSIBLE FOR ANY COSTS RESULTING FROM ADDED OR CHANGED DESIGN OR CONSTRUCTION REQUIREMENTS DUE TO THE FINAL EQUIPMENT CONFIGURATION. 2.THE CUSTOMER WILL RETAIN DESIGN PROFESSIONALS WHO SHALL BE SOLELY RESPONSIBLE FOR PREPARING CONSTRUCTION DRAWINGS AND INCORPORATING THE EQUIPMENT REQUIREMENTS INTO THE CONSTRUCTION DRAWINGS AND THESE EQUIPMENT PLANS WILL NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES. 5.ALL INFORMATION HEREIN IS BASED ON INFORMATION OBTAINED FROM 3RD PARTY SOURCES AND IS INTENDED TO CONVEY TO THE CUSTOMER'S PROFESSIONAL DESIGN TEAM THE EQUIPMENT REQUIREMENTS AND NECESSARY SITE PREPARATION. THE EQUIPMENT PROVIDER AND SITE PLANNER ARE NOT LICENSED DESIGN OR ENGINEERING PROFESSIONALS AND DO NOT PROVIDE ANY SUCH SERVICES. THE CUSTOMER AGREES TO INDEMNIFY AND HOLD HARMLESS THE EQUIPMENT PROVIDER, SITE PLANNER AND ANYONE FOR WHOM THEY ARE LEGALLY LIABLE, FROM AND AGAINST ANY AND ALL DAMAGE, LOSSES OR COST, (INCLUDING ARCHITECT FEE'S, ENGINEERING FEE'S, ATTORNEYS' FEES AND DEFENSE COSTS), CAUSED IN WHOLE OR IN PART BY ITS ACTS, ERRORS OR OMISSIONS IN PROVIDING THESE DRAWINGS AND THE INFORMATION CONTAINED HEREIN. 6.THE CUSTOMER SHALL AT HIS OWN EXPENSE WILL PERFORM ALL RECOMMENDED TESTING AND MONITORING AS IDENTIFIED HEREIN TO CONFIRM SITE SUITABILITY INCLUDING ANY NECESSARY SHIELDING DESIGNS, PERFORM ANY MITIGATION REQUIRED AS A RESULT OF TESTING AND MONITORING AND PROVIDE THE RESULTS OF ANY SUCH ACTIONS TO THE EQUIPMENT SUPPLIER. SHOULD THE CUSTOMER ELECT NOT TO PERFORM ANY RECOMMENDED TESTING, MONITORING, SHIELD DESIGNS OR MITIGATION ,THE CUSTOMER SHALL BE SOLELY RESPONSIBLE FOR ANY IMPACT ON THE INSTALLATION, SERVICE OR OPERATION OF THE EQUIPMENT, SAFETY OF PERSONS, EFFECTS TO OTHER EQUIPMENT AND OPERATIONS. 7.THE CUSTOMER SHALL FURNISH TO THE EQUIPMENT SUPPLIER A COPY OF THE COMPLETED CONSTRUCTION DRAWINGS AND ANY RELEVANT TEST REPORTS AND KEEP THE EQUIPMENT SUPPLIER CURRENT ON ANY CHANGES THAT MAY AFFECT THE EQUIPMENT OR IT'S INSTALLATION AND OPERATION. 8.THE CUSTOMER SHALL COORDINATE AS NECESSARY WITH THE EQUIPMENT SUPPLIER IN ALL ASPECTS TO ENSURE THE PROPER AND TIMELY PREPARATION OF THE SITE TO ACCEPT THE EQUIPMENT. 9.THE CUSTOMER SHALL SUPPLY TO THE EQUIPMENT SUPPLIER AS SOON AS AVAILABLE A COPY OF THE PROJECT SCHEDULE AND IDENTIFY THE EQUIPMENT DELIVERY DATE ON THE PROJECT SCHEDULE. 10.ALL EQUIPMENT BEING FURNISHED BY THE SUPPLIER IS "PRE-OWNED" UNLESS OTHERWISE SPECIFIED IN THE EXECUTED SALES AGREEMENT. 11.THE EQUIPMENT PROVIDER AND SITE PLANNER MAKE NO GUARANTEES AND SHALL NOT BE LIABLE FOR ANY IMPACT THE INSTALLATION OF THE EQUIPMENT MAY HAVE ON OTHER EQUIPMENT IN OR NEAR THE SPACE WHERE THIS EQUIPMENT WILL BE INSTALLED AND THE CUSTOMER ACCEPTS ALL LIABILITIES AND RESPONSIBILITIES THEREOF. 12.WE ARE TENTATIVELY REQUESTING THE EQUIPMENT BE INSTALLED AND READY FOR OPERATION ON: ____________________________________________ THE CUSTOMER HEREBY ACKNOWLEDGES, CONFIRMS, AND ACCEPTS THIS DRAWING PACKAGE AND ALL INFORMATION AND REQUIREMENTS CONTAINED HEREIN AND WILL ENSURE THE SAME FOR THE DESIGN TEAM AND CONTRACTORS: ____________________________________________________________________________________________________________________________________________________________________________ (SIGNATURE OF AUTHORIZED INDIVIDUAL)(DATE SIGNED)(PRINTED NAME & TITLE)(COMPANY) JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED EDA EQUIPMENT DELIVERY AUTHORIZATIONEQUIPMENT DELIVERY AUTHORIZATION IMPORTANT NOTE: THIS FORM MUST BE COMPLETED AND RETURNED TO YOUR EQUIPMENT PROVIDER'S PROJECT MANAGER NOT LESS THAN TEN (10) CALENDAR DAYS PRIOR TO THE SCHEDULED EQUIPMENT DELIVERY DATE, FAILURE TO PROVIDE THIS SHEET MAY DELAY THE SCHEDULED DELIVERY OF YOUR EQUIPMENT! THE CUSTOMER HEREBY CERTIFIES THAT THE SITE WILL BE READY FOR THE DELIVERY AND INSTALLATION OF THE EQUIPMENT IN ALL REGARDS AS SPECIFIED WITHIN THIS DRAWING PACKAGE ON _____________________________________________, AT WHICH TIME THE INSTALLATION OF THE EQUIPMENT WILL IMMEDIATELY WILL COMMENCE. IF THE SITE IS NOT PROPERLY PREPARED UPON DELIVERY OF THE EQUIPMENT AND THE INSTALLATION IS DELAYED AS A RESULT, ADDITIONAL CHARGES MAY BE INCURRED BY THE CUSTOMER, ALL CONSTRUCTION WORK MUST BE COMPLETE IN THE SPACE THE EQUIPMENT IS TO BE INSTALLED. SITE READINESS CHECKLIST 1.THE UNLOADING AREA AND THE DELIVERY PATH ARE CLEAR OF ALL OBSTRUCTIONS THAT WOULD HINDER THE DELIVERY OR INSTALLATION AND THE PATH HAS ADEQUATE SPACE FOR THE MOVEMENT OF THE EQUIPMENT. IF THERE ARE ANY RAMPS, STAIRS, ELEVATORS, DOCKS, GARAGES OR OTHER HINDRANCES ALONG THE DELIVERY PATH THE PROJECT MANAGER MUST BE NOTIFIED IN ADVANCE OF THE DELIVERY. 2.ALL SURFACES ARE TO BE COMPLETELY FINISHED, THIS INCLUDES ALL WALLS, CEILINGS, FLOORS, CABINETS, ETC., (WITH EXCEPTION OF THE DELIVERY PATH). IT IS HIGHLY RECOMMENDED THAT THE GENERAL CONTRACTOR USE HEAVY CRAFT PAPER TO PROTECT ALL FINISHED FLOORING AND COUNTERTOPS PRIOR TO THE DELIVERY OF THE EQUIPMENT. 3.ALL SYSTEMS MUST BE COMPLETE AND OPERATIONAL, INCLUDING ALL HVAC SYSTEMS FOR 72 HRS, ELECTRICAL SYSTEMS , EPO CIRCUIT TESTED, LIGHTING, LIFE SAFETY SYSTEMS , AND LIVE PHONE AND DATA SYSTEMS. 4.THE SPACE THE EQUIPMENT IS BEING INSTALLED MUST BE DUST-FREE AND CLIMATE STABILIZED FOR A MINIMUM OF 72 HOURS PRIOR TO THE ARRIVAL OF THE EQUIPMENT. 5.THERE MUST BE ADEQUATE SPACE FOR THE STAGING OF EQUIPMENT IN THE AREA AND ALL CONTRACTOR TOOLS AND MATERIALS REMOVED FROM THE AREA. 6.A TRASH RECEPTACLE OR DUMPSTER IS AVAILABLE FOR DISPOSAL OF SHIPPING AND PACKING MATERIALS. 7.CEILING TILES MUST BE REMOVED IN AREAS WHERE ACCESS WILL BE REQUIRED FOR SYSTEM CABLING, THE CUSTOMER OR CONTRACTOR MUST HAVE SOMEONE ON SITE ON DELIVERY DAY FOR THIS PURPOSE. THE CUSTOMER OR CONTRACTOR IS RESPONSIBLE FOR RE-INSTALLING SAME AFTER SYSTEM CABLES HAVE BEEN INSTALLED. 8.THE CUSTOMER'S ELECTRICIAN MUST BE ON SITE AND AVAILABLE THE AFTERNOON THE EQUIPMENT IS DELIVERED TO CONFIRM THE POWER SOURCE HAS THE PROPER RATING, GROUNDING, AND ROTATION AND TO ASSIST THE EQUIPMENT INSTALLERS WITH TERMINATION OF THE MAIN POWER ONTO THE EQUIPMENT. IF THERE ARE THERE ANY ITEMS THAT CAN NOT BE COMPLETED AS STATED ABOVE PRIOR TO EQUIPMENT DELIVERY, PLEASE EXPLAIN: ____________________________________________________________________________________________________ ______________________________________________________________________________________________________________________________________________________________________________________________________________________________ EXCEPT AS MAY BE STATED ABOVE, ALL ITEMS WILL BE COMPLETED AND THE SITE READY FOR DELIVERY AND INSTALLATION ON THE DATE INDICATED ABOVE. WE ACKNOWLEDGE THAT ONCE THE EQUIPMENT HAS ARRIVED AT THE SITE, ANY ISSUES OR DELAYS DUE TO NON-CONFORMING SITE CONDITIONS AND PREPARATION MAY IMPACT THE EQUIPMENT INSTALLATION AND MAY RESULT IN ADDITIONAL CHARGES TO THE CUSTOMER. ACKNOWLEDGED BY: __________________________________________ ____________________________ _____________________________________________________________________________________________________ (CUSTOMER SIGNATURE)(DATE)(PRINT NAME)(COMPANY) ACKNOWLEDGED BY: __________________________________________ ____________________________ _____________________________________________________________________________________________________ (CUSTOMER'S CONTRACTOR SIGNATURE)(DATE)(PRINT NAME)(COMPANY) NETWORK CONNECTIVITY FOR MEDICAL EQUIPMENT: THE FOLLOWING INFORMATION IS REQUIRED FOR COMMUNICATION WITH YOUR DATA SYSTEMS , PROVIDE THE FOLLOWING INFORMATION: MEDICAL EQUIPMENT IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________ PACS OR IMAGE DESTINATION IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________ WORKLIST IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________ EMR OR MEDICAL RECORDS IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________ OTHER: _________________________________________ IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________ JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED N-1 GENERAL NOTESDRAWING NOTES: 1.THESE DRAWINGS ARE FOR COORDINATION PURPOSES WITH THE CUSTOMER'S DESIGN TEAM AND ARE NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES. 2.THESE DRAWINGS ARE PROVIDED TO LOCATE THE EQUIPMENT WITHIN THE SPACE, SUGGEST THE PLACEMENT AND ROUTING OF RELATED SUPPORTING SYSTEMS, AND GUIDE THE CUSTOMER'S DESIGN TEAM TO INCORPORATE THE EQUIPMENT REQUIREMENTS INTO THE CONSTRUCTION DOCUMENTS. IN PREPARING THESE DRAWINGS, EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS TO THE ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. HOWEVER, BECAUSE THE INFORMATION CONTAINED HEREIN IS BEING PROVIDED IN ADVANCE OF THE RETENTION OF THE ACTUAL EQUIPMENT BEING PROVIDED, THESE DRAWINGS ARE SUBJECT TO CHANGE. THE INFORMATION CONTAINED HEREIN IS CORRECT TO THE BEST OF OUR KNOWLEDGE AT THE TIME THE INFORMATION WAS PREPARED. 3.THE EQUIPMENT PROVIDER AND SITE PLANNER ARE NOT RESPONSIBLE FOR ANY DAMAGES RESULTING FROM THE USE OF THESE DRAWINGS FOR ANY PURPOSES. THESE DRAWINGS ARE SCHEMATIC IN NATURE AND THE ROUTING AND LOCATIONS OF ANY CONDUITS, TROUGHS, WIREWAYS, BREAKERS, DISCONNECT SWITCHES, ETC. ARE SUGGESTIONS, ACTUAL LOCATIONS AND PLACEMENT ARE THE RESPONSIBILITY OF THE CUSTOMER'S DESIGN PROFESSIONALS. 4.ALL EQUIPMENT LAYOUTS ARE BASED ON THE RECOMMENDED SPACE NECESSARY FOR THE OPERATION AND SERVICEABILITY OF THE EQUIPMENT BEING PROPOSED AND ARE IN THE BEST INTEREST OF BOTH THE CUSTOMER AND EQUIPMENT SUPPLIER BASED ON ARCHITECTURAL DRAWINGS OR INFORMATION PROVIDED BY THE CUSTOMER. IF CHANGES ARE MADE TO THE LAYOUT OR DIMENSIONS THE SITE PLANNER IS TO BE NOTIFIED IMMEDIATELY AS CHANGES COULD RESULT IN CONFLICTS WITH THE EQUIPMENT CLEARANCES, PLACEMENT OF THE EQUIPMENT, AND POSSIBLY IMPACT THE CLINICAL USE OF THE EQUIPMENT. IF CHANGES ARE MADE TO THE FLOOR PLAN AFTER THE SUBMISSION OF THE INITIAL VERSION OF THESE PLANS ADDITIONAL DRAFTING CHARGES MAY BE INCURRED BY THE CUSTOMER. 5.ALL INFORMATION HEREIN IS BASED ON INFORMATION OBTAINED FROM 3RD PARTY SOURCES AND IS INTENDED TO CONVEY TO THE CUSTOMER'S PROFESSIONAL DESIGN TEAM THE EQUIPMENT REQUIREMENTS AND NECESSARY SITE PREPARATION. THE EQUIPMENT PROVIDER AND SITE PLANNER ARE NOT LICENSED DESIGN OR ENGINEERING PROFESSIONALS AND DO NOT PROVIDE ANY SUCH SERVICES. THE CUSTOMER AGREES TO INDEMNIFY AND HOLD HARMLESS THE EQUIPMENT PROVIDER, SITE PLANNER AND ANYONE FOR WHOM THEY ARE LEGALLY LIABLE, FROM AND AGAINST ANY AND ALL DAMAGES, LOSSES OR COSTS, (INCLUDING ARCHITECT FEES, ENGINEERING FEES, ATTORNEYS' FEES AND DEFENSE COSTS), CAUSED IN WHOLE OR IN PART BY ITS ACTS, ERRORS OR OMISSIONS IN PROVIDING THESE DRAWINGS AND THE INFORMATION CONTAINED HEREIN. . 6.THE CUSTOMER'S ARCHITECT AND GENERAL CONTRACTOR SHALL BE ULTIMATELY RESPONSIBLE FOR COMPLIANCE WITH ALL APPLICABLE CODES AND PROFESSIONAL DESIGN REQUIREMENTS. CONSTRUCTION DRAWINGS ARE TO BE SUBMITTED TO THE EQUIPMENT SUPPLIER FOR REVIEW PRIOR TO CONSTRUCTION. 7.ALL DIMENSIONS SHOWN ARE TAKEN FROM FINISHED SURFACES UNLESS SPECIFIED OTHERWISE. 8.THIS DRAWING PACKAGE DOES NOT PROVIDE SHIELDING DESIGNS, IF ANY SHIELDING IS REQUIRED IT SHALL BE THE RESPONSIBILITY OF THE CUSTOMER IN ALL REGARDS. 9.OWNERSHIP OF DOCUMENTS, DRAWINGS, SPECIFICATIONS AND OTHER INFORMATION, INCLUDING THOSE IN ELECTRONIC FORM, PREPARED BY THE EQUIPMENT PROVIDER AND THE EQUIPMENT PROVIDER'S CONSULTANTS ARE INSTRUMENTS OF SERVICE FOR USE SOLELY WITH RESPECT TO THIS PROJECT. THE EQUIPMENT PROVIDER AND THE EQUIPMENT PROVIDER'S CONSULTANTS SHALL BE DEEMED THE AUTHORS AND OWNERS OF THEIR RESPECTIVE INSTRUMENTS OF SERVICE AND SHALL RETAIN ALL COMMON LAW, STATUTORY AND OTHER RESERVED RIGHTS, INCLUDING COPYRIGHTS. SITE USAGE, DELIVERY AND INSTALLATION 1.THE MINIMUM STANDARD DELIVERY REQUIREMENTS FOR THE EQUIPMENT ARE INDICATED HEREIN, SPECIAL ATTENTION SHOULD BE PAID TO: 1.1.THE ENTIRE DELIVERY PATH MUST BEING LARGE ENOUGH TO ACCOMMODATE ALL SYSTEM COMPONENTS AND BE FREE AND CLEAR OF ALL OBSTRUCTIONS. 1.2.THE MINIMUM CEILING HEIGHTS AS NOTED IN THESE PLANS OR AS STATED IN THE O.E.M. REFERENCE DRAWINGS FOR THE SCAN AND EQUIPMENT ROOMS. 1.3.ALL SURFACES AND FLOORS ALONG THE ENTIRE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING THE WEIGHT OF ALL SYSTEM COMPONENTS. 2.FOR STANDARD RIGGING AND DELIVERY, ANY OPENINGS, DOORWAYS, OR TURNS IN THE DELIVERY PATH MUST BE ABLE TO ACCOMMODATE THE GANTRY AND TABLE AS AN ASSEMBLY. SEE DETAILS HEREIN. 3.ADEQUATE SPACE FOR STAGING THE EQUIPMENT DURING INSTALLATION. THE STORAGE SPACE MUST BE INSIDE THE BUILDING IN CLOSE PROXIMITY TO THE WORK AREA, BE CLIMATE CONTROLLED AND ABLE TO BE SECURED AFTER HOURS. 4.THE CUSTOMER SHALL PROVIDE AT NO COST TO THE EQUIPMENT SUPPLIER A CONTAINER, (DUMPSTER, ROLL-OFF, ETC.), FOR THE DISPOSAL OF ALL PACKAGING AND SHIPPING DEBRIS, GENERALLY CONSISTING OF PALLETS, CRATING, BOXES AND PROTECTIVE PLASTIC WRAP. 5.THE ANTICIPATED TIME FRAME FOR INSTALLATION OF THE MEDICAL EQUIPMENT IS TYPICALLY: THREE (3) WORKING DAYS FOR CT SYSTEMS 6.THE INSTALLATION CREW RESERVES THE RIGHT TO HAVE ACCESS TO THE SITE AND PERFORM WORK OUTSIDE OF NORMAL WORKING HOURS, (7:30AM - 5:00PM), WITH ADVANCED NOTICE TO THE CUSTOMER. 7.IF NECESSARY, THE CUSTOMER IS RESPONSIBLE FOR THE REMOVAL AND REPLACEMENT OF CEILING TILES TO FACILITATE THE CABLING AND/OR INSTALLATION OF THE EQUIPMENT. CUSTOMER'S RESPONSIBILITIES 1.THE CUSTOMER IS SOLELY RESPONSIBLE FOR RETAINING A PROFESSIONAL DESIGN TEAM TO PREPARE PERMIT AND CONSTRUCTION DRAWINGS, ENSURING COMPLIANCE WITH ALL EQUIPMENT REQUIREMENTS, CODES, REGULATORY REQUIREMENTS AND SECURING ALL APPROVALS, PERMITS AND LICENSES. 2.ALL WORK EXCEPT THE INSTALLATION OF THE EQUIPMENT IS THE RESPONSIBILITY OF THE CUSTOMER WITHOUT EXCEPTION AND THE EQUIPMENT PROVIDER ASSUMES NO RESPONSIBILITY FOR ANY CONSTRUCTION COSTS, WHETHER OR NOT RELATED TO THE INSTALLATION OF THE EQUIPMENT OR THE USE OF THESE DRAWINGS AND THE INFORMATION CONTAINED HEREIN. 3.THE GENERAL CONTRACTOR SHALL BE RESPONSIBLE FOR THE REINSTALLATION OF CEILING TILES, TOUCH-UP AND ANY COSMETIC WORK AFTER THE INSTALLATION OF THE EQUIPMENT. 4.THE CUSTOMER SHALL BE RESPONSIBLE FOR ALL SAFETY SIGNAGE REQUIRED BY ACR, ISO, JCAHO, OSHA, ETC., INCLUDING SAFETY AND WARNING SIGNAGE FOR THE CRYOGEN EXHAUST SYSTEM. 5.THE CUSTOMER SHALL PERFORM AN INSPECTION AT LEAST FIVE (5) WORKING DAYS PRIOR TO DELIVERY OF THE EQUIPMENT AND VERIFY IN WRITING TO THE EQUIPMENT PROVIDER USING SHEET "PDCL" OF THESE DRAWINGS THAT THE SITE IS READY FOR DELIVERY AND INSTALLATION AS SCHEDULED. UPON DELIVERY OF THE EQUIPMENT, SHOULD THE SITE NOT BE READY AND CAUSE DELAY IN THE INSTALLATION, THE CUSTOMER SHALL REIMBURSE THE EQUIPMENT PROVIDER FOR ANY ASSOCIATED COSTS INCLUDING COSTS FOR ADDITIONAL STORAGE, WAGES, TRAVEL, LODGING, OR RE-MOBILIZATION COSTS. 6.EQUIPMENT WARRANTIES, EXPRESSED OR IMPLIED, ON THE PART OF EQUIPMENT SUPPLIER SHALL BE CONTINGENT UPON STRICT COMPLIANCE WITH ALL RECOMMENDATIONS AND REQUIREMENTS CONTAINED IN THESE DRAWINGS. 7.THE EQUIPMENT SUPPLIER SHALL BE RESPONSIBLE FOR THE EQUIPMENT INSTALLATION AND CALIBRATION WHICH INCLUDES INSTALLING SYSTEM CABLES AND TERMINATING CONTRACTOR PROVIDED CABLES TO THE EQUIPMENT. IN THE EVENT THAT SPECIFIC TRADE RULES OR LICENSE REQUIREMENTS PROHIBIT THIS, THE CUSTOMER SHALL INITIATE THE SERVICES OF APPROVED OTHER CONTRACTORS AND PAY FOR SELECTED, APPROVED PARTIES TO PERFORM THIS WORK WITH JOB SUPERVISION TO BE PROVIDED BY THE EQUIPMENT PROVIDER. CALIBRATION WHEN ACCOMPLISHED OUTSIDE OF NORMAL INSTALLATION SEQUENCES DUE TO CONTRACTOR OR TRADE RULE ACTIONS OR REQUIREMENTS SHALL BE SUPPORTED BY, CHARGED TO, AND ACCEPTED BY THE CUSTOMER AS AN ADDITIONAL INSTALLATION EXPENSE. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED N-2 GENERAL NOTESRECOMMENDED TESTING AND ENVIRONMENTAL EVALUATION THE CUSTOMER IS RESPONSIBLE FOR HAVING A COMPLETE ENVIRONMENTAL SURVEY PERFORMED OF THE PROPOSED SITE TO ENSURE ALL ENVIRONMENTAL CONDITIONS WILL MEET THE REQUIREMENTS OF THE ORIGINAL MANUFACTURER OF THE PROPOSED EQUIPMENT. RECOMMENDED ENVIRONMENTAL TESTING INCLUDES THE FOLLOWING: 1.FOR ALL MEDICAL IMAGING EQUIPMENT THE PRIMARY ELECTRICAL SERVICE MUST BE THOROUGHLY EVALUATED TO INCLUDE A LOAD STUDY ON THE ELECTRICAL SERVICE INTENDED FOR THE IMAGING EQUIPMENT AND MONITORING TO ENSURE THE POWER QUALITY MEETS THE SPECIFICATIONS INCLUDING TRANSIENT VOLTAGES AND GROUNDING. THE CUSTOMER IS RESPONSIBLE FOR VERIFICATION THAT PRIOR TO THE SYSTEM BEING INSTALLED THE REQUIREMENTS ARE MET. IN THE EVENT THERE ARE ANY ENVIRONMENTAL OR POWER DEFICIENCIES AT THE SITE, THE FACILITY WILL BE REQUIRED TO CORRECT SUCH DEFICIENCIES PRIOR TO ENERGIZATION OF THE EQUIPMENT. PLEASE BE ADVISED THAT ANY QUALITY ISSUES OR DAMAGE TO THE IMAGING SYSTEM AS A RESULT OF ENVIRONMENTAL OR POWER ISSUES WILL FALL OUTSIDE ANY WARRANTY OR SERVICE AGREEMENT COVERAGE. A RECOMMENDED RESOURCE FOR THE MONITORING IS: RX MONITORING SERVICES, (888) 329-2321 http://www.rxms.com/ 2.ALL MEDICAL IMAGING EQUIPMENT IS VERY SENSITIVE TO VIBRATION. THE EQUIPMENT MUST BE LOCATED AS FAR AS POSSIBLE FROM VIBRATION SOURCES SUCH AS ADJACENT ROADWAYS, PARKING LOTS, SUBWAYS, TRAIN TRACKS, HALLWAYS, ELEVATORS, WEIGHT MACHINES, ETC. IF ANY ASPECT OF THE SITE IS IN QUESTION, THE CUSTOMER IS REQUIRED TO PERFORM A VIBRATION STUDY. CEILING AND STRUCTURAL SUPPORTS 1.ALL STRUCTURAL SUPPORTS ARE THE RESPONSIBILITY OF THE CUSTOMER AND SHALL BE DESIGNED AND INSTALLED BY THE CUSTOMER'S DESIGN TEAM AND CONTRACTORS. THIS INCLUDES LOAD BEARING SLABS, ELEVATED SLABS, CEILING SUPPORT SYSTEMS, SHORING, BRACING, ETC., WITHOUT EXCEPTION. ALL SUPPORT SYSTEMS MUST BE COMPLETELY INSTALLED AND READY PRIOR TO THE DELIVERY OF EQUIPMENT. FLOOR LEVELNESS 1.THE FLOOR MUST BE LEVEL TO WITHIN 1/8" IN 10'-0" WHERE THE EQUIPMENT IS PLACED, THE EQUIPMENT CAN NOT BE SHIMMED. 2.THE CONTRACTOR MUST VERIFY IN WRITING NOT LESS THAN 5 DAYS BEFORE DELIVERY THAT THE FLOOR MEETS THIS REQUIREMENT. EQUIPMENT FURNISHED AND INSTALLED BY CUSTOMER'S CONTRACTORS 1.SPECIAL CARE AND CONSIDERATION MUST BE GIVEN TO ITEMS THAT ARE TO BE FURNISHED AND INSTALLED BY THE CUSTOMER'S CONTRACTORS THAT ARE NOT INCLUDED WITH THE MEDICAL EQUIPMENT. EVERY ATTEMPT HAS BEEN MADE TO IDENTIFY THESE ITEMS WHICH MAY INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: ·MAIN DISCONNECT CABINETS AND BREAKERS ·CONDUITS, TROUGHS, BREAKERS, WIRING ·LIGHTING, CONTROLS AND REQUIRED WARNING LIGHTS OR CONTROLLERS ·RADIATION SHIELDING SYSTEMS AND DESIGNS LAYOUT PRIORITY ITEMS 1.EQUIPMENT ISO-CENTER IS THE CRITICAL POINT FOR ALL LAYOUT WORK IN THE SCAN ROOM. ANY DEVIATION MUST BE REPORTED TO THE SITE PLANNER IMMEDIATELY. THE GENERAL CONTRACTOR SHALL MAINTAIN MARKS FOR ISO-CENTER THROUGHOUT THE PROJECT. NETWORKING AND NETWORK INFORMATION 1)THE CUSTOMER MUST PROVIDE TO THE EQUIPMENT PROJECT MANAGER NOT LESS THAN 1 WEEK IN ADVANCE OF THE DELIVERY OF THE EQUIPMENT ALL NETWORKING AND IT INFORMATION AS MAY BE REQUIRED TO CONFIGURE THE MEDICAL EQUIPMENT TO BE COMPATIBLE WITH THE CUSTOMER'S SYSTEMS SUCH AS THE EMR, PACS, WORKLIST, REMOTE SEND DESTINATIONS, ETC. 2)CUSTOMER ACKNOWLEDGES THAT DURING THE INSTALLATION OF THE EQUIPMENT THE INSTALLATION TEAM WILL PROGRAM AND TEST ALL NETWORKING. SHOULD NETWORK TESTS FAIL DUE TO EXTERNAL ISSUES THE CUSTOMER WILL RESOLVE ALL SUCH ISSUES WHILE THE INSTALLATION TEAM IS ON SITE. IF FOR ANY EXTERNAL REASON AN ENGINEER MUST RETURN TO THE SITE TO RESOLVE EXTERNAL NETWORKING ISSUES THE CUSTOMER WILL BE BILLED ACCORDINGLY. REQUIRED INFORMATION FROM CUSTOMER FOR CT PROJECTS, THE FOLLOWING INFORMATION IS TO BE PROVIDED BY THE CUSTOMER TO THE EQUIPMENT SUPPLIER: 1.A COPY OF THE PROJECT SCHEDULE INDICATING AN EQUIPMENT DELIVERY DATE. 2.A COPY OF THE COMPLETED CONSTRUCTION DRAWINGS. 3.A COPY OF ANY SHIELDING DRAWINGS AND CALCULATIONS. 4.COPIES OF ANY CORRESPONDENCE AND RFI'S THAT PERTAIN TO THE IMAGING EQUIPMENT THROUGHOUT THE PROJECT. 5.THE COMPLETED EQUIPMENT DELIVERY AUTHORIZATION FORM FROM THIS SET OF DRAWINGS. SITE READINESS AND EQUIPMENT DELIVERY THE CUSTOMER'S CONTRACTOR MUST SUBMIT AS SOON AS IT IS AVAILABLE, A PROJECT SCHEDULE TO THE EQUIPMENT PROVIDER. THE PROJECT SCHEDULE MUST SPECIFICALLY IDENTIFY THE DESIRED DELIVERY DATE FOR THE EQUIPMENT. IF THE DATE CHANGES, THE EQUIPMENT PROVIDER MUST BE NOTIFIED SO ADJUSTMENTS CAN BE MADE TO THE DELIVERY AND INSTALLATION. SCHEDULE. A LETTER SHALL BE PROVIDED TO THE CUSTOMER AT AN APPROPRIATE TIME TO CONFIRM THE SITE READINESS AND DELIVERY DATE. IN GENERAL, SITE READINESS TO ACCEPT DELIVERY OF THE EQUIPMENT INCLUDES, BUT IS NOT LIMITED TO, THE FOLLOWING: 1.ALL CONSTRUCTION WORK MUST BE COMPLETE IN THE SPACE THE EQUIPMENT IS TO BE INSTALLED. 2.ALL SURFACES ARE TO BE COMPLETELY FINISHED, THIS INCLUDES ALL WALLS, CEILINGS, FLOORS, CABINETS, ETC., (WITH EXCEPTION OF THE DELIVERY PATH). THE CUSTOMER'S CONTRACTORS ARE RESPONSIBLE FOR PROTECTING FINISHED SURFACES AS REQUIRED, (CRAFT PAPER ON FLOORS AND COUNTERTOPS ARE SUGGESTED). 3.ALL SYSTEMS MUST BE COMPLETE AND OPERATIONAL, INCLUDING ALL HVAC SYSTEMS, ELECTRICAL SYSTEMS , LIGHTING, LIFE SAFETY, PHONE AND DATA NETWORKS. 4.THE ENVIRONMENT MUST BE DUST-FREE AND CLIMATE STABILIZED FOR A MINIMUM OF 72 HOURS PRIOR TO THE ARRIVAL OF THE EQUIPMENT. 5.THERE MUST BE ADEQUATE SPACE FOR THE STAGING OF EQUIPMENT IN THE AREA AND ALL CONTRACTOR TOOLS AND MATERIALS REMOVED FROM THE AREA. 6.A TRASH RECEPTACLE OR DUMPSTER IS AVAILABLE FOR DISPOSAL OF SHIPPING AND PACKING MATERIALS. 7.WHERE NECESSARY, CEILING TILES MUST BE REMOVED TO FACILITATE THE INSTALLATION OF SYSTEM CABLING AS DIRECTED BY THE INSTALLATION TEAM, IT IS THE CUSTOMER'S CONTRACTORS RESPONSIBILITY TO RE-INSTALL SAME AFTER SYSTEM CABLES HAVE BEEN INSTALLED. 8.THE UNLOADING AREA AND DELIVERY PATH ARE CLEAR OF ALL OBSTRUCTIONS THAT WOULD HINDER THE DELIVERY OR INSTALLATION AND THE PATH HAS ADEQUATE SPACE FOR THE MOVEMENT OF THE EQUIPMENT. IF THERE ARE ANY RAMPS, STAIRS, ELEVATORS, DOCKS, GARAGES OR OTHER HINDRANCES ALONG THE DELIVERY PATH THE PROJECT MANAGER MUST BE NOTIFIED IN ADVANCE OF THE DELIVERY. 9.THE CUSTOMER'S ELECTRICIAN MUST BE ON SITE AND AVAILABLE THE AFTERNOON EQUIPMENT IS DELIVERED TO CONFIRM THE POWER SOURCE HAS THE PROPER RATING, GROUNDING AND ROTATION AND TO ASSIST THE EQUIPMENT INSTALLERS WITH TERMINATION OF THE MAIN POWER ONTO THE EQUIPMENT. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED A-1 FLOOR PLANNorth 19'-101 2" 14'-7" 5" 7' 11'-51 2" 16'-2" 51 2" 6'-61 2" 3'-11 2" 3'-81 2" 4' 4' EXISTING ELECT. TRANS. EXIST ELECT PANEL4' VERIFY 8' VERIFY REV1 DIMENSIONS REVISED DUE TO GANTRY WIDTH REV1 VERIFY EXACT LOCATION OF TRANSFORMER AND PANEL TO REMAIN. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED A-2 CT RIGGING PATHNorth RIGGING AND DELIVERY NOTES 1.REFER TO SECTION 7 OF THE SITE PLANNING GUIDE FOR CLEARANCE REQUIREMENTS. 2.THE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING 6,500 LBS SPREAD OVER FOUR 6" DIAMETER RUBBER WHEELS. 3.MINIMUM OPENINGS 3'-7" CLEAR WIDTH X 81" CLEAR HEIGHT 4.DEPICTED IS THE PRESUMED RIGGING PATH FOR THE EQUIPMENT. THE CUSTOMER MUST ENSURE THE ENTIRE EQUIPMENT DELIVERY PATH IS CLEAR AND ALL HALLWAYS, CORRIDORS, OPENINGS AND POINTS THE EQUIPMENT WILL HAVE TO BE TURNED ARE LARGE ENOUGH TO ACCOMMODATE THE EQUIPMENT, SEE SHEET A-3 FOR TURNING REQUIREMENTS. INADEQUATE TURN AREA FOR GANTRY AND TABLE TO COME IN THROUGH THIS DOOR, IT WOULD BE IMPOSSIBLE TO TURN THE GANTRY INTO THE SCAN ROOM WITH THE CONTROL ROOM WALL BUILT. 3'-3" MIN GANTRY 8' 3'-7" 7'-9" MIN GANTRY 3'-3" MIN GANTRY 8' 3'-7" 7'-9" MIN GANTRY JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED A-3 RIGGING INFORMATIONNorth 5'-6" 5'-6" CLEAR OPENING 8' 6' CLEAR OPENING 5'-6" 5' CLEAR OPENING 8' GANTRY DELIVERY OPENING REQUIREMENTS TABLE DELIVERY OPENING REQUIREMENTS 5'-6" CLEAR OPENING 5'-6" 8' 6' CLEAR OPENING 5'-6" 5' CLEAR OPENING 8' GANTRY DELIVERY OPENING REQUIREMENTS TABLE DELIVERY OPENING REQUIREMENTS 3'-3" MIN GANTRY 8' 3'-7" 7'-9" MIN GANTRY 2'-2" 8' CBTB-021A TABLE JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-1 CT EQUIPMENT LEGENDCEILING HEIGHT ·RECOMMENDED CEILING HEIGHT: 9'-0" ·MINIMUM CEILING HEIGHT: 8'-2 1/2" VIBRATION REQUIREMENTS FOR AQUILION GANTRY .98 M/S² (0.1 G) OR LESS IMPORTANT NOTES 1.THESE EQUIPMENT PLANS AND THE SYSTEM COMPONENTS MUST BE VERIFIED WHEN THE ACTUAL SYSTEM FOR THIS PROJECT HAS BEEN RETAINED, INSPECTED AND STORED. THERE ARE SYSTEM OPTIONS AND VARIANCES THAT CAN IMPACT SITE PREP, SPACE REQUIREMENTS, PERFORMANCE, AND SAFETY. 2.IT IS THE RESPONSIBILITY OF THE CUSTOMER TO RETAIN A REGISTERED PHYSICIST TO PREPARE A SHIELDING PRESCRIPTION AND DETERMINE THE EXTENT AND AMOUNTS OF RADIATION SHIELDING REQUIRED. THIS PACKAGE SHOULD BE PROVIDED TO THE MEDICAL PHYSICIST ALONG WITH WORKFLOW, VOLUME AND ADJACENT OCCUPANCY DATA. CEILING HEIGHT ·RECOMMENDED CEILING HEIGHT: 9'-0" ·MINIMUM CEILING HEIGHT: 8'-2 1/2" SYSTEM ELECTRICAL REQUIREMENTS FOR AQUILION WITH "PCDU" SUPPLY CONFIGURATION:3 PHASE, 3-WIRE DELTA OR WYE CONNECTED STANDARD TRANSFORMER CAPACITY:150 KVA MAXIMUM SYSTEM DEMAND (IMAGING):125 KVA SUPPLY VOLTAGE:480V, 60HZ, 150 AMP *TOSHIBA EQUIPMENT BTU / HOUR: (INCLUDING GANTRY & COUCH ONLY) EXAM ROOM STAND-BY:14,672 BTU/HR EXAM ROOM SCANNING 2 PATIENTS:20,712 BTU/HR EXAM ROOM SCANNING 3 PATIENTS:21,991 BTU/HR EXAM ROOM SCANNING 4 PATIENTS:23,271 BTU/HR EXAM ROOM SCANNING 5 PATIENTS:24,550 BTU/HR EXAM ROOM SCANNING MAXIMUM:25,830 BTU/HR NOTE: FUTURE GROWTH OF FACILITY MUST BE CONSIDERED WHEN FORECASTING PATIENT NUMBERS FOR A/C REQUIREMENTS. VIBRATION REQUIREMENTS FOR AQUILION GANTRY .98 M/S² (0.1 G) OR LESS IMPORTANT NOTES 1.THESE EQUIPMENT PLANS AND THE SYSTEM COMPONENTS MUST BE VERIFIED WHEN THE ACTUAL SYSTEM FOR THIS PROJECT HAS BEEN RETAINED, INSPECTED AND STORED. THERE ARE SYSTEM OPTIONS AND VARIANCES THAT CAN IMPACT SITE PREP, SPACE REQUIREMENTS, PERFORMANCE, AND SAFETY. 2.IT IS THE RESPONSIBILITY OF THE CUSTOMER TO RETAIN A REGISTERED PHYSICIST TO PREPARE A SHIELDING PRESCRIPTION AND DETERMINE THE EXTENT AND AMOUNTS OF RADIATION SHIELDING REQUIRED. THIS PACKAGE SHOULD BE PROVIDED TO THE MEDICAL PHYSICIST ALONG WITH WORKFLOW, VOLUME AND ADJACENT OCCUPANCY DATA. CASEWORK AND ACCESSORIES 1)ALL CASEWORK IS TO BE DESIGNED, DETAILED, FURNISHED, AND INSTALLED BY THE CUSTOMER AND/OR CONTRACTOR.. 2)ALL FURNITURE, (OPERATOR CHAIRS), SIGNAGE, FURNISHINGS, AND ACCESSORIES ARE PROVIDED BY THE CUSTOMER. 3)THE COUNTER TOP AT THE OPERATORS STATION MUST BE INSTALLED PRIOR TO DELIVERY OF THE EQUIPMENT. ELECTRICAL 1)ANY SOUND OR INTERCOM SYSTEMS ARE TO BE FURNISHED AND INSTALLED BY THE CUSTOMER. 2)EMERGENCY POWER OFF (EPO) BUTTONS ARE REQUIRED IN THE CONTROL ROOM AND SCAN ROOM, LOCATIONS TO BE DETERMINED BY THE CUSTOMER'S ELECTRICAL ENGINEER. CUSTOMER'S ELECTRICAL ENGINEER IS RESPONSIBLE FOR DESIGN OF EPO CIRCUIT. 3)DOOR SAFETY SWITCHES AND IN USE SIGNAGE IS REQUIRED ON ALL DOORS ACCESSING THE EXAMINATION ROOM. 4)IT IS HIGHLY SUGGESTED THE CUSTOMER PERFORM A POWER SURVEY PRIOR TO DESIGN AND CONSTRUCTION TO ENSURE THE INCOMING POWER WILL MEET THE EQUIPMENT SPECIFICATIONS AND REQUIREMENTS. DAMAGE TO EQUIPMENT DUE TO IMPROPER POWER IS NOT COVERED BY ANY SERVICE AGREEMENT OR WARRANTY. AQUILION ONE MINIMUM ROOM SIZE REQUIREMENTS SCAN ROOM CONTROL ROOM 13'-2"W X 19'-10"L 6'-0"W X 8'-0"L TOSHIBA AQUILION ONE CT EQUIPMENT LEGEND FURNISHED AS PART OF CT SYSTEM, INSTALLED BY PROVIDER TAG ELEC ITEM DESCRIPTION BTU/HR WEIGHT SYMB (LBS) 1 GANT AQUILION CT GANTRY SEE NOTE 6,063 2 PCH CT PATIENT COUCH SEE NOTE 1,941 3 CPU CENTRAL PROCESSING UNIT (HOST PC)5,123 276 4 MON MONITOR FOR CPU 192 18 5 SPK1 SPEAKER - OPERATOR ---4 6 SPK2 SPEAKER - SCAN ROOM ---4 7 SKBD SCAN CONSOLE KEYBOARD ---6 8 DKBD SCAN CONSOLE KEYBOARD ---6 9 30" DEEP COUNTER BY CUSTOMER 10 CHAIR BY CUSTOMER 11 INV INVERSION WORKSTATION 12 REC1 RECONSTRUCTION UNIT 13,020 816 13 REC2 RECONSTRUCTION UNIT 13,020 816 14 PR PHANTOM RACK 15 PDU POWER DISTRIBUTION UNIT 4,777 1,654 ITEMS FURNISHED AND INSTALLED BY CUSTOMER'S CONTRACTORS CB MAIN CIRCUIT BREAKER FOR CT SYSTEM ONLY. SHUNT TRIP TYPE BREAKER WITH REMOTE POWER OFF STATIONS (EPO). SHOWN IN SUGGESTED LOCATION, ACTUAL LOCATION BY CUSTOMER'S ELECTRICAL ENGINEER. EPO EMERGENCY POWER OFF STATION. SHOWN IN SUGGESTED LOCATION. ACTUAL LOCATION AND DESIGN OF EPO CIRCUITS DETERMINED BY CUSTOMER'S ELECTRICAL ENGINEER. NOTE: FOR HEAT OUTPUT OF SYSTEM IN SCAN ROOM, SEE CHART THIS SHEET FOR BTU/HOUR ENVIRONMENTAL REQUIREMENTS 1.MAINTAIN AT ALL TIMES 24/7/365 AMBIENT TEMPERATURE BETWEEN 68F TO 74F, HUMIDITY RANGE OF 40% TO 70% NON CONDENSING. STATIC AMBIENT TEMPERATURE IS TO BE PROVIDED AND MAINTAINED AS SPECIFIED. ALL CALCULATIONS ARE TO UTILIZE TOSHIBA PROVIDED HEAT OUTPUT SPECIFICATIONS OF EQUIPMENT, (SEE CHART ABOVE). 2.IN GENERAL THE SCAN ROOM MUST BE PROVIDED WITH AN INDEPENDENT AIR CONDITIONING SYSTEM.IF THE ROOM IS MAINTAINED WITHIN THE PERMISSIBLE TEMPERATURE RANGE, GRADUAL TEMPERATURE SHIFTS MAY AFFECT SYSTEM PERFORMANCE, THEREFORE THE ROOM MUST BE KEPT UNDER CONSTANT CONTROL, WITHIN ±3%. 3.A MINIMUM OF 10 AIR CHANGES PER HOUR IS SUGGESTED. CONSULT LOCAL CODE. 4.EQUIPMENT IN ENCLOSED SPACES SUCH AS EQUIPMENT ROOMS, CLOSETS AND COMPUTER ROOMS MUST BE PROVIDED WITH ADEQUATE VENTILATION. 5.A/C SUPPLY OUTLET TO BE PROVIDED BY CUSTOMER AT FLOOR LEVEL OR AS CLOSE AS POSSIBLE TO CONTROL ROOM DESK. 6.DUE TO HEAT GENERATED BY THE "CPU" AND "REC" UNITS, ADDITIONAL VENTILATION IN THE CONTROL AREA IS REQUIRED. CUSTOMER OR CONTRACTOR PROVIDED FANS MAY BE NECESSARY BELOW THE DESKTOP FOR TECHNICIAN COMFORT. THE "CPU/REC" UNITS SHOULD NOT BE BE ENCLOSED IN CASEWORK. 7.AIR SUPPLY DUCTS SHOULD NOT BE PLACED DIRECTLY OVER EXAMINATION TABLES OR GANTRY FOR PATIENT COMFORT CONSIDERATIONS. 8.THE AIRFLOW THROUGH TOSHIBA EQUIPMENT CABINETS IS FROM BOTTOM TO TOP. 9.WHERE POSSIBLE A/C SUPPLY OUTLETS SHOULD BE LOCATED AT FLOOR LEVEL. NO A/C OUTLET SHOULD BE WITHIN THE EXCLUSION ZONE SHOWN BELOW AND AT NO TIME SHOULD THE CT SYSTEM BE EXPOSED TO DIRECT AIRFLOW. NETWORK REQUIREMENTS DUE TO THE LARGE DATA SETS GENERATED BY THE HIGH SPEED SCANS AN OPTIMIZED NETWORK IS REQUIRED TO MAINTAIN FAST IMAGE TRANSFER SPEEDS. THE TWO REQUIREMENTS ARE: 1)HIGH SPEED NETWORK, GIGABIT ETHERNET BETWEEN THE AQUILION AND ITS NETWORK PEERS. 2)AN UNCLUTTERED NETWORK ENVIRONMENT. THE AQUILION'S IMMEDIATE NETWORK, INCLUDING CONNECTION TO THE VITAL WORKSTATION AND TEH MCKESSON THIN SLICE CACHE SHOULD BE ON AN ISOLATED NETWORK SEGMENT FREE OF BROADCASTS AND EXTRANEOUS SIGNALS. THIS CAN BE ACHIEVED BY VLAN, FIREWALL, OR SIMILAR TECHNOLOGIES.THE CUSTOMER WILL NEED TO MAKE ITS OWN INDEPENDENT DETERMINATION ABOUT THE ADEQUACY OF SUCH ITEMS. THE EQUIPMENT PROVIDED WILL NOT BE RESPONSIBLE FOR ISSUES ARISING FROM THE NETWORK ENVIRONMENT. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-2 CT EQUIPMENT PLACEMENTNorth 3 12 15 13 5 7 8 9 10 4 12 CBTB-021A TABLE 2 1 MAX TRAVEL W/HEAD REST EXTENSION 6 11 11 14 6'-7" 7'-4" EXISTING ELECT. TRANS. EXIST ELECT PANEL 12 JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-3 CT EQUIPMENT CLEARANCESNorth MAX TRAVEL W/HEAD REST EXTENSION 3'-6 1 16" 13'-13 8" 11 5 8"101 8" 2'-111 4"10'-11 7 8" 5'-3 3 8" 5'-9 3 16" RECOMMENDED DISTANCE TO WALL 4'-51 8" 4'-9 3 16" 2'-2" 1'-61 8" 8' GANTRY TILTGANTRY TILT MIN. ROOM SIZE MINIMUM SERVICE CLEARANCE N.E.C. CLEARANCE EXISTING ELECT. TRANS. EXIST ELECT PANEL JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-4 CT RADIATION SCATTER PLOTSHIELDING DESIGN NOTES FOR CUSTOMER 1.THE CUSTOMER IS RESPONSIBLE FOR RETAINING THE SERVICES OF A QUALIFIED MEDICAL PHYSICIST TO PREPARE A RADIATION SHIELDING DESIGN AND TO HIRE CONTRACTORS TO FURNISH AND INSTALL THE SHIELDING PRESCRIBED BY THE MEDICAL PHYSICIST. 2.SHIELDING IS REQUIRED TO PROTECT THOSE PERSONS WORKING ADJACENT TO X-RAY ROOMS. LEAD WHICH OFFERS SPECIAL SHIELDING PROPERTIES AT MEDICAL X-RAY ENERGIES IS THE PREFERRED AND MOST USED MATERIAL. 3.ALL DIAGNOSTIC SHIELDING WALLS REQUIRE THE SHIELDING TO EXTEND FROM THE FLOOR TO AT LEAST 7 FEET ABOVE THE FLOOR. THE DOORS AND WINDOWS IN THEM ARE REQUIRED TO BE OF EQUIVALENT SHIELDING IN ALL CASES. THAT IS, WHEN A WALL REQUIRES 1/16 INCH LEAD SHIELDING, A DOOR OR WINDOW INSTALLED IN THAT WALL IS ALSO REQUIRED TO HAVE SHIELDING EQUIVALENT TO 1/16 INCH OF LEAD. PARTICULAR ATTENTION MUST BE GIVEN TO THE CONSTRUCTION OF ALL JOINTS AND WHERE PENETRATIONS OCCUR. 4.IN THIS DESIGN, THERE IS NO OCCUPANCY ABOVE THE CEILING OR BELOW THE FLOOR; THEREFORE, THE USUAL CONSTRUCTION MATERIALS PROVIDE ADEQUATE SHIELDING SO THAT NO ADDITIONAL LEAD IS REQUIRED IN THE FLOORS AND CEILINGS. 5.THE CONTROL BOOTH IS DESIGNED TO MAXIMIZE THE PROTECTION OF THE OPERATOR WHILE PROVIDING THE OPERATOR THE MOST ADVANTAGEOUS VIEW OF THE ROOM. IN ALL CASES, THE CONTROL SWITCH MUST BE POSITIONED AT LEAST 40 INCHES FROM THE OUTSIDE EDGE OF THE CONTROL BOOTH. SCATTER RADIATION NOTES UNITS: MICROGRAY (PER 100 mAs) THE AMOUNT OF SCATTERED RADIATION DURING SCANNING IS SHOWN IN THE SCATTER DIAGRAM. REFER TO THIS DATA WHEN SCANNING TO MINIMIZE X-RAY EXPOSURE. NOTE THAT THE AMOUNT OF SCATTERED RADIATION DIFFERS DEPENDING ON THE PATIENT; THEREFORE, THE DATA SHOWN SHOULD BE USED ONLY AS A GUIDE. 120 kVp/100 mA/1.0 sec/M FOV/0,5 mm x 80/320 mm PMMA PHANTOM THE ABOVE CALCULATION IS BASED ON A MULTI-SLICE EXPOSURE. THE CUSTOMER IS RESPONSIBLE FOR HAVING SHIELDING CALCULATIONS PREPARED BY A LICENSED RADIATION PHYSICIST. SHIELDING DESIGN NOTES FOR CUSTOMER 1.THE CUSTOMER IS RESPONSIBLE FOR RETAINING THE SERVICES OF A QUALIFIED MEDICAL PHYSICIST TO PREPARE A RADIATION SHIELDING DESIGN AND TO HIRE CONTRACTORS TO FURNISH AND INSTALL THE SHIELDING PRESCRIBED BY THE MEDICAL PHYSICIST. 2.SHIELDING IS REQUIRED TO PROTECT THOSE PERSONS WORKING ADJACENT TO X-RAY ROOMS. LEAD WHICH OFFERS SPECIAL SHIELDING PROPERTIES AT MEDICAL X-RAY ENERGIES IS THE PREFERRED AND MOST USED MATERIAL. 3.ALL DIAGNOSTIC SHIELDING WALLS REQUIRE THE SHIELDING TO EXTEND FROM THE FLOOR TO AT LEAST 7 FEET ABOVE THE FLOOR. THE DOORS AND WINDOWS IN THEM ARE REQUIRED TO BE OF EQUIVALENT SHIELDING IN ALL CASES. THAT IS, WHEN A WALL REQUIRES 1/16 INCH LEAD SHIELDING, A DOOR OR WINDOW INSTALLED IN THAT WALL IS ALSO REQUIRED TO HAVE SHIELDING EQUIVALENT TO 1/16 INCH OF LEAD. PARTICULAR ATTENTION MUST BE GIVEN TO THE CONSTRUCTION OF ALL JOINTS AND WHERE PENETRATIONS OCCUR. 4.IN THIS DESIGN, THERE IS NO OCCUPANCY ABOVE THE CEILING OR BELOW THE FLOOR; THEREFORE, THE USUAL CONSTRUCTION MATERIALS PROVIDE ADEQUATE SHIELDING SO THAT NO ADDITIONAL LEAD IS REQUIRED IN THE FLOORS AND CEILINGS. 5.THE CONTROL BOOTH IS DESIGNED TO MAXIMIZE THE PROTECTION OF THE OPERATOR WHILE PROVIDING THE OPERATOR THE MOST ADVANTAGEOUS VIEW OF THE ROOM. IN ALL CASES, THE CONTROL SWITCH MUST BE POSITIONED AT LEAST 40 INCHES FROM THE OUTSIDE EDGE OF THE CONTROL BOOTH. RADIATION SCATTER PLOT NOTE TO CUSTOMER YOUR PHYSICIST WILL NEED ALL THE INFORMATION ON THIS SHEET TO PROVIDE YOU WITH THE "PRESCRIPTION" FOR YOUR RADIATION (LEAD) SHIELDING ALONG WITH YOUR SCAN VOLUMES, SCAN TYPES, AND OCCUPANCY OF THE SPACE AND ADJACENT SPACES. RADIATION SHIELDING IS THE CUSTOMER'S RESPONSIBILITY. RADIATION SCATTER PLOT 150 100 50 20 10 5 100 50 10 5 10 5 150 JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-5 SYSTEM DETAILS7'-4" 8'-0" 6'-9 1 2" 3'-5 5 16" 7'-4" 3'-85 8" 1'-101 2" 18'-10 11 16" 5'-9 3 16" RECOMMENDED DISTANCE TO WALL5'-4" MAX TRAVEL W/HEAD REST EXTENSION 3'-6 1 16" 13'-13 8" 11 5 8"101 8" 2'-111 4"10'-11 7 8" 5'-3 3 8" 4'-51 8" 4'-9 3 16" 2'-2"8' GANTRY TILTGANTRY TILT 9'-5 5 8" 6'-91 2" JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-6 SYSTEM DETAILS2'-8 1 16" 1'-511 16" 2'-31 2" NOT USED 2'-8 1 16" 1'-511 16" 2'-31 2" SIDEFRONT 5 1 2" 7 7 8' 5 5 16" 5 1 2" NOT USED 2'-8" 4'-3 3 4" 3'-5" 2'-8" 14 7 8" 4" 13 1 2" CPU - CENTRAL PROCESSING UNIT HEAT BTU = 5,123 WEIGHT = 276 LBS PLAN 2'-2 1 2" 5'-0 5 8" REC - RECONSTRUCTION UNIT #1 HEAT BTU = 26,040 - 816 LBS #2 HEAT BTU INCL #1 - 728 LBS PCDU - POWER DISTRIBUTION UNIT HEAT BTU = 4,777 WEIGHT = 1,654 LBS 3'-01 2" 2'-21 2" 15 11 16" 15 7 8" 8" SIDEFRONT MON - VIEWING MONITOR HEAT BTU = 192 WEIGHT = 18 LBS SIDEFRONT SIDEFRONT SPK - SPEAKERS HEAT BTU = 0 WEIGHT = 5 LBS SIDEFRONT IV - INVERSION HEAT BTU = T.B.D. WEIGHT = 17 LBS 13 15 16" PLAN PLAN PLAN PLAN SIDEFRONT PLAN JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED CT-7 RIGGING INFORMATIONRIGGING AND DELIVERY NOTES 1.REFER TO SECTION 7 OF THE SITE PLANNING GUIDE FOR CLEARANCE REQUIREMENTS. 2.THE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING 6,500 LBS SPREAD OVER FOUR 6" DIAMETER RUBBER WHEELS. 3.MINIMUM OPENINGS 3'-7" CLEAR WIDTH X 81" CLEAR HEIGHT 4.DEPICTED IS THE PRESUMED RIGGING PATH FOR THE EQUIPMENT. THE CUSTOMER MUST ENSURE THE ENTIRE EQUIPMENT DELIVERY PATH IS CLEAR AND ALL HALLWAYS, CORRIDORS, OPENINGS AND POINTS THE EQUIPMENT WILL HAVE TO BE TURNED ARE LARGE ENOUGH TO ACCOMMODATE THE EQUIPMENT, SEE SHEET A-3 FOR TURNING REQUIREMENTS. INADEQUATE TURN AREA FOR GANTRY AND TABLE TO COME IN THROUGH THIS DOOR, IT WOULD BE IMPOSSIBLE TO TURN THE GANTRY INTO THE SCAN ROOM WITH THE CONTROL ROOM WALL BUILT. 3'-3" MIN GANTRY 8' 3'-7" 7'-9" MIN GANTRY 3'-3" MIN GANTRY 8' 3'-7" 7'-9" MIN GANTRY EXISTING ELECT. TRANS. EXIST ELECT PANEL 6'-8" 7'-0" 8'-0" AQUILION DELIVERY REQUIREMENTS PRIOR TO GANTRY DELIVERY, CHECK THE DELIVERY ENTRANCE, CORRIDOR WIDTH, ELEVATOR CAPACITY, ETC. TO ENSURE EASE OF DELIVERY. COMPENSATION FOR ADDITIONAL CLEARANCES DUE TO THE RIGGING EQUIPMENT AND HARDWARE TO BE USED MAY BE REQUIRED, CONSULT WITH THE RIGGING CONTRACTOR WELL IN ADVANCE OF DELIVERY TO ENSURE THERE IS ADEQUATE CLEARANCES AND MANEUVERABILITY SPACE. SITE USAGE, DELIVERY AND INSTALLATION 1)THE MINIMUM STANDARD DELIVERY REQUIREMENTS FOR THE EQUIPMENT ARE AS FOLLOWS: * THE ENTIRE DELIVERY PATH MUST BE LARGE ENOUGH TO ACCOMMODATE ALL SYSTEM COMPONENTS AND BE FREE AND CLEAR OF ALL OBSTRUCTIONS. * MINIMUM CEILING HEIGHTS ARE TO BE AS NOTED IN THESE PLANS OR AS STATED IN THE O.E.M. REFERENCE DRAWINGS. * FLOORS AND ENTIRE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING THE WEIGHT OF ALL SYSTEM COMPONENTS. 2)FOR STANDARD RIGGING AND DELIVERY ANY OPENINGS, DOORWAYS, OR TURNS IN THE DELIVERY PATH MUST BE ABLE TO ACCOMMODATE THE GANTRY AND TABLE AS AN ASSEMBLY. SEE DETAILS HEREIN. 3)ADEQUATE STORAGE/STAGING SPACE IS REQUIRED FOR THE DELIVERY OF THE EQUIPMENT. THE STORAGE SPACE MUST BE INSIDE THE BUILDING WITHIN CLOSE PROXIMITY TO THE WORK AREA, CLIMATE CONTROLLED, AND BE ABLE TO BE SECURED AFTER HOURS. 4)THE CUSTOMER SHALL PROVIDE AT NO COST TO THE EQUIPMENT SUPPLIER A CONTAINER, (DUMPSTER, ROLL-OFF, ETC.), FOR THE DISPOSAL OF ALL PACKAGING AND SHIPPING DEBRIS, GENERALLY CONSISTING OF PALLETS, CRATING, BOXES AND PROTECTIVE PLASTIC WRAP. 5)THE ANTICIPATED TIME FRAME FOR INSTALLATION OF THE MEDICAL EQUIPMENT IS TYPICALLY THREE (3) WORKING DAYS FOR CT SYSTEMS 6)THE INSTALLATION CREW RESERVES THE RIGHT TO HAVE ACCESS TO THE SITE AND PERFORM WORK OUTSIDE OF NORMAL WORKING HOURS, (7:30AM - 5:00PM), WITH ADVANCED NOTICE TO THE OWNER. 7)THE PROTECTION OF FINISHED SURFACES IS THE RESPONSIBILITY OF THE CUSTOMER AND THE CUSTOMER SHOULD EXPECT MINOR TOUCHUP OF FINISHED SURFACES AND PERFORM A FINAL CLEANING AFTER THE COMPLETION OF EQUIPMENT INSTALLATION, (AT NO COST TO EQUIPMENT PROVIDER). 9'-6" 7'-0" 3'-7" 8'-0" 5'-0" AQUILION DELIVERY REQUIREMENTS PRIOR TO GANTRY DELIVERY, CHECK THE DELIVERY ENTRANCE, CORRIDOR WIDTH, ELEVATOR CAPACITY, ETC. TO ENSURE EASE OF DELIVERY. COMPENSATION FOR ADDITIONAL CLEARANCES DUE TO THE RIGGING EQUIPMENT AND HARDWARE TO BE USED MAY BE REQUIRED, CONSULT WITH THE RIGGING CONTRACTOR WELL IN ADVANCE OF DELIVERY TO ENSURE THERE IS ADEQUATE CLEARANCES AND MANEUVERABILITY SPACE. MINIMUM DELIVERY REQUIREMENTS ALL FLOORS ALONG THE DELIVERY PATHWAY MUST BE CAPABLE OF SUPPORTING THE EQUIPMENT AND RIGGING HARDWARE. ALL DOORWAYS ALONG THE DELIVERY PATH MUST BE AT LEAST 6'-9" CLEAR HEIGHT AND 43" IN CLEAR WIDTH. THE MINIMUM CLEAR WIDTH OF ANY CORRIDOR LEADING INTO THE CT SCAN ROOM MUST BE AT LEAST 6'-7" WIDE. SITE USAGE, DELIVERY AND INSTALLATION 1)THE MINIMUM STANDARD DELIVERY REQUIREMENTS FOR THE EQUIPMENT ARE AS FOLLOWS: * THE ENTIRE DELIVERY PATH MUST BE LARGE ENOUGH TO ACCOMMODATE ALL SYSTEM COMPONENTS AND BE FREE AND CLEAR OF ALL OBSTRUCTIONS. * MINIMUM CEILING HEIGHTS ARE TO BE AS NOTED IN THESE PLANS OR AS STATED IN THE O.E.M. REFERENCE DRAWINGS. * FLOORS AND ENTIRE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING THE WEIGHT OF ALL SYSTEM COMPONENTS. 2)FOR STANDARD RIGGING AND DELIVERY ANY OPENINGS, DOORWAYS, OR TURNS IN THE DELIVERY PATH MUST BE ABLE TO ACCOMMODATE THE GANTRY AND TABLE AS AN ASSEMBLY. SEE DETAILS HEREIN. 3)ADEQUATE STORAGE/STAGING SPACE IS REQUIRED FOR THE DELIVERY OF THE EQUIPMENT. THE STORAGE SPACE MUST BE INSIDE THE BUILDING WITHIN CLOSE PROXIMITY TO THE WORK AREA, CLIMATE CONTROLLED, AND BE ABLE TO BE SECURED AFTER HOURS. 4)THE CUSTOMER SHALL PROVIDE AT NO COST TO THE EQUIPMENT SUPPLIER A CONTAINER, (DUMPSTER, ROLL-OFF, ETC.), FOR THE DISPOSAL OF ALL PACKAGING AND SHIPPING DEBRIS, GENERALLY CONSISTING OF PALLETS, CRATING, BOXES AND PROTECTIVE PLASTIC WRAP. 5)THE ANTICIPATED TIME FRAME FOR INSTALLATION OF THE MEDICAL EQUIPMENT IS TYPICALLY THREE (3) WORKING DAYS FOR CT SYSTEMS 6)THE INSTALLATION CREW RESERVES THE RIGHT TO HAVE ACCESS TO THE SITE AND PERFORM WORK OUTSIDE OF NORMAL WORKING HOURS, (7:30AM - 5:00PM), WITH ADVANCED NOTICE TO THE OWNER. 7)THE PROTECTION OF FINISHED SURFACES IS THE RESPONSIBILITY OF THE CUSTOMER AND THE CUSTOMER SHOULD EXPECT MINOR TOUCHUP OF FINISHED SURFACES AND PERFORM A FINAL CLEANING AFTER THE COMPLETION OF EQUIPMENT INSTALLATION, (AT NO COST TO EQUIPMENT PROVIDER). JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED S-1 STRUCTURAL AND ANCHORINGC.G. WT. = 1,343 LBS. (INCLUDES 350 LB. PATIENT) 30" 6" MIN. C.G. WT. = 3,859 LBS. FRONT ELEVATION 6" (MIN.) STRUCTURAL NOTES 1.THE ANCHORING OF THE EQUIPMENT SHALL BE PERFORMED BY THE EQUIPMENT PROVIDER IN ACCORDANCE WITH THE DESIGN AND SPECIFICATIONS SET FORTH BY THE CUSTOMER'S STRUCTURAL ENGINEER OF RECORD. 2.DUE TO THE DYNAMIC NATURE OF THE LOAD, BOTH HORIZONTAL AND VERTICAL ACCELERATIONS SHOULD BE INCLUDED IN THE DESIGN CALCULATIONS FOR THE SUPPORT STRUCTURE AS WELL AS ANCHORING AND THRU-BOLTING FOR THE EQUIPMENT. 3.IF THE FLOOR THE EQUIPMENT IS TO BE PLACED ON IS ANYTHING OTHER THAN A MINIMUM 6" THICK CONCRETE SLAB THE EQUIPMENT PROVIDER MUST BE IMMEDIATELY NOTIFIED. C.G. WT. = 1,343 LBS. (INCLUDES 350 LB. PATIENT) 30" 6" MIN. M12 ANCHORS C.G. WT. = 3,859 LBS. SEE DETAIL BELOW FRONT ELEVATION 6" (MIN.)M16 ANCHORS FLOOR SURFACE REQUIREMENTS 1.THE GANTRY AND PATIENT TABLE MUST BE INSTALLED ON THE SAME PLANE. IT IS THE CUSTOMER'S RESPONSIBILITY TO MEET FLOOR LEVELNESS SPECIFICATIONS AS OUTLINED IN THIS DETAIL AND IN THE APPENDIX OF THE SITE PLANNING MANUAL, (3 16" MAX OVER ENTIRE AREA). 2.TO ACHIEVE THE FLOOR LEVELNESS SPECIFICATIONS, AN EPOXY PAD IS TO BE INSTALLED BY THE EQUIPMENT PROVIDER. THE EPOXY PAD MUST BE INSTALLED DIRECTLY OVER THE CONCRETE FLOOR, (NOT OVER ANY FINISH FLOORING). SEE APPENDIX SECTION OF THE SITE PLANNING MANUAL FOR FULL DETAILS AND SPECIFICATIONS. 3.THE CURING OF THE PAD IS TEMPERATURE SENSITIVE. IN ORDER TO PROPERLY CURE, THE SLAB MUST BE AT A MINIMUM OF 60°F (48 HR. CURE TIME) OR 77°F (24 HR. CURE TIME). THIS SHOULD BE COORDINATED WITH THE EQUIPMENT PROJECT MANAGER AND MUST BE COMPLETED PRIOR TO DELIVERY OF THE SYSTEM. 4.THE EQUIPMENT MAY BE SET ON TOP OF FINISHED FLOORING ONLY IF THE FINISH FLOORING PRODUCT WILL NOT COMPRESS AND CAN SUPPORT THE EQUIPMENT WITHOUT COMPRESSING OVER TIME. BOLTING REQUIREMENTS 1.A MINIMUM OF A 6" THICK CONCRETE SLAB IS REQUIRED FOR ANCHORING. 2.THE WEIGHT CAPACITY OF THE FLOOR MUST BE EVALUATED BY THE CUSTOMER'S STRUCTURAL ENGINEER. THE EQUIPMENT PROVIDER RESERVES THE RING, IN THE INTEREST OF SAFETY, TO DELAY INSTALLATION UNTIL UNTIL STRUCTURAL DESIGN DRAWINGS STAMPED BY THE CUSTOMERS STRUCTURAL ENGINEER ARE REVIEWED BY THE EQUIPMENT PROVIDER. 3.THE PATIENT TABLE AND GANTRY MUST ALWAYS BE BOLTED TO THE FLOOR THROUGH THE DESIGNATED ATTACHMENT POINTS IN THE GANTRY AND TABLE PEDESTAL. 4.ANCHORS TO MUST PROVIDE A MINIMUM EXTRACTION FORCE OF 621 POUNDS PER ATTACHMENT POINT IS REQUIRED. 5.IT IS THE RESPONSIBILITY OF THE CUSTOMER'S STRUCTURAL ENGINEER TO DETERMINE THE ANCHORING DEPTH AND CONCRETE STRENGTH NEEDED TO INSTALL THE GANTRY AND TABLE BASE. CENTER OF GRAVITY ISO-CENTER 3'-4 7 16" 1'-7 11 16"1'-11 7 16" 4 1 2" 10 5 8"3'-5 13 16" STANDARD ANCHOR LOCATIONS SHOWN SHOWN, IF FACILITY REQUIRES SEISMIC ANCHORING, CONTACT YOUR PROJECT PLANNER. COORDINATE ANCHOR HOLE SIZES AND LOCATION WITH INSTALLERS INSTALLERS. GANTRY AND TABLE BASE DETAIL REFER TO APPENDIX, PAGE 77 OF SITE PLANNING MANUAL FOR FULL REQUIREMENT, DIMENSIONS, SPECIFICATIONS AND EPOXY LEVELING PAD INSTRUCTIONS 3'-10 7 16" 13" 2 3 4" 3 1 8" 3'-1 3 8" 7 1 16" 215 16" 7 7 8"14 15 16" 6'-0 7 16" 17 11 16" 7 7 8" 20 7 16" 12 3 16" THE AREA BEHIND THE GANTRY MUST BE KEPT FLAT AND LEVEL EPOXY IN THIS AREA MUST BE REMOVED TO USE MAIN DETECTOR REPLACEMENT JIG. FLOOR TO BE FLAT SHADED AREA INDICATES LEVELING EPOXY, MUST BE INSTALLED OVER CLEAN-DRY CONCRETE. 13 4" 1 3 16" 10'-4 13 16" 3'-4 151 16 " 3'-5 3 8" 3'-10 7 16" JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED M-1 ENVIRONMENTAL SPECIFICATIONSCT SUITE HVAC REQUIREMENTS 1)AMBIENT TEMPERATURE SHOULD BE 68F TO 74F, HUMIDITY RANGE OF 40% TO 70% NON CONDENSING. STATIC AMBIENT TEMPERATURE IS TO BE PROVIDED AND MAINTAINED AS SPECIFIED. ALL CALCULATIONS ARE TO UTILIZE PROVIDED HEAT OUTPUT SPECIFICATIONS OF EQUIPMENT. 2)A MINIMUM OF 10 AIR CHANGES PER HOUR IS SUGGESTED. CONSULT LOCAL CODE. 3)EQUIPMENT IN ENCLOSED SPACES SUCH AS EQUIPMENT ROOMS, CLOSETS AND COMPUTER ROOMS MUST BE PROVIDED WITH ADEQUATE VENTILATION. 4)A/C SUPPLY OUTLET TO BE PROVIDED BY CUSTOMER AT FLOOR LEVEL OR AS CLOSE AS POSSIBLE TO CONTROL ROOM DESK. 5)DUE TO HEAT GENERATED BY THE "CPU" AND "REC" UNITS, ADDITIONAL VENTILATION IN THE CONTROL AREA IS REQUIRED. CUSTOMER OR CONTRACTOR PROVIDED FANS MAY BE NECESSARY BELOW THE DESKTOP FOR TECHNICIAN COMFORT. THE "CPU/REC" UNITS SHOULD NOT BE BE ENCLOSED IN CASEWORK. 6)AIR SUPPLY DUCTS SHOULD NOT BE PLACED DIRECTLY OVER EXAMINATION TABLES OR GANTRY FOR PATIENT COMFORT CONSIDERATIONS. 7)THE AIRFLOW THROUGH TOSHIBA EQUIPMENT CABINETS IS FROM BOTTOM TO TOP. 8)WHERE POSSIBLE A/C SUPPLY OUTLETS SHOULD BE LOCATED AT FLOOR LEVEL. NO A/C OUTLET SHOULD BE WITHIN THE EXCLUSION ZONE SHOWN BELOW AND AT NO TIME SHOULD THE CT SYSTEM BE EXPOSED TO DIRECT AIRFLOW. TIMER ON A/C ON 1.5 HRS SCAN START TIME TOLERANCE: ±3°F RECOMMENDED TEMPERATURE CONDITIONS IN CT SCAN ROOM NOTES: 1)BTU / HR SHOWN FOR GANTRY AND PATIENT COUCH ONLY. 2)FUTURE GROWTH OF FACILITY MUST BE CONSIDERED WHEN FORECASTING PATIENT NUMBERS FOR A/C REQUIREMENTS. 3'-3" 2'-9" 4'-6"4'-6" A/C SUPPLY AIR EXCLUSION ZONE *TOSHIBA EQUIPMENT BTU / HOUR: (INCLUDING GANTRY & COUCH ONLY) EXAM ROOM STAND-BY:14,672 BTU/HR EXAM ROOM SCANNING 2 PATIENTS:20,712 BTU/HR EXAM ROOM SCANNING 3 PATIENTS:21,991 BTU/HR EXAM ROOM SCANNING 4 PATIENTS:23,271 BTU/HR EXAM ROOM SCANNING 5 PATIENTS:24,550 BTU/HR EXAM ROOM SCANNING MAXIMUM:25,830 BTU/HR NOTE: FUTURE GROWTH OF FACILITY MUST BE CONSIDERED WHEN FORECASTING PATIENT NUMBERS FOR A/C REQUIREMENTS. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-1 CT ELECTRICAL NOTESELECTRICAL NOTES 1.THESE SITE PLANS ARE INTENDED TO DEPICT THE CONCEPT OF THE ELECTRICAL REQUIREMENTS FOR THE EQUIPMENT. THE DESIGN OF ALL ELECTRICAL ELEMENTS MUST BE SPECIFIED BY A LICENSED ELECTRICAL ENGINEER IN ACCORDANCE WITH THE EQUIPMENT SPECIFICATIONS AND ALL APPLICABLE CODES. 2.IN ACCORDANCE WITH N.E.C. ARTICLE 517-72(B), THE EQUIPMENT CIRCUIT BREAKER(S) DISCONNECT MUST BE LOCATED SO THAT THEY SHALL BE OPERABLE FROM A LOCATION READILY ACCESSIBLE FROM THE CONTROL AREA. IF THIS IS IMPOSSIBLE OR IMPRACTICAL, THE USE OF A SHUNT TRIP TYPE BREAKER WILL BE NECESSARY TO SATISFY THIS REQUIREMENT. THE EMERGENCY POWER OFF BUTTON FOR THE SHUNT TRIP SHOULD BE LOCATED IN THE CONTROL AREA. 3.THE CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL CIRCUIT BREAKERS, CONDUITS, WIRES, JUNCTION BOXES, DUCTS, TROUGHS, ETC. 4.THESE PLANS DO NOT SPECIFY ELECTRICAL REQUIREMENTS FOR EQUIPMENT NOT PROVIDED WITH THE CT SYSTEM. 5.INTERCONNECTING CABLES FOR THE CT EQUIPMENT ARE FURNISHED WITH THE CT SYSTEM AND INSTALLED BY THE CT INSTALLERS. 6.EXCEPT FOR THEIR USE IN POWER LINE CONNECTIONS TO EQUIPMENT CABINETS, FLEXIBLE CONDUIT SHALL NOT BE USED IN THIS INSTALLATION. ONLY FACTORY CONDUIT ELBOWS SHALL BE USED. DUCT WORK SHALL BE PROVIDED WITH SWEEP ELBOWS. 7.ALL JUNCTION BOXES AND DUCTS THAT PENETRATE THE FLOOR SHALL BE WATERPROOF TYPE AND PROVIDED WITH GASKETED WATERPROOF COVERS. ALL FLOOR JUNCTION BOXES AND DUCT COVERS SHALL BE CAPABLE OF SUPPORTING 200 LBS OF CONCENTRATED LOAD. 8.GROMMETED OPENINGS ARE SHOWN FOR REFERENCE PURPOSES ONLY. VERIFY SIZE AND LOCATION WITH INSTALLATION REPRESENTATIVE. ALL GROMMETED OPENINGS SHALL HAVE NO SHARP EDGES. 9.ALL CHASE & GROMMETED OPENINGS SHALL HAVE PLASTIC BUSHINGS. 10.ALL DUCT WORK SHALL HAVE A MINIMUM OF THREE COMPARTMENTS WITH TRANSITIONS FROM HORIZONTAL TO VERTICAL WALL DUCT OR JUNCTION BOXES. LOCAL CODES MAY REQUIRE THE USE OF CROSS-OVER TUNNELS OR OTHER SUCH DEVICES TO MAINTAIN CABLE SEPARATION. 11.ALL DUCTS, TROUGHS AND CONDUITS MUST BE ELECTRICALLY BONDED AS A GROUNDING PATH IN ACCORDANCE WITH N.E.C. ARTICLE 517-13(B). 12.THE ELECTRICAL CONTRACTOR SHALL SUPPLY AND INSTALL NYLON MEASURING PULL STRING OR EQUIVALENT IN ALL CONDUITS AND CLOSED DUCT WORK. 13.CONDUIT RUNS SHOWN ARE FOR REFERENCE ONLY. ALL CONDUIT RUNS MUST TAKE THE SHORTEST MOST DIRECT ROUTE POSSIBLE. CONDUIT RUNS MAY HAVE A MAXIMUM OF (3) 90 DEGREE BENDS. 14.110 VAC GROUNDED OUTLETS SHALL BE PROVIDED ON WALLS NEAR EQUIPMENT FOR USE DURING EQUIPMENT SERVICE. 15.THE ELECTRICAL CONTRACTOR MUST SUPPLY AND INSTALL ALL INCOMING POWER CABLES FROM CIRCUIT BREAKER(S) TO THE EQUIPMENT. 16.CABLE TYPE MUST BE MTW MULTI-STRAND COPPER - NO ALUMINUM IS PERMITTED. CABLE SIZE MUST BE IN ACCORDANCE WITH POWER QUALITY REQUIREMENTS SPECIFICATIONS. 17.CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL NECESSARY HARDWARE TO ENCLOSE INCOMING POWER CABLES IN WATER TIGHT CONDUIT FROM CIRCUIT BREAKER(S) TO EQUIPMENT CABINET(S). 18.ANY CHANGES IN THE LOCATION OR TYPE OF CONDUIT, DUCT WORK, JUNCTION BOXES, ETC. MUST BE SUBMITTED IN WRITING TO THE INSTALLATION PROJECT MANAGER FOR APPROVAL. 19.ALL DUCT WORK MAKING A 90 DEGREE ANGLE MUST BE CHAMFERED FOR CABLE ACCESS. 20.JUNCTION BOX SIZES MAY BE INCREASED AS NEEDED (WITH EXCEPTION OF THE "PCDU" / "VUC" / "PCDU" / OR "PDU" JUNCTION BOX IF APPLICABLE.). 21.FIBER OPTIC CABLES FROM "REC" TO "GANT" & "REC" TO "CPU" REQUIRE A MINIMUM RADIUS OF 4 1/2". DUCT WORK DESIGN MUST ACCOMMODATE THIS REQUIREMENT. 22.EMERGENCY POWER OFF (EPO) BUTTONS REQUIRED IN CONTROL AREA, EXAMINATION ROOM AND EQUIPMENT AREA.. 23.DOOR (SAFETY) SWITCH REQUIRED ON ALL DOORS ACCESSING THE EXAMINATION ROOM IN ACCORDANCE WITH LOCAL CODES. ELECTRICAL NOTES 1.THESE SITE PLANS ARE INTENDED TO DEPICT THE CONCEPT OF THE ELECTRICAL REQUIREMENTS FOR THE EQUIPMENT. THE DESIGN OF ALL ELECTRICAL ELEMENTS MUST BE SPECIFIED BY A LICENSED ELECTRICAL ENGINEER IN ACCORDANCE WITH THE EQUIPMENT SPECIFICATIONS AND ALL APPLICABLE CODES. 2.IN ACCORDANCE WITH N.E.C. ARTICLE 517-72(B), THE EQUIPMENT CIRCUIT BREAKER(S) DISCONNECT MUST BE LOCATED SO THAT THEY SHALL BE OPERABLE FROM A LOCATION READILY ACCESSIBLE FROM THE CONTROL AREA. IF THIS IS IMPOSSIBLE OR IMPRACTICAL, THE USE OF A SHUNT TRIP TYPE BREAKER WILL BE NECESSARY TO SATISFY THIS REQUIREMENT. THE EMERGENCY POWER OFF BUTTON FOR THE SHUNT TRIP SHOULD BE LOCATED IN THE CONTROL AREA. 3.THE CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL CIRCUIT BREAKERS, CONDUITS, WIRES, JUNCTION BOXES, DUCTS, TROUGHS, ETC. 4.THESE PLANS DO NOT SPECIFY ELECTRICAL REQUIREMENTS FOR EQUIPMENT NOT PROVIDED WITH THE CT SYSTEM. 5.INTERCONNECTING CABLES FOR THE CT EQUIPMENT ARE FURNISHED WITH THE CT SYSTEM AND INSTALLED BY THE CT INSTALLERS. 6.EXCEPT FOR THEIR USE IN POWER LINE CONNECTIONS TO EQUIPMENT CABINETS, FLEXIBLE CONDUIT SHALL NOT BE USED IN THIS INSTALLATION. ONLY FACTORY CONDUIT ELBOWS SHALL BE USED. DUCT WORK SHALL BE PROVIDED WITH SWEEP ELBOWS. 7.ALL JUNCTION BOXES AND DUCTS THAT PENETRATE THE FLOOR SHALL BE WATERPROOF TYPE AND PROVIDED WITH GASKETED WATERPROOF COVERS. ALL FLOOR JUNCTION BOXES AND DUCT COVERS SHALL BE CAPABLE OF SUPPORTING 200 LBS OF CONCENTRATED LOAD. 8.GROMMETED OPENINGS ARE SHOWN FOR REFERENCE PURPOSES ONLY. VERIFY SIZE AND LOCATION WITH INSTALLATION REPRESENTATIVE. ALL GROMMETED OPENINGS SHALL HAVE NO SHARP EDGES. 9.ALL CHASE & GROMMETED OPENINGS SHALL HAVE PLASTIC BUSHINGS. 10.ALL DUCT WORK SHALL HAVE A MINIMUM OF THREE COMPARTMENTS WITH TRANSITIONS FROM HORIZONTAL TO VERTICAL WALL DUCT OR JUNCTION BOXES. LOCAL CODES MAY REQUIRE THE USE OF CROSS-OVER TUNNELS OR OTHER SUCH DEVICES TO MAINTAIN CABLE SEPARATION. 11.ALL DUCTS, TROUGHS AND CONDUITS MUST BE ELECTRICALLY BONDED AS A GROUNDING PATH IN ACCORDANCE WITH N.E.C. ARTICLE 517-13(B). 12.THE ELECTRICAL CONTRACTOR SHALL SUPPLY AND INSTALL NYLON MEASURING PULL STRING OR EQUIVALENT IN ALL CONDUITS AND CLOSED DUCT WORK. 13.CONDUIT RUNS SHOWN ARE FOR REFERENCE ONLY. ALL CONDUIT RUNS MUST TAKE THE SHORTEST MOST DIRECT ROUTE POSSIBLE. CONDUIT RUNS MAY HAVE A MAXIMUM OF (3) 90 DEGREE BENDS. 14.110 VAC GROUNDED OUTLETS SHALL BE PROVIDED ON WALLS NEAR EQUIPMENT FOR USE DURING EQUIPMENT SERVICE. 15.THE ELECTRICAL CONTRACTOR MUST SUPPLY AND INSTALL ALL INCOMING POWER CABLES FROM CIRCUIT BREAKER(S) TO THE EQUIPMENT. 16.CABLE TYPE MUST BE MTW MULTI-STRAND COPPER - NO ALUMINUM IS PERMITTED. CABLE SIZE MUST BE IN ACCORDANCE WITH POWER QUALITY REQUIREMENTS SPECIFICATIONS. 17.CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL NECESSARY HARDWARE TO ENCLOSE INCOMING POWER CABLES IN WATER TIGHT CONDUIT FROM CIRCUIT BREAKER(S) TO EQUIPMENT CABINET(S). 18.ANY CHANGES IN THE LOCATION OR TYPE OF CONDUIT, DUCT WORK, JUNCTION BOXES, ETC. MUST BE SUBMITTED IN WRITING TO THE INSTALLATION PROJECT MANAGER FOR APPROVAL. 19.ALL DUCT WORK MAKING A 90 DEGREE ANGLE MUST BE CHAMFERED FOR CABLE ACCESS. 20.JUNCTION BOX SIZES MAY BE INCREASED AS NEEDED (WITH EXCEPTION OF THE "PCDU" / "VUC" / "PCDU" / OR "PDU" JUNCTION BOX IF APPLICABLE.). 21.FIBER OPTIC CABLES FROM "REC" TO "GANT" & "REC" TO "CPU" REQUIRE A MINIMUM RADIUS OF 4 1/2". DUCT WORK DESIGN MUST ACCOMMODATE THIS REQUIREMENT. 22.EMERGENCY POWER OFF (EPO) BUTTONS REQUIRED IN CONTROL AREA, EXAMINATION ROOM AND EQUIPMENT AREA.. 23.DOOR (SAFETY) SWITCH REQUIRED ON ALL DOORS ACCESSING THE EXAMINATION ROOM IN ACCORDANCE WITH LOCAL CODES. AQUILION ONE CT SCANNER POWER REQUIREMENTS SUPPLY CONFIGURATION: 3 PHASE, 3 WIRE POWER, AND GROUND DELTA OR WYE (SEE NOTE A) NOMINAL LINE VOLTAGE: 480 VAC 60 HZ (SEE NOTE B) LINE VOLTAGE VARIATION:±10 % STEADY-STATE INCLUDING SAGS AND SURGES LINE VOLTAGE BALANCE: 2 % MAXIMUM OF NOMINAL VOLTAGE BETWEEN PHASES FREQUENCY VARIATIONS: ±1 HZ HARMONIC DISTORTION: 3% STEADY-STATE, 5% FOR SHORT PERIODS (1 MINUTE OR LESS) GROUND CONDUCTOR IMPEDANCE: 0.1 OHMS @ 60HZ, TO NEUTRAL-GROUND BONDING POINT (SEE NOTE D) STANDARD TRANSFORMER CAPACITY:150 KVA MAXIMUM SYSTEM DEMAND:125 KVA (IMAGING) CONDUCTOR SIZES (SEE NOTE E) FOR 1.5% IMPEDANCE OF BRANCH CONDUCTORS (20°C CONDUCTOR SIZE 480 VAC BREAKER FRAME SIZE 1/0 AWG 377 FT.200 A 2/0 AWG 444 FT.200 A CIRCUIT BREAKER SIZE:150 A (SEE NOTE F) MOMENTARY MAX CURRENT:163 A 150 A MAXIMUM VOLTAGE DROP:24.0 V REGULATION %:5.0% STANDARD POWER QUALITY NOTES: 1.A GROUNDED NEUTRAL POWER SOURCE IS REQUIRED TO ASSURE RELIABLE EQUIPMENT OPERATION. THE NEUTRAL CONDUCTOR MAY NOT BE USED FOR A PARTICULAR SYSTEM. 2.IN CASES WHERE MULTIPLE VOLTAGES ARE PERMITTED, THE PREFERRED SYSTEM VOLTAGE IS SPECIFIED. 3.DUE TO THE HIGH INSTANTANEOUS POWER OF MEDICAL IMAGING SYSTEMS, USE THE HIGHEST AVAILABLE VOLTAGE SOURCE. ENSURE THAT LOWER VOLTAGE SOURCES ARE DERIVED DIRECTLY FROM THE SERVICE ENTRANCE OF THE FACILITY. 4.GROUND CONDUCTORS ARE REQUIRED TO BE THE SAME SIZE AS THE PHASE CONDUCTORS UNLESS A LARGER SIZE IS REQUIRED BY CODE. 5.ALL FEEDER AND BRANCH CIRCUIT CONDUCTORS MUST BE COPPER - ALUMINUM IS NOT PERMITTED. 6.IF THE EQUIPMENT CIRCUIT BREAKER IS NOT LOCATED IN THE CONTROL AREA, A SHUNT TRIP BREAKER MUST BE USED IN ORDER TO COMPLY WITH N.E.C. 517-72(B). A PUSH-BUTTON TO OPERATE THE SHUNT TRIP MUST BE LOCATED IN THE CONTROL AREA. 7.A SEPARATE CIRCUIT, FED FROM THE FACILITY RADIOLOGY PANEL OR A MAIN SERVICE PANEL IS REQUIRED. USE OF A SUB PANEL WITH LOADS SUCH AS ELEVATORS, HVAC, MOTORS, ETC., IS NOT PERMITTED. 8.DEVICES SUCH AS UNINTERRUPTIBLE POWER SUPPLIES, POWER CONDITIONERS, VOLTAGE REGULATORS, AND FILTERS MAY BE INCOMPATIBLE WITH THIS IMAGING EQUIPMENT. CONSULT YOUR TOSHIBA SERVICE REPRESENTATIVE PRIOR TO PURCHASING OR INSTALLING THESE DEVICES. 9.IF THE MAINS POWER GROUND CONDUCTOR IS TO BE RUN WITH THE POWER PHASE CONDUCTORS, THE GROUNDS TO BUILDING STEEL OR EARTH GROUND ARE NOT TO BE RUN WITH THESE PHASE CONDUCTORS. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-2 ELECTRICAL NOTESGENERAL ELECTRICAL NOTES 1.ALL CONDUITS, TROUGHS, JUNCTION BOXES, BREAKERS, WIRES, DEVICES AND ACCESSORIES ARE FURNISHED AND INSTALLED BY CUSTOMER'S ELECTRICAL CONTRACTOR. 2.CONDUIT RUN IS NOT TO BE RUN IN SUCH A MANNER THAT WILL ALLOW CABLE POINT TO POINT LENGTHS TO BE EXCEEDED AS SHOWN IN CONDUIT LEGEND ON SHEET E-7. 3.ALL ELECTRICAL WORK SHALL BE IN COMPLIANCE WITH THE LATEST EDITION OF THE NATIONAL ELECTRICAL CODE (NFPA-70), O.S.H.A. REGULATIONS, AS WELL AS APPLICABLE REGULATIONS OF CITY, COUNTY, STATE AND FEDERAL AGENCIES. PROVIDE MATERIALS AND EQUIPMENT THAT COMPLY TO ANSI, IEEE AND NEMA STANDARDS. WHERE APPLICABLE, PROVIDE ONLY MATERIALS AND PRODUCTS THAT ARE U.L. LISTED AND LABELED. CUSTOMER'S/CONTRACTOR'S WORK SHALL COMPLY WITH THE LATEST EDITION OF NECA STANDARD OF INSTALLATION. 4.THE CONTRACTOR SHALL VERIFY EXISTING CONDITIONS IN THE FIELD TO INSURE THAT THE NEW WORK WILL FIT TO THE EXISTING STRUCTURE AS SHOWN ON THE DRAWINGS. SHOULD ANY CONDITIONS EXIST OR BE DISCOVERED THAT PREVENT THE INSTALLATION OF WORK AS SHOWN, THE CONTRACTOR SHALL NOTIFY THE OWNER'S REPRESENTATIVE PRIOR TO FABRICATION OF EQUIPMENT, OR THE PERFORMANCE OF ANY WORK THAT MAY BE AFFECTED. DO NOT ALTER DRAWINGS, DIMENSIONS, OR SPECIFICATIONS IN ANY WAY WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM THE SITE PLANNER. ALL DIMENSIONS ARE FROM FINISHED SURFACES. CONDUIT AND PULL BOXES TO BE INSTALLED BY THE CUSTOMER/CONTRACTOR WITH LOCATIONS BEING FIELD VERIFIED WITH THE EQUIPMENT INSTALLERS. 5.POWER SUPPLIES FOR THE EQUIPMENT SHALL BE DEDICATED SERVICES KEPT ENTIRELY FREE AND INDEPENDENT OF ALL OTHER BUILDING WIRING AND EQUIPMENT, SUCH AS: ELEVATORS, GENERATORS, PUMPS, HVAC SYSTEMS, ETC. THE CONTRACTOR SHALL COORDINATE THIS WORK WITH THE CUSTOMER/UTILITY COMPANY FIELD REPRESENTATIVE. 6.WORK NOT PROVIDED BY THE EQUIPMENT INSTALLATION TEAM BUT SHOWN ON DRAWINGS TO BE FURNISHED AND INSTALLED BY CUSTOMER/CONTRACTOR INCLUDES, BUT IS NOT LIMITED TO, THE FOLLOWING UNLESS NOTED OTHERWISE: ELECTRICAL RACEWAYS AND DUCTS, WIRING TROUGHS, PULL BOXES, CONDUITS, CIRCUIT BREAKERS AND EMERGENCY OFF BUTTONS, DOOR SWITCHES AND WARNING LIGHTS, WIRING, WIRING DEVICES, CONNECTORS, LIGHT FIXTURES AND GROUNDING 7.RACEWAY SHALL BE ELECTRIC METALLIC TUBING (EMT) FOR RIGID CONDUIT WORK, OR WHERE SHORT OFFSET CONNECTIONS ARE REQUIRED LIQUID TIGHT FLEXIBLE METAL CONDUIT SHALL BE USED. 8.FIELD BENDS SHALL NOT BE LESS THAN AS SHOWN IN THE NATIONAL ELECTRICAL CODE. 9.PROVIDE A JETLINE "SUPER TRUE TAPE", OR EQUIVALENT CONDUIT MEASURING TAPE FISH LINE IN ALL RACEWAYS AND CONDUITS. 10.CONDUIT BODIES SHALL NOT BE USED. WHERE A CONDUIT ENTERS A BOX, FITTING, OR OTHER ENCLOSURE, AN INSULATED THROAT CONNECTOR SHALL BE PROVIDED TO PROTECT THE WIRE FROM ABRASION. CONNECTORS SHALL BE DOUBLE SET SCREW TYPE, STEEL CONCRETE TIGHT. KEEP RACEWAYS AT LEAST 6 INCHES AWAY FROM PARALLEL RUNS OF FLUES OR STEAM AND HOT WATER PIPES. INSTALL RACEWAY RUNS ABOVE WATER AND STEAM PIPES PROVIDED THAT CABLE RUN DISTANCES ARE MAINTAINED. USE TEMPORARY CLOSURES TO PREVENT FOREIGN MATTER FROM ENTERING RACEWAY. 11.CONDUIT RUNS ARE SHOWN SCHEMATICALLY. INSTALL CONDUIT WITH A MINIMUM OF BENDS IN THE SHORTEST PRACTICAL DISTANCE CONSIDERING THE BUILDING CONSTRUCTION AND OBSTRUCTIONS, EXCEPT AS OTHERWISE INDICATED. THE CONTRACTOR SHALL MAKE CERTAIN THAT ANY CONDUIT/RACEWAY RUNS CONTAINING MEDICAL EQUIPMENT SYSTEMS CABLES DO NOT EXCEED THE SPECIFIED MAXIMUM DISTANCES AS SHOWN ON THE ELECTRICAL DETAILS. 12.PROVIDE ENCLOSED METAL RACEWAY SYSTEM (WIRE DUCT) WHERE SHOWN ON DRAWINGS WITH DIVIDERS TO SEPARATE THE DUCT (FOR POWER AND MEDICAL EQUIPMENT CABLING). DIVIDERS AND CROSSOVER PIECES TO BE PROVIDED AS NECESSARY. 13.FOR UL CERTIFIED SYSTEMS, THE CABLE TO CABLE AS WELL AS THE CIRCUIT TO CIRCUIT SEPARATION REQUIREMENT WAS EVALUATED DURING THE UL SYSTEM INVESTIGATION OF THIS EQUIPMENT. ADDITIONAL SEPARATION OF THE SYSTEM CABLE ASSEMBLIES INTO SEPARATE OR PARTITIONED RACEWAYS, UNLESS OTHERWISE NOTED, IS NOT NECESSARY TO INSURE SEPARATION OF CIRCUITS, AS THEY CAN BE IN THE SAME RACEWAY. 14.PROVIDE WIRE DUCT/RACEWAY WITH ACCESSIBLE REMOVABLE COVERS. LOCATIONS OF OPENINGS (I.E. ACCESS PANELS) TO BE CUT IN FIELD ARE TO BE COORDINATED WITH THE EQUIPMENT INSTALLATION TEAM. ELECTRICAL PULL BOXES AND RACEWAY COVERS SHALL BE INSTALLED IN A MANNER TO ALLOW ACCESSIBILITY FOR INSTALLATION AND MAINTENANCE. 15.IN-FLOOR TRENCH DUCT AND FLUSH FLOOR BOXES SHALL BE PROVIDED WITH FULLY CASKETED REMOVABLE COVERS. 16.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED HIGHER THAN 14 FEET ABOVE FINISHED FLOOR, THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL EQUIPMENT CABLES AT THE CUSTOMER EXPENSE. 17.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED ABOVE A HARD CEILING (I.E. SHEET ROCK), A 24" x 24" ACCESS PANEL IS REQUIRED AT EACH JUNCTION BOX AND WITHIN 2 FEET OF EACH 90 DEGREE ELBOW OR TEE IN WIRE DUCT/RACEWAY. THERE MUST BE FREE AND CLEAR ACCESS TO JUNCTION BOXES AND WIRE DUCT/RACEWAY. WHEN ACCESS PANELS ARE LOCATED MORE THAN 3 FEET FROM JUNCTION BOXES AND WIRE DUCT/RACEWAY THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL SUPPLIED SYSTEM CABLES AT THE CUSTOMER EXPENSE. 18.WIRING SHALL BE INSTALLED IN METAL RACEWAY, 600 VOLT CLASS, STRANDED TYPE THHN-THWN, SINGLE CONDUCTOR ANNEALED COPPER FOR A MAXIMUM OPERATING TEMPERATURE OF 75 DEGREES C (165F). SIZED AS INDICATED. THE CUSTOMER/CONTRACTOR SHALL LEAVE MINIMUM 10 FT. WIRE TAILS AT ALL OUTLET POINTS WITH WIRE IDENTIFICATION TAGGED AT BOTH ENDS FOR FINAL CONNECTION BY THE EQUIPMENT INSTALL TEAM. 19.IN ADDITION TO THE CIRCUIT BREAKER LOAD CURRENT RATING, CONSIDERATION MUST ALSO BE GIVEN TO SELECTING CIRCUIT BREAKERS THAT HAVE A HIGH ENOUGH SHORT CIRCUIT CURRENT WITHSTAND RATING TO SAFELY COORDINATE WITH THE POWER SYSTEM AVAILABLE SHORT CIRCUIT CURRENT. GENERALLY, WHEN THE 480 VOLT, 3 PHASE, EQUIPMENT IS SERVED FROM A POWER SUPPLY SYSTEM THAT IS PROVIDED WITH A 500 kVA OR SMALLER TRANSFORMER, A STANDARD 14,000 RMS AMPERE WITHSTAND RATED CIRCUIT BREAKER WILL BE ADEQUATE. HOWEVER, IF THE POWER SUPPLY SYSTEM TRANSFORMER IS LARGER THAN 500 kVA, THEN THE CIRCUIT BREAKERS HAVING A SHORT CIRCUIT WITHSTAND RATING GREATER THAN 14,000 RMS AMPERES MAY BE REQUIRED. 20.THE ELECTRICAL CONTRACTOR MUST ADJUST CONDUIT RUN LENGTHS FOR SYSTEM CABLES BASED ON THE MAXIMUM LENGTHS LISTED IN THE SYSTEM CABLE DIAGRAM. 21.THE EPO'S MUST BE INSTALLED BY A QUALIFIED ELECTRICAL CONTRACTOR ACCORDING TO NATIONAL ELECTRICAL CODE, STATE AND LOCAL REGULATIONS. MEASURES SHOULD BE TAKEN TO DESIGN THE CIRCUIT IN SUCH A WAY THAT IT WILL ALWAYS WORK WHEN THE MEDICAL EQUIPMENT IS POWERED. THE CUSTOMER IS SOLELY RESPONSIBLE FOR THE IMPLEMENTATION OF THE EPO's AND THEIR ASSOCIATED CIRCUITS AND MUST MAKE THE FINAL DETERMINATION CONSIDERING ALL SITE CONDITIONS AND REGULATORY FACTORS. THE UPS WITH THE IMS SYSTEM, LOCATED IN THE CONTROL ROOM, SUPPLIES IT'S OWN 24V DC CONTROL VOLTAGE AND NEEDS A NORMALLY CLOSED CONTACT FOR EACH EPO. THIS 24V DC CONTROL VOLTAGE IS TO BE USED FOR THIS UPS ONLY. GENERAL ELECTRICAL NOTES 1.ALL CONDUITS, TROUGHS, JUNCTION BOXES, BREAKERS, WIRES, DEVICES AND ACCESSORIES ARE FURNISHED AND INSTALLED BY CUSTOMER'S ELECTRICAL CONTRACTOR. 2.CONDUIT RUN IS NOT TO BE RUN IN SUCH A MANNER THAT WILL ALLOW CABLE POINT TO POINT LENGTHS TO BE EXCEEDED AS SHOWN IN CONDUIT LEGEND ON SHEET E-7. 3.ALL ELECTRICAL WORK SHALL BE IN COMPLIANCE WITH THE LATEST EDITION OF THE NATIONAL ELECTRICAL CODE (NFPA-70), O.S.H.A. REGULATIONS, AS WELL AS APPLICABLE REGULATIONS OF CITY, COUNTY, STATE AND FEDERAL AGENCIES. PROVIDE MATERIALS AND EQUIPMENT THAT COMPLY TO ANSI, IEEE AND NEMA STANDARDS. WHERE APPLICABLE, PROVIDE ONLY MATERIALS AND PRODUCTS THAT ARE U.L. LISTED AND LABELED. CUSTOMER'S/CONTRACTOR'S WORK SHALL COMPLY WITH THE LATEST EDITION OF NECA STANDARD OF INSTALLATION. 4.THE CONTRACTOR SHALL VERIFY EXISTING CONDITIONS IN THE FIELD TO INSURE THAT THE NEW WORK WILL FIT TO THE EXISTING STRUCTURE AS SHOWN ON THE DRAWINGS. SHOULD ANY CONDITIONS EXIST OR BE DISCOVERED THAT PREVENT THE INSTALLATION OF WORK AS SHOWN, THE CONTRACTOR SHALL NOTIFY THE OWNER'S REPRESENTATIVE PRIOR TO FABRICATION OF EQUIPMENT, OR THE PERFORMANCE OF ANY WORK THAT MAY BE AFFECTED. DO NOT ALTER DRAWINGS, DIMENSIONS, OR SPECIFICATIONS IN ANY WAY WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM THE SITE PLANNER. ALL DIMENSIONS ARE FROM FINISHED SURFACES. CONDUIT AND PULL BOXES TO BE INSTALLED BY THE CUSTOMER/CONTRACTOR WITH LOCATIONS BEING FIELD VERIFIED WITH THE EQUIPMENT INSTALLERS. 5.POWER SUPPLIES FOR THE EQUIPMENT SHALL BE DEDICATED SERVICES KEPT ENTIRELY FREE AND INDEPENDENT OF ALL OTHER BUILDING WIRING AND EQUIPMENT, SUCH AS: ELEVATORS, GENERATORS, PUMPS, HVAC SYSTEMS, ETC. THE CONTRACTOR SHALL COORDINATE THIS WORK WITH THE CUSTOMER/UTILITY COMPANY FIELD REPRESENTATIVE. 6.WORK NOT PROVIDED BY THE EQUIPMENT INSTALLATION TEAM BUT SHOWN ON DRAWINGS TO BE FURNISHED AND INSTALLED BY CUSTOMER/CONTRACTOR INCLUDES, BUT IS NOT LIMITED TO, THE FOLLOWING UNLESS NOTED OTHERWISE: ELECTRICAL RACEWAYS AND DUCTS, WIRING TROUGHS, PULL BOXES, CONDUITS, CIRCUIT BREAKERS AND EMERGENCY OFF BUTTONS, DOOR SWITCHES AND WARNING LIGHTS, WIRING, WIRING DEVICES, CONNECTORS, LIGHT FIXTURES AND GROUNDING 7.RACEWAY SHALL BE ELECTRIC METALLIC TUBING (EMT) FOR RIGID CONDUIT WORK, OR WHERE SHORT OFFSET CONNECTIONS ARE REQUIRED LIQUID TIGHT FLEXIBLE METAL CONDUIT SHALL BE USED. 8.FIELD BENDS SHALL NOT BE LESS THAN AS SHOWN IN THE NATIONAL ELECTRICAL CODE. 9.PROVIDE A JETLINE "SUPER TRUE TAPE", OR EQUIVALENT CONDUIT MEASURING TAPE FISH LINE IN ALL RACEWAYS AND CONDUITS. 10.CONDUIT BODIES SHALL NOT BE USED. WHERE A CONDUIT ENTERS A BOX, FITTING, OR OTHER ENCLOSURE, AN INSULATED THROAT CONNECTOR SHALL BE PROVIDED TO PROTECT THE WIRE FROM ABRASION. CONNECTORS SHALL BE DOUBLE SET SCREW TYPE, STEEL CONCRETE TIGHT. KEEP RACEWAYS AT LEAST 6 INCHES AWAY FROM PARALLEL RUNS OF FLUES OR STEAM AND HOT WATER PIPES. INSTALL RACEWAY RUNS ABOVE WATER AND STEAM PIPES PROVIDED THAT CABLE RUN DISTANCES ARE MAINTAINED. USE TEMPORARY CLOSURES TO PREVENT FOREIGN MATTER FROM ENTERING RACEWAY. 11.CONDUIT RUNS ARE SHOWN SCHEMATICALLY. INSTALL CONDUIT WITH A MINIMUM OF BENDS IN THE SHORTEST PRACTICAL DISTANCE CONSIDERING THE BUILDING CONSTRUCTION AND OBSTRUCTIONS, EXCEPT AS OTHERWISE INDICATED. THE CONTRACTOR SHALL MAKE CERTAIN THAT ANY CONDUIT/RACEWAY RUNS CONTAINING MEDICAL EQUIPMENT SYSTEMS CABLES DO NOT EXCEED THE SPECIFIED MAXIMUM DISTANCES AS SHOWN ON THE ELECTRICAL DETAILS. 12.PROVIDE ENCLOSED METAL RACEWAY SYSTEM (WIRE DUCT) WHERE SHOWN ON DRAWINGS WITH DIVIDERS TO SEPARATE THE DUCT (FOR POWER AND MEDICAL EQUIPMENT CABLING). DIVIDERS AND CROSSOVER PIECES TO BE PROVIDED AS NECESSARY. 13.FOR UL CERTIFIED SYSTEMS, THE CABLE TO CABLE AS WELL AS THE CIRCUIT TO CIRCUIT SEPARATION REQUIREMENT WAS EVALUATED DURING THE UL SYSTEM INVESTIGATION OF THIS EQUIPMENT. ADDITIONAL SEPARATION OF THE SYSTEM CABLE ASSEMBLIES INTO SEPARATE OR PARTITIONED RACEWAYS, UNLESS OTHERWISE NOTED, IS NOT NECESSARY TO INSURE SEPARATION OF CIRCUITS, AS THEY CAN BE IN THE SAME RACEWAY. 14.PROVIDE WIRE DUCT/RACEWAY WITH ACCESSIBLE REMOVABLE COVERS. LOCATIONS OF OPENINGS (I.E. ACCESS PANELS) TO BE CUT IN FIELD ARE TO BE COORDINATED WITH THE EQUIPMENT INSTALLATION TEAM. ELECTRICAL PULL BOXES AND RACEWAY COVERS SHALL BE INSTALLED IN A MANNER TO ALLOW ACCESSIBILITY FOR INSTALLATION AND MAINTENANCE. 15.IN-FLOOR TRENCH DUCT AND FLUSH FLOOR BOXES SHALL BE PROVIDED WITH FULLY CASKETED REMOVABLE COVERS. 16.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED HIGHER THAN 14 FEET ABOVE FINISHED FLOOR, THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL EQUIPMENT CABLES AT THE CUSTOMER EXPENSE. 17.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED ABOVE A HARD CEILING (I.E. SHEET ROCK), A 24" x 24" ACCESS PANEL IS REQUIRED AT EACH JUNCTION BOX AND WITHIN 2 FEET OF EACH 90 DEGREE ELBOW OR TEE IN WIRE DUCT/RACEWAY. THERE MUST BE FREE AND CLEAR ACCESS TO JUNCTION BOXES AND WIRE DUCT/RACEWAY. WHEN ACCESS PANELS ARE LOCATED MORE THAN 3 FEET FROM JUNCTION BOXES AND WIRE DUCT/RACEWAY THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL SUPPLIED SYSTEM CABLES AT THE CUSTOMER EXPENSE. 18.WIRING SHALL BE INSTALLED IN METAL RACEWAY, 600 VOLT CLASS, STRANDED TYPE THHN-THWN, SINGLE CONDUCTOR ANNEALED COPPER FOR A MAXIMUM OPERATING TEMPERATURE OF 75 DEGREES C (165F). SIZED AS INDICATED. THE CUSTOMER/CONTRACTOR SHALL LEAVE MINIMUM 10 FT. WIRE TAILS AT ALL OUTLET POINTS WITH WIRE IDENTIFICATION TAGGED AT BOTH ENDS FOR FINAL CONNECTION BY THE EQUIPMENT INSTALL TEAM. 19.IN ADDITION TO THE CIRCUIT BREAKER LOAD CURRENT RATING, CONSIDERATION MUST ALSO BE GIVEN TO SELECTING CIRCUIT BREAKERS THAT HAVE A HIGH ENOUGH SHORT CIRCUIT CURRENT WITHSTAND RATING TO SAFELY COORDINATE WITH THE POWER SYSTEM AVAILABLE SHORT CIRCUIT CURRENT. GENERALLY, WHEN THE 480 VOLT, 3 PHASE, EQUIPMENT IS SERVED FROM A POWER SUPPLY SYSTEM THAT IS PROVIDED WITH A 500 kVA OR SMALLER TRANSFORMER, A STANDARD 14,000 RMS AMPERE WITHSTAND RATED CIRCUIT BREAKER WILL BE ADEQUATE. HOWEVER, IF THE POWER SUPPLY SYSTEM TRANSFORMER IS LARGER THAN 500 kVA, THEN THE CIRCUIT BREAKERS HAVING A SHORT CIRCUIT WITHSTAND RATING GREATER THAN 14,000 RMS AMPERES MAY BE REQUIRED. 20.THE ELECTRICAL CONTRACTOR MUST ADJUST CONDUIT RUN LENGTHS FOR SYSTEM CABLES BASED ON THE MAXIMUM LENGTHS LISTED IN THE SYSTEM CABLE DIAGRAM. 21.THE EPO'S MUST BE INSTALLED BY A QUALIFIED ELECTRICAL CONTRACTOR ACCORDING TO NATIONAL ELECTRICAL CODE, STATE AND LOCAL REGULATIONS. MEASURES SHOULD BE TAKEN TO DESIGN THE CIRCUIT IN SUCH A WAY THAT IT WILL ALWAYS WORK WHEN THE MEDICAL EQUIPMENT IS POWERED. THE CUSTOMER IS SOLELY RESPONSIBLE FOR THE IMPLEMENTATION OF THE EPO's AND THEIR ASSOCIATED CIRCUITS AND MUST MAKE THE FINAL DETERMINATION CONSIDERING ALL SITE CONDITIONS AND REGULATORY FACTORS. THE UPS WITH THE IMS SYSTEM, LOCATED IN THE CONTROL ROOM, SUPPLIES IT'S OWN 24V DC CONTROL VOLTAGE AND NEEDS A NORMALLY CLOSED CONTACT FOR EACH EPO. THIS 24V DC CONTROL VOLTAGE IS TO BE USED FOR THIS UPS ONLY. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-3 CT ELECTRICAL LEGENDELEC SYM. ITEM DESCRIPTION ALL SUPPLIED AND INSTALLED BY CUSTOMER / CONTRACTOR UPS JB1 SPK2 RECSPK1 CPU GANT EPO WL CB MAIN MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F. TO BOTTOM OF BOX. 6' WHIP OF PROVIDE FLEXIBLE CONDUIT AND CONDUCTORS FOR PHYSICAL CONNECTION TO "PCDU". 4" STANDARD J-BOX FOR "WARNING LIGHT" MOUNTED ABOVE PATIENT ENTRY DOOR, SEE WIRING DETAIL, FIXTURE AND CONTROLLER BY CUSTOMER. 4" STANDARD J-BOX FOR "EMERGENCY POWER OFF" BUTTON. DPDT, NORMALLY OPEN MUSHROOM HEAD PUSH BUTTON MOUNTED AT 60" A.F.F. WITH PROTECTIVE COVER TIED INTO "CB". 4" SHALLOW J-BOX FOR SCAN ROOM SPEAKER, SURFACE MOUNTED ON WALL AT 50" A.F.F. TO BOTTOM OF BOX ATOP "WM". MINIMUM 10"W X 10"H X 3-1/2"D PULL-BOX, FLUSH MOUNTED IN WALL @ 12" A.F.F. TO BOTTOM OF BOX, NO COVER, OPENS INTO BACK OF "WD". GROMMETED OPENING IN COVER OF WALL TROUGH, COORDINATE ACTUAL SIZE AND LOCATION WITH INSTALLERS, (IF BEING PROVIDED). GROMMETED OPENINGS IN TROUGHS FOR CABLING, COORDINATE ACTUAL SIZE AND LOCATION WITH INSTALLERS. MAIN 480V OR 208V, 3-PHASE, 3-WIRE ELECTRICAL SERVICE FOR CT EQUIPMENT. WIRE SIZED PER REQUIREMENTS SHT. E-1, PROVIDE FULL SIZE GROUND WIRE. MINIMUM 16"X16"X6" DEEP JUNCTION BOX SET FLUSH WITH FINISH FLOOR, PROVIDE GROMMETED OPENING IN COVER, COORDINATE ACTUAL SIZE AND LOCATION OF OPENING WITH INSTALLERS. NOTES: 1.ALL CONDUITS, TROUGHS AND RACEWAYS ARE FURNISHED AND INSTALLED BY THE CUSTOMER'S ELECTRICAL CONTRACTOR. 2.ALL CONDUITS, TROUGHS AND RACEWAYS MUST TAKE THE SHORTEST POSSIBLE ROUTE. 3.CONDUIT RUNS 7 & 11 ARE FOR FIBER OPTIC CABLES, MINIMUM 4.5" RADIUS BENDS. 4.JUNCTION BOX SIZES ARE APPROXIMATE, BOX SIZES MAY BE INCREASED AS NECESSARY TO ACCOMMODATE ALL INCOMING CONDUITS. 5.CUSTOMER'S ELECTRICIAN MUST BE ON SITE WHEN THE EQUIPMENT IS DELIVERED TO VERIFY PHASE ROTATION, (CLOCKWISE), TO LAND PRIMARY WIRES, AND TO COORDINATE ALL GROMMETED OPENINGS REQUIRED.OC DEDICATED 120V ELECTRICAL OUTLETS (QUAD) FOR SYSTEM EQUIPMENT AND SERVICE EQUIPMENT. "EMERGENCY POWER OFF" BUTTON MOUNTED AT 60" A.F.F. WITH PROTECTIVE COVER OR GUARD WARNING LIGHT - REQUIRED AT ALL ENTRY POINTS INTO CT SCAN ROOM, COMES ON WHEN SCANNING ELECTRICAL JUNCTION BOX WALL TROUGH FLOOR TROUGH, SURFACE OR RECESSED MINIMUM 10"W X 2-1/2"D SURFACE MOUNTED WALL DUCT WITH 2 EQUAL COMPARTMENTS AND REMOVABLE COVERS ENTIRE LENGTH. MOUNT 12" A.F.F. TO BOTTOM OF DUCT. PROVIDE GROMMETED OPENING INTO "JB1". WT WLS "WARNING LIGHT SWITCH" LOCATED IN SCAN ROOM FOR SERVICE PERSONNEL TO TURN OFF LINE VOLTAGE TO CIRCUIT INSIDE CT GANTRY. REC MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F. TO BOTTOM OF BOX. GROMMETED OPENING IN BOTTOM OF BOX FOR SYSTEM CABLES. ELEC SYM. ITEM DESCRIPTION ALL SUPPLIED AND INSTALLED BY CUSTOMER / CONTRACTOR UPS JB1 SPK2 RECSPK1 CPU GANT EPO WL PDU CB MAIN MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F. TO BOTTOM OF BOX. 6' WHIP OF PROVIDE FLEXIBLE CONDUIT AND CONDUCTORS FOR PHYSICAL CONNECTION TO "PCDU". 4" STANDARD J-BOX FOR "WARNING LIGHT" MOUNTED ABOVE PATIENT ENTRY DOOR, SEE WIRING DETAIL, FIXTURE AND CONTROLLER BY CUSTOMER. 4" STANDARD J-BOX FOR "EMERGENCY POWER OFF" BUTTON. DPDT, NORMALLY OPEN MUSHROOM HEAD PUSH BUTTON MOUNTED AT 60" A.F.F. WITH PROTECTIVE COVER TIED INTO "CB". 4" SHALLOW J-BOX FOR SCAN ROOM SPEAKER, SURFACE MOUNTED ON WALL AT 50" A.F.F. TO BOTTOM OF BOX ATOP "WM". MINIMUM 10"W X 10"H X 3-1/2"D PULL-BOX, FLUSH MOUNTED IN WALL @ 12" A.F.F. TO BOTTOM OF BOX, NO COVER, OPENS INTO BACK OF "WD". GROMMETED OPENING IN COVER OF WALL TROUGH, COORDINATE ACTUAL SIZE AND LOCATION WITH INSTALLERS, (IF BEING PROVIDED). GROMMETED OPENINGS IN TROUGHS FOR CABLING, COORDINATE ACTUAL SIZE AND LOCATION WITH INSTALLERS. MAIN 480V OR 208V, 3-PHASE, 3-WIRE ELECTRICAL SERVICE FOR CT EQUIPMENT. WIRE SIZED PER REQUIREMENTS SHT. E-1, PROVIDE FULL SIZE GROUND WIRE. DEDICATED SHUNT-TRIP CIRCUIT BREAKER FOR CT EQUIPMENT PER ELECTRICAL REQUIREMENTS. ACTUAL LOCATION TO BE DETERMINED BY CUSTOMER'S ELECTRICAL ENGINEER. RUN CONDUIT AND WIRES AS REQUIRED INTO "PDU" BOX, PROVIDE 6' LIQUID TIGHT WHIP AT "PDU" BOX TO CONNECT EQUIPMENT. NOTES: 1.ALL CONDUITS, TROUGHS AND RACEWAYS ARE FURNISHED AND INSTALLED BY THE CUSTOMER'S ELECTRICAL CONTRACTOR. 2.ALL CONDUITS, TROUGHS AND RACEWAYS MUST TAKE THE SHORTEST POSSIBLE ROUTE. 3.CONDUIT RUNS 7 & 11 ARE FOR FIBER OPTIC CABLES, MINIMUM 4.5" RADIUS BENDS. 4.JUNCTION BOX SIZES ARE APPROXIMATE, BOX SIZES MAY BE INCREASED AS NECESSARY TO ACCOMMODATE ALL INCOMING CONDUITS. 5.CUSTOMER'S ELECTRICIAN MUST BE ON SITE WHEN THE EQUIPMENT IS DELIVERED TO VERIFY PHASE ROTATION, (CLOCKWISE), TO LAND PRIMARY WIRES, AND TO COORDINATE ALL GROMMETED OPENINGS REQUIRED. RUN CONDUIT CONDUIT CONDUIT CABLE CABLE LENGTH CABLE NO.POINT TO POINT ROUTING DIAMETER POINT TO POINT USABLE LENGTH FURNISH/INSTALLED BY 1 CB PDU AS REQ'D PER NEC CB PCDU AS REQ'D + 6' WHIP CONTRACTOR 2 CB EPO AS REQ'D PER NEC CB EPO AS REQ'D CONTRACTOR 3 PDU GANT UNDER SLAB (2) 3"PCDU GANT PER OEM INSTALLERS 4 PDU JB1 UNDER SLAB (2) 3"PCDU JB1 PER OEM INSTALLERS 5 PDU EPO AS REQ'D 1 2"PCDU EPO AS REQ'D CONTRACTOR 6 REC1 GANT UNDER SLAB 3"REC GANT PER OEM INSTALLERS 7 REC2 JB1 UNDER SLAB (2) 3"REC JB1 PER OEM INSTALLERS 8 REC JB1 UNDER SLAB 1"REC JB1 PER OEM INSTALLERS 9 GANT JB1 UNDER SLAB (2) 3"GANT JB1 PER OEM INSTALLERS 10 WL GANT UNDER SLAB 1 2"WL GANT (2) #12 WIRES CONTRACTOR 11 SPK2 JB1 OVER CEILING 1 2"SPK2 CPU PER OEM INSTALLERS OC RJ45 CONNECTOR, NETWORK DATA CONNECTION TO CUSTOMER NETWORK. DEDICATED 120V ELECTRICAL OUTLETS (QUAD) FOR SYSTEM EQUIPMENT AND SERVICE EQUIPMENT. "EMERGENCY POWER OFF" BUTTON MOUNTED AT 60" A.F.F. WITH PROTECTIVE COVER OR GUARD WARNING LIGHT - REQUIRED AT ALL ENTRY POINTS INTO CT SCAN ROOM, COMES ON WHEN SCANNING ELECTRICAL JUNCTION BOX WALL TROUGH FLOOR TROUGH, SURFACE OR RECESSED MINIMUM 10"W X 2-1/2"D SURFACE MOUNTED WALL DUCT WITH 2 EQUAL COMPARTMENTS AND REMOVABLE COVERS ENTIRE LENGTH. MOUNT 12" A.F.F. TO BOTTOM OF DUCT. PROVIDE GROMMETED OPENING INTO "JB1". WT WLS "WARNING LIGHT SWITCH" LOCATED IN SCAN ROOM FOR SERVICE PERSONNEL TO TURN OFF LINE VOLTAGE TO CIRCUIT INSIDE CT GANTRY. REC MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F. TO BOTTOM OF BOX. GROMMETED OPENING IN BOTTOM OF BOX FOR SYSTEM CABLES. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-4 CT ELECTRICAL PLANNorth EPO EPO 2 PCDU 1'-8" CL REMOTE CONNECTIVITY CONNECTION W/110V RECEPT. SPK2 WL CB 10 120V POWER SOURCE, SEE DETAIL E-5 GANT CABLE ACCESS AT GANTRY EXISTING ELECT. TRANS. EXIST ELECT PANEL 12' CONDUIT, WIRES BY E.C. EMPTY CONDUIT BY E.C. JB1 WD 1 5 3 4 REC1 REC2 6 7 8 9 11 NOTES: 1.ALL CONDUITS ARE SHOWN SCHEMATICALLY, ALL CONDUITS TO TAKE THE MOST DIRECT ROUTE. 2.J-BOXES MAY BE INCREASED IN SIZE AS NECESSARY. 3.THE EPO CIRCUIT DESIGN IS BY THE ELECTRICAL ENGINEER OF RECORDS. OC INV SPK1 REC JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-5 ELECTRICAL DETAILSPCDU BSCBEPOUPS IN USE OR WARNING LIGHT DETAIL SHUT OFF SWITCH, LOCATE IN CT SCAN ROOM FOR SERVICE X-RAY IN USE OR WARNING LIGHT FURNISHED AND INSTALLED BY CUSTOMER, LOCATED ABOVE ENTRY DOOR. INTERLOCK WITH CT SYSTEM SUPPLIED TOSHIBA SUPPLIED CUSTOMER NEUTRAL 120V ONLY REQUIRED IF OPTIONAL UPS IS INSTALLED GROUND SOURCE (PER NEC SECT. 250) CB MAIN BREAKER SHUNT TRIP EPO "Y" * G "X" * 1 Hr ACCEPTABLE VOLTAGE RANGE SYSTEM DAMAGE MAY OCCUR (DISRUPTIVE IMPULSES) SYSTEM DISRUPTION MAY OCCUR (DISRUPTIVE IMPULSES) 200% ABOVE NOMINAL VOLTAGE 100% ABOVE NOMINAL VOLTAGE NOMINAL VOLTAGE +10 -10% BELOW NOMINAL SYSTEM DISRUPTION MAY OCCUR (LOW VOLTAGE, SAGS, & OUTAGE) 1 Min1 Sec100 mS10 mS1 mS100 µS10 µS SYSTEM ELECTRICAL REQUIREMENTS AQUILION WITH "PCDU" SUPPLY CONFIGURATION:3 PHASE DELTA OR WYE CONNECTED, 150KVA SERVICE SUPPLY VOLTAGE:480V, 60HZ, 100 AMP OR 208V, 60HZ, 225 AMP FOR CONDUCTOR SIZE USING COMBINED LENGTHS FOR "X" AND "Y" SEE CONDUCTOR SIZE TABLE. NOTES: 1.UPS MAY, OR MAY NOT, APPLY. 2.ALL CONDUITS, CABLES, AND CONNECTIONS ARE BY CUSTOMER'S ELECTRICAL CONTRACTOR. MAIN BREAKER AND PCDU WIRING DIAGRAM POWER QUALITY REQUIREMENTS CT GANTRY SYSTEM DAMAGE IS LIKELY OVER OR UNDER NOMINAL VOLTAGE INCLUDING HIGH VOLTAGE, SWELLS, SURGES, & SAGS. FAILURES DUE TO INCOMING POWER IS NOT COVERED BY WARRANTY OR SERVICE AGREEMENT PCDU BSCBEPOUPS SHUT OFF SWITCH, LOCATE IN CT SCAN ROOM FOR SERVICE X-RAY IN USE OR WARNING LIGHT FURNISHED AND INSTALLED BY CUSTOMER, LOCATED ABOVE ENTRY DOOR. INTERLOCK WITH CT SYSTEM SUPPLIED TOSHIBA SUPPLIED CUSTOMER NEUTRAL 120V ONLY REQUIRED IF OPTIONAL UPS IS INSTALLED PUSH BUTTON IN WITH NORMALLY OPEN CONTACT IN CONTROL & SCAN ROOM TO CONTROL SHUNT TRIP BREAKER 200V 150 amp 200V 30 amp 200V 50 amp CONSOLE GROUND SOURCE (PER NEC SECT. 250) SUPPLIED WITH SYSTEM SEE NOTE 1 NOTE 1 - TOSHIBA SUPPLIED POWER CABLES BETWEEN "PDU" AND TOSHIBA EQUIPMENT ARE TO USED IF THE LENGTHS PROVIDED ARE SUFFICIENT, IN CASES WHERE EXTENDED CABLES ARE REQUIRED THE CUSTOMER'S ELECTRICAL CONTRACTOR MUST PROVIDE CABLES PER LOCAL/NATIONAL CODES. MAIN SERVICE 480V 3-P GANTRY PDU CB MAIN BREAKER SHUNT TRIP EPO "Y" * G "X" * 1 Hr ACCEPTABLE VOLTAGE RANGE SYSTEM DAMAGE MAY OCCUR (DISRUPTIVE IMPULSES) SYSTEM DISRUPTION MAY OCCUR (DISRUPTIVE IMPULSES) 200% ABOVE NOMINAL VOLTAGE 100% ABOVE NOMINAL VOLTAGE NOMINAL VOLTAGE -10% BELOW NOMINAL SYSTEM DISRUPTION MAY OCCUR (LOW VOLTAGE, SAGS, & OUTAGE) 1 Min1 Sec100 mS10 mS1 mS100 µS10 µS SYSTEM ELECTRICAL REQUIREMENTS AQUILION WITH "PCDU" SUPPLY CONFIGURATION:3 PHASE DELTA OR WYE CONNECTED, 150KVA SERVICE SUPPLY VOLTAGE:480V, 60HZ, 100 AMP OR 208V, 60HZ, 225 AMP FOR CONDUCTOR SIZE USING COMBINED LENGTHS FOR "X" AND "Y" SEE CONDUCTOR SIZE TABLE. NOTES: 1.UPS MAY, OR MAY NOT, APPLY. 2.ALL CONDUITS, CABLES, AND CONNECTIONS ARE BY CUSTOMER'S ELECTRICAL CONTRACTOR. MAIN BREAKER AND PCDU WIRING DIAGRAM POWER QUALITY REQUIREMENTS LIGHT ON SIGNAL FROM GANTRY CT GANTRY SYSTEM DAMAGE IS LIKELY OVER OR UNDER NOMINAL VOLTAGE INCLUDING HIGH VOLTAGE, SWELLS, SURGES, & SAGS. FAILURES DUE TO INCOMING POWER IS NOT COVERED BY WARRANTY OR SERVICE AGREEMENT! % ABOVE NOMINAL G GROUND ELECTRODE PER NEC 250-30(A) BUILDING STEEL, WATER PIPE OR GROUNDING ELECTRODE AS CLOSE AS PHYSICALLY POSSIBLE TO THE "PDU". GROUND CONDUCTORS MUST BE THE SAME SIZE AS PHASE CONDUCTORS. NOTES: 1."X-RAY ON" OR WARNING LIGHT TO BE PROVIDED BY CUSTOMER/CONTRACTOR. 2.DO NOT USE FLUORESCENT FIXTURES, LED CAN BE USED IF CUSTOMER/CONTRACTOR PROVIDES REPLAY PER LAMP WATTAGE, MINIMUM RATING 2 AMP. JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-6 ELECTRICAL DETAILSCT VIEW WINDOW TD CT SCAN RM. FLOOR TYPICAL TRENCH AND WALL DUCT DETAILS TYPICAL CONTROL ROOM CONFIGURATION NOTE: VERIFY WITH EQUIPMENT VENDOR IF A CONSOLE WAS INCLUDED IN YOUR ORDR, IF NOT THE CUSTOMER MUST SUPPLY A COUNTERTOP OR TABLE. TYPICAL SCAN ROOM ELEVATION "JB" FLUSH MOUNTED, OPEN TO "WD" ALL TROUGHS TO HAVE REMOVABLE COVERS ALL TROUGHS TO BE 3 COMPARTMENTS 45 DEGREE CORNERS SURFACE MOUNTED TROUGHS, 3 EQUAL COMPARTMENTS TYPICAL PCDU AND J-BOX ARRANGEMENT SHOWN USING WALL TROUGH FINISHED WALL PCDU CONDUITS OR TROUGH FROM "PCDU" TO "GANT" AS SITE CONDITIONS ALLOW CT VIEW WINDOW CPU WT PDU VT PDU TD CT SCAN RM. FLOOR TYPICAL TRENCH AND WALL DUCT DETAILS RECUPS TYPICAL CONTROL ROOM CONFIGURATION NOTE: VERIFY WITH EQUIPMENT VENDOR IF A CONSOLE WAS INCLUDED IN YOUR ORDR, IF NOT THE CUSTOMER MUST SUPPLY A COUNTERTOP OR TABLE. TYPICAL SCAN ROOM ELEVATION "JB" FLUSH MOUNTED, OPEN TO "WD" ALL TROUGHS TO HAVE REMOVABLE COVERS ALL TROUGHS TO BE 3 COMPARTMENTS 45 DEGREE CORNERS SURFACE MOUNTED TROUGHS, 3 EQUAL COMPARTMENTS TYPICAL PCDU AND J-BOX ARRANGEMENT SHOWN USING WALL TROUGH FINISHED WALL PDU CONDUITS OR TROUGH FROM "PCDU" TO "GANT" AS SITE CONDITIONS ALLOW CONDUITS FROM GANT & PCDU JB1 CONDUITS TO JB1 10" X 10" X 4"D J-BOX FLUSH IN WALL, OPEN TO WALL DUCT. WALL DUCT FLEX CONDUIT TO "PDU" BY ELECTRICAL CONTRACTOR JOB NUMBER SHEET PROJECT DATE REVISION HISTORY 1. 2. 3. 4. 5. 6. 2-14-24 2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES 5335 Avion Park Drive Unit A Highland Heights, Ohio (888)505-0319 These drawings and all information contained herein are the property of PrizMED Imaging Solutions and are to be treated as confidential. These drawings are not to be reproduced, copied, distributed, disclosed or published directly or indirectly without the expressed written authorization of PrizMED Imaging Solutions. WORK ORDER NUMBER W-4849 2-22-24 WILL IT FIT SKETCH 2-28-24 PRELIMS 3-10-24 FINALS ISSUED E-7 SYSTEM CABLE DIAGRAMRUN NO. CABLE (END TO END) CABLE LENGTH (MAX. USABLE) 39'-4" 59'-0" 55'-9" 75'-0" 18'-0" 52'-5" 39'-4" 52'-5" 26'-2" 52'-5" 75'-0" 39'-4" 52'-5" 18'-4" 18'-4" 18'-4" "POWER SYSTEM" TO "GANT" (39'-4") - (POWER & GROUND) "POWER SYSTEM" TO "CPU" (59'-0") - (POWER) "GANT" TO "CPU" (55'-9") - (POWER) "CPU" TO "SPK1" (26'-2") - (SIGNAL) "SPK2" TO "CPU" (52'-5") - (SIGNAL) "INV" TO "GANT" (75'-0") - (SIGNAL) "CPU" TO "INV" (75'-0") - (SIGNAL) "REC" TO "CPU" (52'-5" POWER) "REC" TO "CPU" (65'-7" FIBER & SIGNAL) "REC" TO "GANT" (36'-1" FIBER) "GANT" TO "CPU" (52'-5") - (SIGNAL) A B C D F G E H I J K POWER SYSTEM POWER SYSTEM NOTES: 1)THIS CABLE DIAGRAM IS FOR REFERENCE ONLY AND MAY NOT CONTAIN THE EXACT EQUIPMENT AS SHOWN IN THE ATTACHED DRAWING PACKAGE. 2)THE ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR ENSURING THE CONDUIT RUNS DO NOT EXCEED THE CABLE LENGTHS SHOWN. 3)ALL CABLE CONNECTIONS SHOWN ARE THE MAXIMUM LENGTH AND CABLES CANNOT BE EXTENDED. 4)THIRD PARTY ITEM CONNECTIONS TO BE VERIFIED WITH VENDOR OF SUCH ITEMS. 5)POWER AND SIGNAL CABLES MUST BE RUN IN SEPARATE CONDUIT OR SEPARATE COMPARTMENTS IN TROUGHS. SPK2 GANTREC CPU SPK1 INV GANT CPU CPUGANT GANT CPU SPK1CPU SPK2 CPU INV INV REC REC REC GANT CPU CPU CPU GANT A B C D E F G H I J K PCDU RUN NO. CABLE (END TO END) CABLE LENGTH (MAX. USABLE) 39'-4" 59'-0" 55'-9" 75'-0" 18'-0" 52'-5" 39'-4" 52'-5" 26'-2" 52'-5" 75'-0" 39'-4" 52'-5" 18'-4" 18'-4" 18'-4" "POWER SYSTEM" TO "GANT" (39'-4") - (POWER & GROUND) "POWER SYSTEM" TO "CPU" (59'-0") - (POWER) "GANT" TO "CPU" (55'-9") - (POWER) "CPU" TO "SPK1" (26'-2") - (SIGNAL) "SPK2" TO "CPU" (52'-5") - (SIGNAL) "INV" TO "GANT" (75'-0") - (SIGNAL) "CPU" TO "INV" (75'-0") - (SIGNAL) "REC" TO "CPU" (52'-5" POWER) "REC" TO "CPU" (65'-7" FIBER & SIGNAL) "REC" TO "GANT" (36'-1" FIBER) "GANT" TO "CPU" (52'-5") - (SIGNAL) A B C D F G E H I J K POWER SYSTEM POWER SYSTEM NOTES: 1)THIS CABLE DIAGRAM IS FOR REFERENCE ONLY AND MAY NOT CONTAIN THE EXACT EQUIPMENT AS SHOWN IN THE ATTACHED DRAWING PACKAGE. 2)THE ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR ENSURING THE CONDUIT RUNS DO NOT EXCEED THE CABLE LENGTHS SHOWN. 3)ALL CABLE CONNECTIONS SHOWN ARE THE MAXIMUM LENGTH AND CABLES CANNOT BE EXTENDED. 4)THIRD PARTY ITEM CONNECTIONS TO BE VERIFIED WITH VENDOR OF SUCH ITEMS. 5)POWER AND SIGNAL CABLES MUST BE RUN IN SEPARATE CONDUIT OR SEPARATE COMPARTMENTS IN TROUGHS. SPK2 GANTREC CPU SPK1 INV GANT CPU CPUGANT GANT CPU SPK1CPU SPK2 CPU INV INV REC REC REC GANT CPU CPU CPU GANT A B C D E F G H I J K AQUILION 6ONE CT SYSTEM CABLE LENGTHS * THESE ARE THE ORIGINAL LENGTHS, PRE-OWNED SYTEMS MAY HAVE DIFFERENT CABLE SETS PCDU