W4849 - Toshiba AQ One Final Drawings - Integrity VeterinaryTHESE DRAWINGS ARE FORMATTED FOR 11 X 17 PAPER
S-1 STRUCTURAL & ANCHORING INFORMATION
M-1 ENVIRONMENTAL INFORMATION
E-1 ELECTRICAL NOTES
E-2 ELECTRICAL NOTES
E-3 CT ELECTRICAL LEGEND
E-4 CT ELECTRICAL PLAN
E-5 ELECTRICAL DETAILS
E-6 ELECTRICAL DETAILS
E-7 CT SYSTEM CABLES
SUBMISSION HISTORY
2-21-24 TEST FIT SKETCH
2-28-24 PRELIMS ISSUED
3-10-24 FINALS ISSUED
MEDICAL EQUIPMENT SITE PLANNING INFORMATION
PROVIDED BY
*** ADDITIONAL REQUIRED REFERENCE MATERIAL, SEPARATE DOCUMENTS ***
TOSHIBA SITE PLANNING GUIDE NO. 2A201 536EN*G
THESE DRAWINGS ARE PROVIDED FOR COORDINATION PURPOSES AND ARE NOT TO BE USED FOR PERMITTING OR CONSTRUCTION
5335 AVION PARK DR. UNIT A
HIGHLAND HEIGHTS, OH 44143
THESE DRAWINGS ARE ISSUED TO THE CUSTOMER FOR THE SOLE PURPOSE OF COORDINATING THE EQUIPMENT
REQUIREMENTS WITH THE CUSTOMER'S PROFESSIONAL DESIGN TEAM AND ARE NOT TO BE USED FOR PERMITTING
OR CONSTRUCTION PURPOSES.
THIS IS YOUR FINAL SET OF DRAWINGS ISSUED FOR USE OF YOUR DESIGN TEAM, PLEASE SIGN WHERE INDICATED
ON SHEET "APRV" AND RETURN TO YOUR PROJECT MANGER WITHIN 5 BUSINESS DAYS TO RELEASE YOUR
EQUIPMENT FOR STAGING AND PREP.
FINAL DRAWING INDEX
COVER SHEET
APRV CUSTOMER ACKNOWLEDGEMENT AND APPROVAL
EDA EQUIPMENT DELIVERY AUTHORIZATION
N-1 GENERAL NOTES
N-2 GENERAL NOTES
A-1 FLOOR PLAN
A-2 CT RIGGING PATH
CT-1 CT EQUIPMENT LEGEND
CT-2 CT EQUIPMENT PLACEMENT
CT-3 CT EQUIPMENT CLEARANCES
CT-4 CT RADIATION SCATTER PLOT
CT-5 CT EQUIPMENT DETAILS & NOTES
CT-6 CT EQUIPMENT DETAILS & NOTES
CT-7 CT RIGGING AND DELIVERY INFORMATION
PROJECT IDENTIFICATION INFORMATION
JOB NAME:INTEGRITY VETERINARY
WO NUMBER:W4849
PATTERSON NO:N/A
DRAFTING NO.2024-01
CUSTOMER INTEGRITY VETERINARY CENTER
518 PLEASANT ST
NORTHAMPTON, MA 01060
LAUREN PAYNE
LAUREN@INTEGRITYVETCENTER.COM
ARCHITECT:T.B.D
CONTRACTOR WAINSCOTT BUILDING
MATTHEW WAINSCOTT
37 STAGE ROAD
WILLIAMSBURG, MA 01096
413.559.0875
WAINSCOTTBUILDING@GMAIL.COM
EQUIPMENT PRE-OWNED
TOSHIBA AQUILION ONE CT
STANDARD TABLE
PROJECT MANAGER:BRIAN MALONEY
OFFICE:(888) 505-0319
CELL:(216) 633-3854
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
APRV CUSTOMER ACKNOWLEDGEMENT AND APPROVALNOTE TO CUSTOMER - FINAL EQUIPMENT PLANS:
1.THIS IS YOUR FINAL EQUIPMENT PLANNING PACKAGE, REVIEW THESE PLANS CAREFULLY, AND IF NECESSARY, MARK-UP WITH ANY DESIRED CHANGES WITHIN FIVE (5) BUSINESS DAYS.
2.RETURN A COPY OF THIS SHEET AND ANY SHEETS WITH MARK-UPS TO THE PROJECT MANAGER NOT LATER THAN FIVE (5) WORKING DAYS OF ISSUANCE TO CONFIRM YOUR RECEIPT AND ACKNOWLEDGEMENT OF THIS DRAWING PACKAGE. YOUR
APPROVAL AUTHORIZES THE STAGING OF YOUR EQUIPMENT AND MOVES YOUR EQUIPMENT ORDER STATUS FROM SALES & PLANNING TO STAGING & PRE-DELIVERY.
3.REVISIONS REQUIRED TO THE FINAL EQUIPMENT PLANS AFTER THE INITIAL ISSUANCE IS SUBJECT TO ADDITIONAL DRAFTING FEES.
PART 2 - FINAL EQUIPMENT PLAN APPROVAL AND ACKNOWLEDGEMENT
THIS IS YOUR PHASE II, "FINAL" DRAWING PACKAGE ISSUED FOR USE OF YOUR DESIGN PROFESSIONALS TO GENERATE CONSTRUCTION DRAWINGS. THE CUSTOMER MUST SIGN BELOW AND RETURN A COPY OF THIS SHEET WITH SIGNATURE, ALONG
WITH ANY COMMENTS, TO THE PRIZMED PROJECT MANAGER WITHIN FIVE (5) WORKING DAYS TO MOVE YOUR EQUIPMENT ORDER FROM THE SALES PHASE TO THE STAGING PHASE. IF A SIGNED APPROVAL IS NOT RECEIVED WITHIN THE SPECIFIED
TIME THIS PACKAGE WILL BE CONSIDERED "APPROVED WITHOUT COMMENT" AND THE CUSTOMER AGREES AS SET FORTH BELOW.
THE DRAWINGS WILL BE DEEMED TO BE ACKNOWLEDGED AND APPROVED BY THE CUSTOMER, AS IF SIGNED BY THE CUSTOMER, IF THIS SHEET IS NOT RETURNED AS STATED ABOVE.
HAVING REVIEWED THE FINAL DRAWINGS WE HEREBY ACKNOWLEDGE, CONFIRM AND ACCEPT THAT:
1.THE EQUIPMENT PLANNING REQUIREMENTS INDICATED HEREIN ARE SUBJECT TO CHANGE UNTIL THE ACTUAL EQUIPMENT FOR THE PROJECT HAS BEEN PHYSICALLY RETAINED AND INSPECTED AND WE UNDERSTAND THAT THE EQUIPMENT
SUPPLIER IS NOT RESPONSIBLE FOR ANY COSTS RESULTING FROM ADDED OR CHANGED DESIGN OR CONSTRUCTION REQUIREMENTS DUE TO THE FINAL EQUIPMENT CONFIGURATION.
2.THE CUSTOMER WILL RETAIN DESIGN PROFESSIONALS WHO SHALL BE SOLELY RESPONSIBLE FOR PREPARING CONSTRUCTION DRAWINGS AND INCORPORATING THE EQUIPMENT REQUIREMENTS INTO THE CONSTRUCTION DRAWINGS AND
THESE EQUIPMENT PLANS WILL NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES.
5.ALL INFORMATION HEREIN IS BASED ON INFORMATION OBTAINED FROM 3RD PARTY SOURCES AND IS INTENDED TO CONVEY TO THE CUSTOMER'S PROFESSIONAL DESIGN TEAM THE EQUIPMENT REQUIREMENTS AND NECESSARY SITE
PREPARATION. THE EQUIPMENT PROVIDER AND SITE PLANNER ARE NOT LICENSED DESIGN OR ENGINEERING PROFESSIONALS AND DO NOT PROVIDE ANY SUCH SERVICES. THE CUSTOMER AGREES TO INDEMNIFY AND HOLD HARMLESS THE
EQUIPMENT PROVIDER, SITE PLANNER AND ANYONE FOR WHOM THEY ARE LEGALLY LIABLE, FROM AND AGAINST ANY AND ALL DAMAGE, LOSSES OR COST, (INCLUDING ARCHITECT FEE'S, ENGINEERING FEE'S, ATTORNEYS' FEES AND DEFENSE
COSTS), CAUSED IN WHOLE OR IN PART BY ITS ACTS, ERRORS OR OMISSIONS IN PROVIDING THESE DRAWINGS AND THE INFORMATION CONTAINED HEREIN.
6.THE CUSTOMER SHALL AT HIS OWN EXPENSE WILL PERFORM ALL RECOMMENDED TESTING AND MONITORING AS IDENTIFIED HEREIN TO CONFIRM SITE SUITABILITY INCLUDING ANY NECESSARY SHIELDING DESIGNS, PERFORM ANY MITIGATION
REQUIRED AS A RESULT OF TESTING AND MONITORING AND PROVIDE THE RESULTS OF ANY SUCH ACTIONS TO THE EQUIPMENT SUPPLIER. SHOULD THE CUSTOMER ELECT NOT TO PERFORM ANY RECOMMENDED TESTING, MONITORING,
SHIELD DESIGNS OR MITIGATION ,THE CUSTOMER SHALL BE SOLELY RESPONSIBLE FOR ANY IMPACT ON THE INSTALLATION, SERVICE OR OPERATION OF THE EQUIPMENT, SAFETY OF PERSONS, EFFECTS TO OTHER EQUIPMENT AND
OPERATIONS.
7.THE CUSTOMER SHALL FURNISH TO THE EQUIPMENT SUPPLIER A COPY OF THE COMPLETED CONSTRUCTION DRAWINGS AND ANY RELEVANT TEST REPORTS AND KEEP THE EQUIPMENT SUPPLIER CURRENT ON ANY CHANGES THAT MAY AFFECT
THE EQUIPMENT OR IT'S INSTALLATION AND OPERATION.
8.THE CUSTOMER SHALL COORDINATE AS NECESSARY WITH THE EQUIPMENT SUPPLIER IN ALL ASPECTS TO ENSURE THE PROPER AND TIMELY PREPARATION OF THE SITE TO ACCEPT THE EQUIPMENT.
9.THE CUSTOMER SHALL SUPPLY TO THE EQUIPMENT SUPPLIER AS SOON AS AVAILABLE A COPY OF THE PROJECT SCHEDULE AND IDENTIFY THE EQUIPMENT DELIVERY DATE ON THE PROJECT SCHEDULE.
10.ALL EQUIPMENT BEING FURNISHED BY THE SUPPLIER IS "PRE-OWNED" UNLESS OTHERWISE SPECIFIED IN THE EXECUTED SALES AGREEMENT.
11.THE EQUIPMENT PROVIDER AND SITE PLANNER MAKE NO GUARANTEES AND SHALL NOT BE LIABLE FOR ANY IMPACT THE INSTALLATION OF THE EQUIPMENT MAY HAVE ON OTHER EQUIPMENT IN OR NEAR THE SPACE WHERE THIS EQUIPMENT
WILL BE INSTALLED AND THE CUSTOMER ACCEPTS ALL LIABILITIES AND RESPONSIBILITIES THEREOF.
12.WE ARE TENTATIVELY REQUESTING THE EQUIPMENT BE INSTALLED AND READY FOR OPERATION ON: ____________________________________________
THE CUSTOMER HEREBY ACKNOWLEDGES, CONFIRMS, AND ACCEPTS THIS DRAWING PACKAGE AND ALL INFORMATION AND REQUIREMENTS CONTAINED HEREIN AND WILL ENSURE THE SAME FOR THE DESIGN TEAM AND CONTRACTORS:
____________________________________________________________________________________________________________________________________________________________________________
(SIGNATURE OF AUTHORIZED INDIVIDUAL)(DATE SIGNED)(PRINTED NAME & TITLE)(COMPANY)
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
EDA EQUIPMENT DELIVERY AUTHORIZATIONEQUIPMENT DELIVERY AUTHORIZATION
IMPORTANT NOTE: THIS FORM MUST BE COMPLETED AND RETURNED TO YOUR EQUIPMENT PROVIDER'S PROJECT MANAGER NOT LESS THAN TEN (10) CALENDAR DAYS PRIOR TO THE SCHEDULED EQUIPMENT DELIVERY DATE, FAILURE TO PROVIDE
THIS SHEET MAY DELAY THE SCHEDULED DELIVERY OF YOUR EQUIPMENT!
THE CUSTOMER HEREBY CERTIFIES THAT THE SITE WILL BE READY FOR THE DELIVERY AND INSTALLATION OF THE EQUIPMENT IN ALL REGARDS AS SPECIFIED WITHIN THIS DRAWING PACKAGE ON _____________________________________________,
AT WHICH TIME THE INSTALLATION OF THE EQUIPMENT WILL IMMEDIATELY WILL COMMENCE. IF THE SITE IS NOT PROPERLY PREPARED UPON DELIVERY OF THE EQUIPMENT AND THE INSTALLATION IS DELAYED AS A RESULT, ADDITIONAL CHARGES
MAY BE INCURRED BY THE CUSTOMER, ALL CONSTRUCTION WORK MUST BE COMPLETE IN THE SPACE THE EQUIPMENT IS TO BE INSTALLED.
SITE READINESS CHECKLIST
1.THE UNLOADING AREA AND THE DELIVERY PATH ARE CLEAR OF ALL OBSTRUCTIONS THAT WOULD HINDER THE DELIVERY OR INSTALLATION AND THE PATH HAS ADEQUATE SPACE FOR THE MOVEMENT OF THE EQUIPMENT. IF THERE ARE ANY
RAMPS, STAIRS, ELEVATORS, DOCKS, GARAGES OR OTHER HINDRANCES ALONG THE DELIVERY PATH THE PROJECT MANAGER MUST BE NOTIFIED IN ADVANCE OF THE DELIVERY.
2.ALL SURFACES ARE TO BE COMPLETELY FINISHED, THIS INCLUDES ALL WALLS, CEILINGS, FLOORS, CABINETS, ETC., (WITH EXCEPTION OF THE DELIVERY PATH). IT IS HIGHLY RECOMMENDED THAT THE GENERAL CONTRACTOR USE HEAVY CRAFT
PAPER TO PROTECT ALL FINISHED FLOORING AND COUNTERTOPS PRIOR TO THE DELIVERY OF THE EQUIPMENT.
3.ALL SYSTEMS MUST BE COMPLETE AND OPERATIONAL, INCLUDING ALL HVAC SYSTEMS FOR 72 HRS, ELECTRICAL SYSTEMS , EPO CIRCUIT TESTED, LIGHTING, LIFE SAFETY SYSTEMS , AND LIVE PHONE AND DATA SYSTEMS.
4.THE SPACE THE EQUIPMENT IS BEING INSTALLED MUST BE DUST-FREE AND CLIMATE STABILIZED FOR A MINIMUM OF 72 HOURS PRIOR TO THE ARRIVAL OF THE EQUIPMENT.
5.THERE MUST BE ADEQUATE SPACE FOR THE STAGING OF EQUIPMENT IN THE AREA AND ALL CONTRACTOR TOOLS AND MATERIALS REMOVED FROM THE AREA.
6.A TRASH RECEPTACLE OR DUMPSTER IS AVAILABLE FOR DISPOSAL OF SHIPPING AND PACKING MATERIALS.
7.CEILING TILES MUST BE REMOVED IN AREAS WHERE ACCESS WILL BE REQUIRED FOR SYSTEM CABLING, THE CUSTOMER OR CONTRACTOR MUST HAVE SOMEONE ON SITE ON DELIVERY DAY FOR THIS PURPOSE. THE CUSTOMER OR
CONTRACTOR IS RESPONSIBLE FOR RE-INSTALLING SAME AFTER SYSTEM CABLES HAVE BEEN INSTALLED.
8.THE CUSTOMER'S ELECTRICIAN MUST BE ON SITE AND AVAILABLE THE AFTERNOON THE EQUIPMENT IS DELIVERED TO CONFIRM THE POWER SOURCE HAS THE PROPER RATING, GROUNDING, AND ROTATION AND TO ASSIST THE EQUIPMENT
INSTALLERS WITH TERMINATION OF THE MAIN POWER ONTO THE EQUIPMENT.
IF THERE ARE THERE ANY ITEMS THAT CAN NOT BE COMPLETED AS STATED ABOVE PRIOR TO EQUIPMENT DELIVERY, PLEASE EXPLAIN: ____________________________________________________________________________________________________
______________________________________________________________________________________________________________________________________________________________________________________________________________________________
EXCEPT AS MAY BE STATED ABOVE, ALL ITEMS WILL BE COMPLETED AND THE SITE READY FOR DELIVERY AND INSTALLATION ON THE DATE INDICATED ABOVE. WE ACKNOWLEDGE THAT ONCE THE EQUIPMENT HAS ARRIVED AT THE SITE, ANY ISSUES
OR DELAYS DUE TO NON-CONFORMING SITE CONDITIONS AND PREPARATION MAY IMPACT THE EQUIPMENT INSTALLATION AND MAY RESULT IN ADDITIONAL CHARGES TO THE CUSTOMER.
ACKNOWLEDGED BY: __________________________________________ ____________________________ _____________________________________________________________________________________________________
(CUSTOMER SIGNATURE)(DATE)(PRINT NAME)(COMPANY)
ACKNOWLEDGED BY: __________________________________________ ____________________________ _____________________________________________________________________________________________________
(CUSTOMER'S CONTRACTOR SIGNATURE)(DATE)(PRINT NAME)(COMPANY)
NETWORK CONNECTIVITY FOR MEDICAL EQUIPMENT: THE FOLLOWING INFORMATION IS REQUIRED FOR COMMUNICATION WITH YOUR DATA SYSTEMS , PROVIDE THE FOLLOWING INFORMATION:
MEDICAL EQUIPMENT
IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________
PACS OR IMAGE DESTINATION
IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________
WORKLIST
IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________
EMR OR MEDICAL RECORDS
IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________
OTHER: _________________________________________
IP ADDRESS: ______________________________AE TITLE: ______________________________SUBNET MASK: ______________________________DEFAULT GATEWAY: ______________________________PORT NUMBER: _________________
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
N-1 GENERAL NOTESDRAWING NOTES:
1.THESE DRAWINGS ARE FOR COORDINATION PURPOSES WITH THE CUSTOMER'S DESIGN TEAM AND ARE NOT TO BE
USED FOR PERMITTING OR CONSTRUCTION PURPOSES.
2.THESE DRAWINGS ARE PROVIDED TO LOCATE THE EQUIPMENT WITHIN THE SPACE, SUGGEST THE PLACEMENT
AND ROUTING OF RELATED SUPPORTING SYSTEMS, AND GUIDE THE CUSTOMER'S DESIGN TEAM TO INCORPORATE
THE EQUIPMENT REQUIREMENTS INTO THE CONSTRUCTION DOCUMENTS. IN PREPARING THESE DRAWINGS,
EVERY EFFORT HAS BEEN MADE TO CONFORM DETAILS TO THE ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED.
HOWEVER, BECAUSE THE INFORMATION CONTAINED HEREIN IS BEING PROVIDED IN ADVANCE OF THE RETENTION
OF THE ACTUAL EQUIPMENT BEING PROVIDED, THESE DRAWINGS ARE SUBJECT TO CHANGE. THE INFORMATION
CONTAINED HEREIN IS CORRECT TO THE BEST OF OUR KNOWLEDGE AT THE TIME THE INFORMATION WAS
PREPARED.
3.THE EQUIPMENT PROVIDER AND SITE PLANNER ARE NOT RESPONSIBLE FOR ANY DAMAGES RESULTING FROM THE
USE OF THESE DRAWINGS FOR ANY PURPOSES. THESE DRAWINGS ARE SCHEMATIC IN NATURE AND THE ROUTING
AND LOCATIONS OF ANY CONDUITS, TROUGHS, WIREWAYS, BREAKERS, DISCONNECT SWITCHES, ETC. ARE
SUGGESTIONS, ACTUAL LOCATIONS AND PLACEMENT ARE THE RESPONSIBILITY OF THE CUSTOMER'S DESIGN
PROFESSIONALS.
4.ALL EQUIPMENT LAYOUTS ARE BASED ON THE RECOMMENDED SPACE NECESSARY FOR THE OPERATION AND
SERVICEABILITY OF THE EQUIPMENT BEING PROPOSED AND ARE IN THE BEST INTEREST OF BOTH THE CUSTOMER
AND EQUIPMENT SUPPLIER BASED ON ARCHITECTURAL DRAWINGS OR INFORMATION PROVIDED BY THE
CUSTOMER. IF CHANGES ARE MADE TO THE LAYOUT OR DIMENSIONS THE SITE PLANNER IS TO BE NOTIFIED
IMMEDIATELY AS CHANGES COULD RESULT IN CONFLICTS WITH THE EQUIPMENT CLEARANCES, PLACEMENT OF
THE EQUIPMENT, AND POSSIBLY IMPACT THE CLINICAL USE OF THE EQUIPMENT. IF CHANGES ARE MADE TO THE
FLOOR PLAN AFTER THE SUBMISSION OF THE INITIAL VERSION OF THESE PLANS ADDITIONAL DRAFTING CHARGES
MAY BE INCURRED BY THE CUSTOMER.
5.ALL INFORMATION HEREIN IS BASED ON INFORMATION OBTAINED FROM 3RD PARTY SOURCES AND IS INTENDED
TO CONVEY TO THE CUSTOMER'S PROFESSIONAL DESIGN TEAM THE EQUIPMENT REQUIREMENTS AND
NECESSARY SITE PREPARATION. THE EQUIPMENT PROVIDER AND SITE PLANNER ARE NOT LICENSED DESIGN OR
ENGINEERING PROFESSIONALS AND DO NOT PROVIDE ANY SUCH SERVICES. THE CUSTOMER AGREES TO
INDEMNIFY AND HOLD HARMLESS THE EQUIPMENT PROVIDER, SITE PLANNER AND ANYONE FOR WHOM THEY ARE
LEGALLY LIABLE, FROM AND AGAINST ANY AND ALL DAMAGES, LOSSES OR COSTS, (INCLUDING ARCHITECT FEES,
ENGINEERING FEES, ATTORNEYS' FEES AND DEFENSE COSTS), CAUSED IN WHOLE OR IN PART BY ITS ACTS,
ERRORS OR OMISSIONS IN PROVIDING THESE DRAWINGS AND THE INFORMATION CONTAINED HEREIN. .
6.THE CUSTOMER'S ARCHITECT AND GENERAL CONTRACTOR SHALL BE ULTIMATELY RESPONSIBLE FOR
COMPLIANCE WITH ALL APPLICABLE CODES AND PROFESSIONAL DESIGN REQUIREMENTS. CONSTRUCTION
DRAWINGS ARE TO BE SUBMITTED TO THE EQUIPMENT SUPPLIER FOR REVIEW PRIOR TO CONSTRUCTION.
7.ALL DIMENSIONS SHOWN ARE TAKEN FROM FINISHED SURFACES UNLESS SPECIFIED OTHERWISE.
8.THIS DRAWING PACKAGE DOES NOT PROVIDE SHIELDING DESIGNS, IF ANY SHIELDING IS REQUIRED IT SHALL BE
THE RESPONSIBILITY OF THE CUSTOMER IN ALL REGARDS.
9.OWNERSHIP OF DOCUMENTS, DRAWINGS, SPECIFICATIONS AND OTHER INFORMATION, INCLUDING THOSE IN
ELECTRONIC FORM, PREPARED BY THE EQUIPMENT PROVIDER AND THE EQUIPMENT PROVIDER'S CONSULTANTS
ARE INSTRUMENTS OF SERVICE FOR USE SOLELY WITH RESPECT TO THIS PROJECT. THE EQUIPMENT PROVIDER
AND THE EQUIPMENT PROVIDER'S CONSULTANTS SHALL BE DEEMED THE AUTHORS AND OWNERS OF THEIR
RESPECTIVE INSTRUMENTS OF SERVICE AND SHALL RETAIN ALL COMMON LAW, STATUTORY AND OTHER
RESERVED RIGHTS, INCLUDING COPYRIGHTS.
SITE USAGE, DELIVERY AND INSTALLATION
1.THE MINIMUM STANDARD DELIVERY REQUIREMENTS FOR THE EQUIPMENT ARE INDICATED HEREIN, SPECIAL
ATTENTION SHOULD BE PAID TO:
1.1.THE ENTIRE DELIVERY PATH MUST BEING LARGE ENOUGH TO ACCOMMODATE ALL SYSTEM COMPONENTS
AND BE FREE AND CLEAR OF ALL OBSTRUCTIONS.
1.2.THE MINIMUM CEILING HEIGHTS AS NOTED IN THESE PLANS OR AS STATED IN THE O.E.M. REFERENCE
DRAWINGS FOR THE SCAN AND EQUIPMENT ROOMS.
1.3.ALL SURFACES AND FLOORS ALONG THE ENTIRE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING THE
WEIGHT OF ALL SYSTEM COMPONENTS.
2.FOR STANDARD RIGGING AND DELIVERY, ANY OPENINGS, DOORWAYS, OR TURNS IN THE DELIVERY PATH MUST BE
ABLE TO ACCOMMODATE THE GANTRY AND TABLE AS AN ASSEMBLY. SEE DETAILS HEREIN.
3.ADEQUATE SPACE FOR STAGING THE EQUIPMENT DURING INSTALLATION. THE STORAGE SPACE MUST BE INSIDE
THE BUILDING IN CLOSE PROXIMITY TO THE WORK AREA, BE CLIMATE CONTROLLED AND ABLE TO BE SECURED
AFTER HOURS.
4.THE CUSTOMER SHALL PROVIDE AT NO COST TO THE EQUIPMENT SUPPLIER A CONTAINER, (DUMPSTER, ROLL-OFF,
ETC.), FOR THE DISPOSAL OF ALL PACKAGING AND SHIPPING DEBRIS, GENERALLY CONSISTING OF PALLETS,
CRATING, BOXES AND PROTECTIVE PLASTIC WRAP.
5.THE ANTICIPATED TIME FRAME FOR INSTALLATION OF THE MEDICAL EQUIPMENT IS TYPICALLY:
THREE (3) WORKING DAYS FOR CT SYSTEMS
6.THE INSTALLATION CREW RESERVES THE RIGHT TO HAVE ACCESS TO THE SITE AND PERFORM WORK OUTSIDE OF
NORMAL WORKING HOURS, (7:30AM - 5:00PM), WITH ADVANCED NOTICE TO THE CUSTOMER.
7.IF NECESSARY, THE CUSTOMER IS RESPONSIBLE FOR THE REMOVAL AND REPLACEMENT OF CEILING TILES TO
FACILITATE THE CABLING AND/OR INSTALLATION OF THE EQUIPMENT.
CUSTOMER'S RESPONSIBILITIES
1.THE CUSTOMER IS SOLELY RESPONSIBLE FOR RETAINING A PROFESSIONAL DESIGN TEAM TO PREPARE PERMIT
AND CONSTRUCTION DRAWINGS, ENSURING COMPLIANCE WITH ALL EQUIPMENT REQUIREMENTS, CODES,
REGULATORY REQUIREMENTS AND SECURING ALL APPROVALS, PERMITS AND LICENSES.
2.ALL WORK EXCEPT THE INSTALLATION OF THE EQUIPMENT IS THE RESPONSIBILITY OF THE CUSTOMER WITHOUT
EXCEPTION AND THE EQUIPMENT PROVIDER ASSUMES NO RESPONSIBILITY FOR ANY CONSTRUCTION COSTS,
WHETHER OR NOT RELATED TO THE INSTALLATION OF THE EQUIPMENT OR THE USE OF THESE DRAWINGS AND THE
INFORMATION CONTAINED HEREIN.
3.THE GENERAL CONTRACTOR SHALL BE RESPONSIBLE FOR THE REINSTALLATION OF CEILING TILES, TOUCH-UP AND
ANY COSMETIC WORK AFTER THE INSTALLATION OF THE EQUIPMENT.
4.THE CUSTOMER SHALL BE RESPONSIBLE FOR ALL SAFETY SIGNAGE REQUIRED BY ACR, ISO, JCAHO, OSHA, ETC.,
INCLUDING SAFETY AND WARNING SIGNAGE FOR THE CRYOGEN EXHAUST SYSTEM.
5.THE CUSTOMER SHALL PERFORM AN INSPECTION AT LEAST FIVE (5) WORKING DAYS PRIOR TO DELIVERY OF THE
EQUIPMENT AND VERIFY IN WRITING TO THE EQUIPMENT PROVIDER USING SHEET "PDCL" OF THESE DRAWINGS
THAT THE SITE IS READY FOR DELIVERY AND INSTALLATION AS SCHEDULED. UPON DELIVERY OF THE EQUIPMENT,
SHOULD THE SITE NOT BE READY AND CAUSE DELAY IN THE INSTALLATION, THE CUSTOMER SHALL REIMBURSE THE
EQUIPMENT PROVIDER FOR ANY ASSOCIATED COSTS INCLUDING COSTS FOR ADDITIONAL STORAGE, WAGES,
TRAVEL, LODGING, OR RE-MOBILIZATION COSTS.
6.EQUIPMENT WARRANTIES, EXPRESSED OR IMPLIED, ON THE PART OF EQUIPMENT SUPPLIER SHALL BE CONTINGENT
UPON STRICT COMPLIANCE WITH ALL RECOMMENDATIONS AND REQUIREMENTS CONTAINED IN THESE DRAWINGS.
7.THE EQUIPMENT SUPPLIER SHALL BE RESPONSIBLE FOR THE EQUIPMENT INSTALLATION AND CALIBRATION WHICH
INCLUDES INSTALLING SYSTEM CABLES AND TERMINATING CONTRACTOR PROVIDED CABLES TO THE EQUIPMENT. IN
THE EVENT THAT SPECIFIC TRADE RULES OR LICENSE REQUIREMENTS PROHIBIT THIS, THE CUSTOMER SHALL
INITIATE THE SERVICES OF APPROVED OTHER CONTRACTORS AND PAY FOR SELECTED, APPROVED PARTIES TO
PERFORM THIS WORK WITH JOB SUPERVISION TO BE PROVIDED BY THE EQUIPMENT PROVIDER. CALIBRATION
WHEN ACCOMPLISHED OUTSIDE OF NORMAL INSTALLATION SEQUENCES DUE TO CONTRACTOR OR TRADE RULE
ACTIONS OR REQUIREMENTS SHALL BE SUPPORTED BY, CHARGED TO, AND ACCEPTED BY THE CUSTOMER AS AN
ADDITIONAL INSTALLATION EXPENSE.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
N-2 GENERAL NOTESRECOMMENDED TESTING AND ENVIRONMENTAL EVALUATION
THE CUSTOMER IS RESPONSIBLE FOR HAVING A COMPLETE ENVIRONMENTAL SURVEY PERFORMED OF THE PROPOSED
SITE TO ENSURE ALL ENVIRONMENTAL CONDITIONS WILL MEET THE REQUIREMENTS OF THE ORIGINAL MANUFACTURER
OF THE PROPOSED EQUIPMENT.
RECOMMENDED ENVIRONMENTAL TESTING INCLUDES THE FOLLOWING:
1.FOR ALL MEDICAL IMAGING EQUIPMENT THE PRIMARY ELECTRICAL SERVICE MUST BE THOROUGHLY EVALUATED
TO INCLUDE A LOAD STUDY ON THE ELECTRICAL SERVICE INTENDED FOR THE IMAGING EQUIPMENT AND
MONITORING TO ENSURE THE POWER QUALITY MEETS THE SPECIFICATIONS INCLUDING TRANSIENT VOLTAGES
AND GROUNDING. THE CUSTOMER IS RESPONSIBLE FOR VERIFICATION THAT PRIOR TO THE SYSTEM BEING
INSTALLED THE REQUIREMENTS ARE MET. IN THE EVENT THERE ARE ANY ENVIRONMENTAL OR POWER
DEFICIENCIES AT THE SITE, THE FACILITY WILL BE REQUIRED TO CORRECT SUCH DEFICIENCIES PRIOR TO
ENERGIZATION OF THE EQUIPMENT. PLEASE BE ADVISED THAT ANY QUALITY ISSUES OR DAMAGE TO THE
IMAGING SYSTEM AS A RESULT OF ENVIRONMENTAL OR POWER ISSUES WILL FALL OUTSIDE ANY WARRANTY OR
SERVICE AGREEMENT COVERAGE. A RECOMMENDED RESOURCE FOR THE MONITORING IS: RX MONITORING
SERVICES, (888) 329-2321 http://www.rxms.com/
2.ALL MEDICAL IMAGING EQUIPMENT IS VERY SENSITIVE TO VIBRATION. THE EQUIPMENT MUST BE LOCATED AS FAR
AS POSSIBLE FROM VIBRATION SOURCES SUCH AS ADJACENT ROADWAYS, PARKING LOTS, SUBWAYS, TRAIN
TRACKS, HALLWAYS, ELEVATORS, WEIGHT MACHINES, ETC. IF ANY ASPECT OF THE SITE IS IN QUESTION, THE
CUSTOMER IS REQUIRED TO PERFORM A VIBRATION STUDY.
CEILING AND STRUCTURAL SUPPORTS
1.ALL STRUCTURAL SUPPORTS ARE THE RESPONSIBILITY OF THE CUSTOMER AND SHALL BE DESIGNED AND
INSTALLED BY THE CUSTOMER'S DESIGN TEAM AND CONTRACTORS. THIS INCLUDES LOAD BEARING SLABS,
ELEVATED SLABS, CEILING SUPPORT SYSTEMS, SHORING, BRACING, ETC., WITHOUT EXCEPTION. ALL SUPPORT
SYSTEMS MUST BE COMPLETELY INSTALLED AND READY PRIOR TO THE DELIVERY OF EQUIPMENT.
FLOOR LEVELNESS
1.THE FLOOR MUST BE LEVEL TO WITHIN 1/8" IN 10'-0" WHERE THE EQUIPMENT IS PLACED, THE EQUIPMENT CAN NOT
BE SHIMMED.
2.THE CONTRACTOR MUST VERIFY IN WRITING NOT LESS THAN 5 DAYS BEFORE DELIVERY THAT THE FLOOR MEETS
THIS REQUIREMENT.
EQUIPMENT FURNISHED AND INSTALLED BY CUSTOMER'S CONTRACTORS
1.SPECIAL CARE AND CONSIDERATION MUST BE GIVEN TO ITEMS THAT ARE TO BE FURNISHED AND INSTALLED BY
THE CUSTOMER'S CONTRACTORS THAT ARE NOT INCLUDED WITH THE MEDICAL EQUIPMENT. EVERY ATTEMPT HAS
BEEN MADE TO IDENTIFY THESE ITEMS WHICH MAY INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING:
·MAIN DISCONNECT CABINETS AND BREAKERS
·CONDUITS, TROUGHS, BREAKERS, WIRING
·LIGHTING, CONTROLS AND REQUIRED WARNING LIGHTS OR CONTROLLERS
·RADIATION SHIELDING SYSTEMS AND DESIGNS
LAYOUT PRIORITY ITEMS
1.EQUIPMENT ISO-CENTER IS THE CRITICAL POINT FOR ALL LAYOUT WORK IN THE SCAN ROOM. ANY DEVIATION
MUST BE REPORTED TO THE SITE PLANNER IMMEDIATELY. THE GENERAL CONTRACTOR SHALL MAINTAIN MARKS
FOR ISO-CENTER THROUGHOUT THE PROJECT.
NETWORKING AND NETWORK INFORMATION
1)THE CUSTOMER MUST PROVIDE TO THE EQUIPMENT PROJECT MANAGER NOT LESS THAN 1 WEEK IN ADVANCE OF
THE DELIVERY OF THE EQUIPMENT ALL NETWORKING AND IT INFORMATION AS MAY BE REQUIRED TO CONFIGURE
THE MEDICAL EQUIPMENT TO BE COMPATIBLE WITH THE CUSTOMER'S SYSTEMS SUCH AS THE EMR, PACS,
WORKLIST, REMOTE SEND DESTINATIONS, ETC.
2)CUSTOMER ACKNOWLEDGES THAT DURING THE INSTALLATION OF THE EQUIPMENT THE INSTALLATION TEAM WILL
PROGRAM AND TEST ALL NETWORKING. SHOULD NETWORK TESTS FAIL DUE TO EXTERNAL ISSUES THE
CUSTOMER WILL RESOLVE ALL SUCH ISSUES WHILE THE INSTALLATION TEAM IS ON SITE. IF FOR ANY EXTERNAL
REASON AN ENGINEER MUST RETURN TO THE SITE TO RESOLVE EXTERNAL NETWORKING ISSUES THE CUSTOMER
WILL BE BILLED ACCORDINGLY.
REQUIRED INFORMATION FROM CUSTOMER
FOR CT PROJECTS, THE FOLLOWING INFORMATION IS TO BE PROVIDED BY THE CUSTOMER TO THE EQUIPMENT
SUPPLIER:
1.A COPY OF THE PROJECT SCHEDULE INDICATING AN EQUIPMENT DELIVERY DATE.
2.A COPY OF THE COMPLETED CONSTRUCTION DRAWINGS.
3.A COPY OF ANY SHIELDING DRAWINGS AND CALCULATIONS.
4.COPIES OF ANY CORRESPONDENCE AND RFI'S THAT PERTAIN TO THE IMAGING EQUIPMENT THROUGHOUT THE
PROJECT.
5.THE COMPLETED EQUIPMENT DELIVERY AUTHORIZATION FORM FROM THIS SET OF DRAWINGS.
SITE READINESS AND EQUIPMENT DELIVERY
THE CUSTOMER'S CONTRACTOR MUST SUBMIT AS SOON AS IT IS AVAILABLE, A PROJECT SCHEDULE TO THE EQUIPMENT
PROVIDER. THE PROJECT SCHEDULE MUST SPECIFICALLY IDENTIFY THE DESIRED DELIVERY DATE FOR THE EQUIPMENT.
IF THE DATE CHANGES, THE EQUIPMENT PROVIDER MUST BE NOTIFIED SO ADJUSTMENTS CAN BE MADE TO THE
DELIVERY AND INSTALLATION. SCHEDULE. A LETTER SHALL BE PROVIDED TO THE CUSTOMER AT AN APPROPRIATE TIME
TO CONFIRM THE SITE READINESS AND DELIVERY DATE.
IN GENERAL, SITE READINESS TO ACCEPT DELIVERY OF THE EQUIPMENT INCLUDES, BUT IS NOT LIMITED TO, THE
FOLLOWING:
1.ALL CONSTRUCTION WORK MUST BE COMPLETE IN THE SPACE THE EQUIPMENT IS TO BE INSTALLED.
2.ALL SURFACES ARE TO BE COMPLETELY FINISHED, THIS INCLUDES ALL WALLS, CEILINGS, FLOORS, CABINETS,
ETC., (WITH EXCEPTION OF THE DELIVERY PATH). THE CUSTOMER'S CONTRACTORS ARE RESPONSIBLE FOR
PROTECTING FINISHED SURFACES AS REQUIRED, (CRAFT PAPER ON FLOORS AND COUNTERTOPS ARE
SUGGESTED).
3.ALL SYSTEMS MUST BE COMPLETE AND OPERATIONAL, INCLUDING ALL HVAC SYSTEMS, ELECTRICAL SYSTEMS ,
LIGHTING, LIFE SAFETY, PHONE AND DATA NETWORKS.
4.THE ENVIRONMENT MUST BE DUST-FREE AND CLIMATE STABILIZED FOR A MINIMUM OF 72 HOURS PRIOR TO THE
ARRIVAL OF THE EQUIPMENT.
5.THERE MUST BE ADEQUATE SPACE FOR THE STAGING OF EQUIPMENT IN THE AREA AND ALL CONTRACTOR TOOLS
AND MATERIALS REMOVED FROM THE AREA.
6.A TRASH RECEPTACLE OR DUMPSTER IS AVAILABLE FOR DISPOSAL OF SHIPPING AND PACKING MATERIALS.
7.WHERE NECESSARY, CEILING TILES MUST BE REMOVED TO FACILITATE THE INSTALLATION OF SYSTEM CABLING
AS DIRECTED BY THE INSTALLATION TEAM, IT IS THE CUSTOMER'S CONTRACTORS RESPONSIBILITY TO RE-INSTALL
SAME AFTER SYSTEM CABLES HAVE BEEN INSTALLED.
8.THE UNLOADING AREA AND DELIVERY PATH ARE CLEAR OF ALL OBSTRUCTIONS THAT WOULD HINDER THE
DELIVERY OR INSTALLATION AND THE PATH HAS ADEQUATE SPACE FOR THE MOVEMENT OF THE EQUIPMENT. IF
THERE ARE ANY RAMPS, STAIRS, ELEVATORS, DOCKS, GARAGES OR OTHER HINDRANCES ALONG THE DELIVERY
PATH THE PROJECT MANAGER MUST BE NOTIFIED IN ADVANCE OF THE DELIVERY.
9.THE CUSTOMER'S ELECTRICIAN MUST BE ON SITE AND AVAILABLE THE AFTERNOON EQUIPMENT IS DELIVERED TO
CONFIRM THE POWER SOURCE HAS THE PROPER RATING, GROUNDING AND ROTATION AND TO ASSIST THE
EQUIPMENT INSTALLERS WITH TERMINATION OF THE MAIN POWER ONTO THE EQUIPMENT.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
A-1 FLOOR PLANNorth
19'-101
2"
14'-7"
5"
7'
11'-51
2"
16'-2"
51
2"
6'-61
2"
3'-11
2"
3'-81
2"
4'
4'
EXISTING
ELECT.
TRANS.
EXIST
ELECT
PANEL4'
VERIFY
8'
VERIFY
REV1
DIMENSIONS REVISED DUE TO GANTRY WIDTH
REV1
VERIFY EXACT LOCATION OF
TRANSFORMER AND PANEL
TO REMAIN.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
A-2 CT RIGGING PATHNorth
RIGGING AND DELIVERY NOTES
1.REFER TO SECTION 7 OF THE SITE PLANNING GUIDE FOR CLEARANCE REQUIREMENTS.
2.THE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING 6,500 LBS SPREAD OVER FOUR 6" DIAMETER RUBBER
WHEELS.
3.MINIMUM OPENINGS 3'-7" CLEAR WIDTH X 81" CLEAR HEIGHT
4.DEPICTED IS THE PRESUMED RIGGING PATH FOR THE EQUIPMENT. THE CUSTOMER MUST ENSURE THE ENTIRE
EQUIPMENT DELIVERY PATH IS CLEAR AND ALL HALLWAYS, CORRIDORS, OPENINGS AND POINTS THE
EQUIPMENT WILL HAVE TO BE TURNED ARE LARGE ENOUGH TO ACCOMMODATE THE EQUIPMENT, SEE SHEET
A-3 FOR TURNING REQUIREMENTS.
INADEQUATE TURN AREA
FOR GANTRY AND TABLE TO COME IN THROUGH
THIS DOOR, IT WOULD BE IMPOSSIBLE TO TURN
THE GANTRY INTO THE SCAN ROOM WITH THE
CONTROL ROOM WALL BUILT.
3'-3"
MIN
GANTRY
8'
3'-7"
7'-9"
MIN GANTRY
3'-3"
MIN
GANTRY
8'
3'-7"
7'-9"
MIN GANTRY
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
A-3 RIGGING INFORMATIONNorth
5'-6"
5'-6"
CLEAR OPENING
8'
6'
CLEAR OPENING
5'-6"
5'
CLEAR OPENING
8'
GANTRY DELIVERY OPENING REQUIREMENTS TABLE DELIVERY OPENING REQUIREMENTS
5'-6"
CLEAR OPENING
5'-6"
8'
6'
CLEAR OPENING
5'-6"
5'
CLEAR OPENING
8'
GANTRY DELIVERY OPENING REQUIREMENTS TABLE DELIVERY OPENING REQUIREMENTS
3'-3"
MIN
GANTRY
8'
3'-7"
7'-9"
MIN GANTRY
2'-2"
8'
CBTB-021A TABLE
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-1 CT EQUIPMENT LEGENDCEILING HEIGHT
·RECOMMENDED CEILING HEIGHT: 9'-0"
·MINIMUM CEILING HEIGHT: 8'-2 1/2"
VIBRATION REQUIREMENTS FOR AQUILION GANTRY
.98 M/S² (0.1 G) OR LESS
IMPORTANT NOTES
1.THESE EQUIPMENT PLANS AND THE SYSTEM
COMPONENTS MUST BE VERIFIED WHEN THE ACTUAL
SYSTEM FOR THIS PROJECT HAS BEEN RETAINED,
INSPECTED AND STORED. THERE ARE SYSTEM OPTIONS
AND VARIANCES THAT CAN IMPACT SITE PREP, SPACE
REQUIREMENTS, PERFORMANCE, AND SAFETY.
2.IT IS THE RESPONSIBILITY OF THE CUSTOMER TO RETAIN
A REGISTERED PHYSICIST TO PREPARE A SHIELDING
PRESCRIPTION AND DETERMINE THE EXTENT AND
AMOUNTS OF RADIATION SHIELDING REQUIRED. THIS
PACKAGE SHOULD BE PROVIDED TO THE MEDICAL
PHYSICIST ALONG WITH WORKFLOW, VOLUME AND
ADJACENT OCCUPANCY DATA.
CEILING HEIGHT
·RECOMMENDED CEILING HEIGHT: 9'-0"
·MINIMUM CEILING HEIGHT: 8'-2 1/2"
SYSTEM ELECTRICAL REQUIREMENTS FOR AQUILION WITH "PCDU"
SUPPLY CONFIGURATION:3 PHASE, 3-WIRE
DELTA OR WYE CONNECTED
STANDARD TRANSFORMER CAPACITY:150 KVA
MAXIMUM SYSTEM DEMAND (IMAGING):125 KVA
SUPPLY VOLTAGE:480V, 60HZ, 150 AMP
*TOSHIBA EQUIPMENT BTU / HOUR:
(INCLUDING GANTRY & COUCH ONLY)
EXAM ROOM STAND-BY:14,672 BTU/HR
EXAM ROOM SCANNING 2 PATIENTS:20,712 BTU/HR
EXAM ROOM SCANNING 3 PATIENTS:21,991 BTU/HR
EXAM ROOM SCANNING 4 PATIENTS:23,271 BTU/HR
EXAM ROOM SCANNING 5 PATIENTS:24,550 BTU/HR
EXAM ROOM SCANNING MAXIMUM:25,830 BTU/HR
NOTE: FUTURE GROWTH OF FACILITY MUST BE CONSIDERED
WHEN FORECASTING PATIENT NUMBERS FOR A/C
REQUIREMENTS.
VIBRATION REQUIREMENTS FOR AQUILION GANTRY
.98 M/S² (0.1 G) OR LESS
IMPORTANT NOTES
1.THESE EQUIPMENT PLANS AND THE SYSTEM
COMPONENTS MUST BE VERIFIED WHEN THE ACTUAL
SYSTEM FOR THIS PROJECT HAS BEEN RETAINED,
INSPECTED AND STORED. THERE ARE SYSTEM OPTIONS
AND VARIANCES THAT CAN IMPACT SITE PREP, SPACE
REQUIREMENTS, PERFORMANCE, AND SAFETY.
2.IT IS THE RESPONSIBILITY OF THE CUSTOMER TO RETAIN
A REGISTERED PHYSICIST TO PREPARE A SHIELDING
PRESCRIPTION AND DETERMINE THE EXTENT AND
AMOUNTS OF RADIATION SHIELDING REQUIRED. THIS
PACKAGE SHOULD BE PROVIDED TO THE MEDICAL
PHYSICIST ALONG WITH WORKFLOW, VOLUME AND
ADJACENT OCCUPANCY DATA.
CASEWORK AND ACCESSORIES
1)ALL CASEWORK IS TO BE DESIGNED, DETAILED,
FURNISHED, AND INSTALLED BY THE CUSTOMER
AND/OR CONTRACTOR..
2)ALL FURNITURE, (OPERATOR CHAIRS), SIGNAGE,
FURNISHINGS, AND ACCESSORIES ARE PROVIDED BY
THE CUSTOMER.
3)THE COUNTER TOP AT THE OPERATORS STATION MUST
BE INSTALLED PRIOR TO DELIVERY OF THE EQUIPMENT.
ELECTRICAL
1)ANY SOUND OR INTERCOM SYSTEMS ARE TO BE
FURNISHED AND INSTALLED BY THE CUSTOMER.
2)EMERGENCY POWER OFF (EPO) BUTTONS ARE
REQUIRED IN THE CONTROL ROOM AND SCAN ROOM,
LOCATIONS TO BE DETERMINED BY THE CUSTOMER'S
ELECTRICAL ENGINEER. CUSTOMER'S ELECTRICAL
ENGINEER IS RESPONSIBLE FOR DESIGN OF EPO
CIRCUIT.
3)DOOR SAFETY SWITCHES AND IN USE SIGNAGE IS
REQUIRED ON ALL DOORS ACCESSING THE
EXAMINATION ROOM.
4)IT IS HIGHLY SUGGESTED THE CUSTOMER PERFORM A
POWER SURVEY PRIOR TO DESIGN AND CONSTRUCTION
TO ENSURE THE INCOMING POWER WILL MEET THE
EQUIPMENT SPECIFICATIONS AND REQUIREMENTS.
DAMAGE TO EQUIPMENT DUE TO IMPROPER POWER IS
NOT COVERED BY ANY SERVICE AGREEMENT OR
WARRANTY.
AQUILION ONE MINIMUM ROOM SIZE REQUIREMENTS
SCAN ROOM CONTROL ROOM
13'-2"W X 19'-10"L 6'-0"W X 8'-0"L
TOSHIBA AQUILION ONE CT EQUIPMENT LEGEND
FURNISHED AS PART OF CT SYSTEM, INSTALLED BY PROVIDER
TAG ELEC ITEM DESCRIPTION BTU/HR WEIGHT
SYMB (LBS)
1 GANT AQUILION CT GANTRY SEE NOTE 6,063
2 PCH CT PATIENT COUCH SEE NOTE 1,941
3 CPU CENTRAL PROCESSING UNIT (HOST PC)5,123 276
4 MON MONITOR FOR CPU 192 18
5 SPK1 SPEAKER - OPERATOR ---4
6 SPK2 SPEAKER - SCAN ROOM ---4
7 SKBD SCAN CONSOLE KEYBOARD ---6
8 DKBD SCAN CONSOLE KEYBOARD ---6
9 30" DEEP COUNTER BY CUSTOMER
10 CHAIR BY CUSTOMER
11 INV INVERSION WORKSTATION
12 REC1 RECONSTRUCTION UNIT 13,020 816
13 REC2 RECONSTRUCTION UNIT 13,020 816
14 PR PHANTOM RACK
15 PDU POWER DISTRIBUTION UNIT 4,777 1,654
ITEMS FURNISHED AND INSTALLED BY CUSTOMER'S CONTRACTORS
CB MAIN CIRCUIT BREAKER FOR CT SYSTEM ONLY. SHUNT TRIP TYPE
BREAKER WITH REMOTE POWER OFF STATIONS (EPO). SHOWN IN
SUGGESTED LOCATION, ACTUAL LOCATION BY CUSTOMER'S
ELECTRICAL ENGINEER.
EPO EMERGENCY POWER OFF STATION. SHOWN IN SUGGESTED LOCATION.
ACTUAL LOCATION AND DESIGN OF EPO CIRCUITS DETERMINED BY
CUSTOMER'S ELECTRICAL ENGINEER.
NOTE: FOR HEAT OUTPUT OF SYSTEM IN SCAN ROOM, SEE CHART THIS SHEET FOR
BTU/HOUR
ENVIRONMENTAL REQUIREMENTS
1.MAINTAIN AT ALL TIMES 24/7/365 AMBIENT TEMPERATURE BETWEEN 68F TO 74F, HUMIDITY RANGE OF 40% TO 70% NON CONDENSING. STATIC AMBIENT TEMPERATURE IS TO BE
PROVIDED AND MAINTAINED AS SPECIFIED. ALL CALCULATIONS ARE TO UTILIZE TOSHIBA PROVIDED HEAT OUTPUT SPECIFICATIONS OF EQUIPMENT, (SEE CHART ABOVE).
2.IN GENERAL THE SCAN ROOM MUST BE PROVIDED WITH AN INDEPENDENT AIR CONDITIONING SYSTEM.IF THE ROOM IS MAINTAINED WITHIN THE PERMISSIBLE TEMPERATURE
RANGE, GRADUAL TEMPERATURE SHIFTS MAY AFFECT SYSTEM PERFORMANCE, THEREFORE THE ROOM MUST BE KEPT UNDER CONSTANT CONTROL, WITHIN ±3%.
3.A MINIMUM OF 10 AIR CHANGES PER HOUR IS SUGGESTED. CONSULT LOCAL CODE.
4.EQUIPMENT IN ENCLOSED SPACES SUCH AS EQUIPMENT ROOMS, CLOSETS AND COMPUTER ROOMS MUST BE PROVIDED WITH ADEQUATE VENTILATION.
5.A/C SUPPLY OUTLET TO BE PROVIDED BY CUSTOMER AT FLOOR LEVEL OR AS CLOSE AS POSSIBLE TO CONTROL ROOM DESK.
6.DUE TO HEAT GENERATED BY THE "CPU" AND "REC" UNITS, ADDITIONAL VENTILATION IN THE CONTROL AREA IS REQUIRED. CUSTOMER OR CONTRACTOR PROVIDED FANS
MAY BE NECESSARY BELOW THE DESKTOP FOR TECHNICIAN COMFORT. THE "CPU/REC" UNITS SHOULD NOT BE BE ENCLOSED IN CASEWORK.
7.AIR SUPPLY DUCTS SHOULD NOT BE PLACED DIRECTLY OVER EXAMINATION TABLES OR GANTRY FOR PATIENT COMFORT CONSIDERATIONS.
8.THE AIRFLOW THROUGH TOSHIBA EQUIPMENT CABINETS IS FROM BOTTOM TO TOP.
9.WHERE POSSIBLE A/C SUPPLY OUTLETS SHOULD BE LOCATED AT FLOOR LEVEL. NO A/C OUTLET SHOULD BE WITHIN THE EXCLUSION ZONE SHOWN BELOW AND AT NO TIME
SHOULD THE CT SYSTEM BE EXPOSED TO DIRECT AIRFLOW.
NETWORK REQUIREMENTS
DUE TO THE LARGE DATA SETS GENERATED BY THE HIGH
SPEED SCANS AN OPTIMIZED NETWORK IS REQUIRED TO
MAINTAIN FAST IMAGE TRANSFER SPEEDS. THE TWO
REQUIREMENTS ARE:
1)HIGH SPEED NETWORK, GIGABIT ETHERNET BETWEEN
THE AQUILION AND ITS NETWORK PEERS.
2)AN UNCLUTTERED NETWORK ENVIRONMENT. THE
AQUILION'S IMMEDIATE NETWORK, INCLUDING
CONNECTION TO THE VITAL WORKSTATION AND TEH
MCKESSON THIN SLICE CACHE SHOULD BE ON AN
ISOLATED NETWORK SEGMENT FREE OF BROADCASTS
AND EXTRANEOUS SIGNALS. THIS CAN BE ACHIEVED BY
VLAN, FIREWALL, OR SIMILAR TECHNOLOGIES.THE
CUSTOMER WILL NEED TO MAKE ITS OWN INDEPENDENT
DETERMINATION ABOUT THE ADEQUACY OF SUCH ITEMS.
THE EQUIPMENT PROVIDED WILL NOT BE RESPONSIBLE
FOR ISSUES ARISING FROM THE NETWORK
ENVIRONMENT.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-2 CT EQUIPMENT PLACEMENTNorth
3
12
15 13
5
7 8
9 10
4
12
CBTB-021A TABLE 2
1
MAX TRAVEL
W/HEAD REST
EXTENSION
6
11
11
14
6'-7"
7'-4"
EXISTING
ELECT.
TRANS.
EXIST
ELECT
PANEL
12
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-3 CT EQUIPMENT CLEARANCESNorth
MAX TRAVEL
W/HEAD REST
EXTENSION
3'-6 1
16"
13'-13
8"
11 5
8"101
8"
2'-111
4"10'-11 7
8"
5'-3 3
8"
5'-9 3
16"
RECOMMENDED
DISTANCE
TO WALL
4'-51
8"
4'-9 3
16"
2'-2"
1'-61
8"
8'
GANTRY TILTGANTRY TILT
MIN. ROOM SIZE
MINIMUM SERVICE CLEARANCE
N.E.C. CLEARANCE
EXISTING
ELECT.
TRANS.
EXIST
ELECT
PANEL
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-4 CT RADIATION SCATTER PLOTSHIELDING DESIGN NOTES FOR CUSTOMER
1.THE CUSTOMER IS RESPONSIBLE FOR RETAINING THE SERVICES OF A QUALIFIED
MEDICAL PHYSICIST TO PREPARE A RADIATION SHIELDING DESIGN AND TO HIRE
CONTRACTORS TO FURNISH AND INSTALL THE SHIELDING PRESCRIBED BY THE
MEDICAL PHYSICIST.
2.SHIELDING IS REQUIRED TO PROTECT THOSE PERSONS WORKING ADJACENT TO X-RAY
ROOMS. LEAD WHICH OFFERS SPECIAL SHIELDING PROPERTIES AT MEDICAL X-RAY
ENERGIES IS THE PREFERRED AND MOST USED MATERIAL.
3.ALL DIAGNOSTIC SHIELDING WALLS REQUIRE THE SHIELDING TO EXTEND FROM THE
FLOOR TO AT LEAST 7 FEET ABOVE THE FLOOR. THE DOORS AND WINDOWS IN THEM
ARE REQUIRED TO BE OF EQUIVALENT SHIELDING IN ALL CASES. THAT IS, WHEN A
WALL REQUIRES 1/16 INCH LEAD SHIELDING, A DOOR OR WINDOW INSTALLED IN THAT
WALL IS ALSO REQUIRED TO HAVE SHIELDING EQUIVALENT TO 1/16 INCH OF LEAD.
PARTICULAR ATTENTION MUST BE GIVEN TO THE CONSTRUCTION OF ALL JOINTS AND
WHERE PENETRATIONS OCCUR.
4.IN THIS DESIGN, THERE IS NO OCCUPANCY ABOVE THE CEILING OR BELOW THE FLOOR;
THEREFORE, THE USUAL CONSTRUCTION MATERIALS PROVIDE ADEQUATE SHIELDING
SO THAT NO ADDITIONAL LEAD IS REQUIRED IN THE FLOORS AND CEILINGS.
5.THE CONTROL BOOTH IS DESIGNED TO MAXIMIZE THE PROTECTION OF THE OPERATOR
WHILE PROVIDING THE OPERATOR THE MOST ADVANTAGEOUS VIEW OF THE ROOM. IN
ALL CASES, THE CONTROL SWITCH MUST BE POSITIONED AT LEAST 40 INCHES FROM
THE OUTSIDE EDGE OF THE CONTROL BOOTH.
SCATTER RADIATION NOTES
UNITS: MICROGRAY (PER 100 mAs)
THE AMOUNT OF SCATTERED RADIATION DURING SCANNING IS SHOWN IN THE SCATTER
DIAGRAM.
REFER TO THIS DATA WHEN SCANNING TO MINIMIZE X-RAY EXPOSURE. NOTE THAT THE
AMOUNT OF SCATTERED RADIATION DIFFERS DEPENDING ON THE PATIENT; THEREFORE,
THE DATA SHOWN SHOULD BE USED ONLY AS A GUIDE.
120 kVp/100 mA/1.0 sec/M FOV/0,5 mm x 80/320 mm PMMA PHANTOM
THE ABOVE CALCULATION IS BASED ON A MULTI-SLICE EXPOSURE.
THE CUSTOMER IS RESPONSIBLE FOR HAVING SHIELDING CALCULATIONS PREPARED BY A
LICENSED RADIATION PHYSICIST.
SHIELDING DESIGN NOTES FOR CUSTOMER
1.THE CUSTOMER IS RESPONSIBLE FOR RETAINING THE SERVICES OF A QUALIFIED
MEDICAL PHYSICIST TO PREPARE A RADIATION SHIELDING DESIGN AND TO HIRE
CONTRACTORS TO FURNISH AND INSTALL THE SHIELDING PRESCRIBED BY THE
MEDICAL PHYSICIST.
2.SHIELDING IS REQUIRED TO PROTECT THOSE PERSONS WORKING ADJACENT TO X-RAY
ROOMS. LEAD WHICH OFFERS SPECIAL SHIELDING PROPERTIES AT MEDICAL X-RAY
ENERGIES IS THE PREFERRED AND MOST USED MATERIAL.
3.ALL DIAGNOSTIC SHIELDING WALLS REQUIRE THE SHIELDING TO EXTEND FROM THE
FLOOR TO AT LEAST 7 FEET ABOVE THE FLOOR. THE DOORS AND WINDOWS IN THEM
ARE REQUIRED TO BE OF EQUIVALENT SHIELDING IN ALL CASES. THAT IS, WHEN A
WALL REQUIRES 1/16 INCH LEAD SHIELDING, A DOOR OR WINDOW INSTALLED IN THAT
WALL IS ALSO REQUIRED TO HAVE SHIELDING EQUIVALENT TO 1/16 INCH OF LEAD.
PARTICULAR ATTENTION MUST BE GIVEN TO THE CONSTRUCTION OF ALL JOINTS AND
WHERE PENETRATIONS OCCUR.
4.IN THIS DESIGN, THERE IS NO OCCUPANCY ABOVE THE CEILING OR BELOW THE FLOOR;
THEREFORE, THE USUAL CONSTRUCTION MATERIALS PROVIDE ADEQUATE SHIELDING
SO THAT NO ADDITIONAL LEAD IS REQUIRED IN THE FLOORS AND CEILINGS.
5.THE CONTROL BOOTH IS DESIGNED TO MAXIMIZE THE PROTECTION OF THE OPERATOR
WHILE PROVIDING THE OPERATOR THE MOST ADVANTAGEOUS VIEW OF THE ROOM. IN
ALL CASES, THE CONTROL SWITCH MUST BE POSITIONED AT LEAST 40 INCHES FROM
THE OUTSIDE EDGE OF THE CONTROL BOOTH.
RADIATION SCATTER PLOT
NOTE TO CUSTOMER
YOUR PHYSICIST WILL NEED ALL THE INFORMATION ON THIS SHEET TO PROVIDE YOU WITH THE
"PRESCRIPTION" FOR YOUR RADIATION (LEAD) SHIELDING ALONG WITH YOUR SCAN VOLUMES, SCAN
TYPES, AND OCCUPANCY OF THE SPACE AND ADJACENT SPACES. RADIATION SHIELDING IS THE
CUSTOMER'S RESPONSIBILITY.
RADIATION SCATTER PLOT
150
100
50
20
10
5
100
50
10
5
10
5
150
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-5 SYSTEM DETAILS7'-4"
8'-0"
6'-9 1
2"
3'-5 5
16"
7'-4"
3'-85
8"
1'-101
2"
18'-10 11
16"
5'-9 3
16"
RECOMMENDED
DISTANCE
TO WALL5'-4"
MAX TRAVEL
W/HEAD REST
EXTENSION
3'-6 1
16"
13'-13
8"
11 5
8"101
8"
2'-111
4"10'-11 7
8"
5'-3 3
8"
4'-51
8"
4'-9 3
16"
2'-2"8'
GANTRY TILTGANTRY TILT
9'-5 5
8"
6'-91
2"
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-6 SYSTEM DETAILS2'-8 1
16"
1'-511
16"
2'-31
2"
NOT USED
2'-8 1
16"
1'-511
16"
2'-31
2"
SIDEFRONT
5 1
2"
7 7
8'
5 5
16"
5 1
2"
NOT USED
2'-8"
4'-3 3
4"
3'-5"
2'-8"
14 7
8"
4"
13 1
2"
CPU - CENTRAL PROCESSING UNIT
HEAT BTU = 5,123
WEIGHT = 276 LBS
PLAN 2'-2 1
2"
5'-0 5
8"
REC - RECONSTRUCTION UNIT
#1 HEAT BTU = 26,040 - 816 LBS
#2 HEAT BTU INCL #1 - 728 LBS
PCDU - POWER DISTRIBUTION UNIT
HEAT BTU = 4,777
WEIGHT = 1,654 LBS
3'-01
2"
2'-21
2"
15 11
16"
15 7
8"
8"
SIDEFRONT
MON - VIEWING MONITOR
HEAT BTU = 192
WEIGHT = 18 LBS
SIDEFRONT SIDEFRONT
SPK - SPEAKERS
HEAT BTU = 0
WEIGHT = 5 LBS
SIDEFRONT
IV - INVERSION
HEAT BTU = T.B.D.
WEIGHT = 17 LBS
13 15
16"
PLAN
PLAN
PLAN
PLAN
SIDEFRONT
PLAN
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
CT-7 RIGGING INFORMATIONRIGGING AND DELIVERY NOTES
1.REFER TO SECTION 7 OF THE SITE PLANNING GUIDE FOR CLEARANCE REQUIREMENTS.
2.THE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING 6,500 LBS SPREAD OVER FOUR 6" DIAMETER RUBBER
WHEELS.
3.MINIMUM OPENINGS 3'-7" CLEAR WIDTH X 81" CLEAR HEIGHT
4.DEPICTED IS THE PRESUMED RIGGING PATH FOR THE EQUIPMENT. THE CUSTOMER MUST ENSURE THE ENTIRE
EQUIPMENT DELIVERY PATH IS CLEAR AND ALL HALLWAYS, CORRIDORS, OPENINGS AND POINTS THE
EQUIPMENT WILL HAVE TO BE TURNED ARE LARGE ENOUGH TO ACCOMMODATE THE EQUIPMENT, SEE SHEET
A-3 FOR TURNING REQUIREMENTS.
INADEQUATE TURN AREA
FOR GANTRY AND TABLE TO COME IN THROUGH
THIS DOOR, IT WOULD BE IMPOSSIBLE TO TURN
THE GANTRY INTO THE SCAN ROOM WITH THE
CONTROL ROOM WALL BUILT.
3'-3"
MIN
GANTRY
8'
3'-7"
7'-9"
MIN GANTRY
3'-3"
MIN
GANTRY
8'
3'-7"
7'-9"
MIN GANTRY
EXISTING
ELECT.
TRANS.
EXIST
ELECT
PANEL
6'-8"
7'-0"
8'-0"
AQUILION DELIVERY REQUIREMENTS
PRIOR TO GANTRY DELIVERY, CHECK THE DELIVERY
ENTRANCE, CORRIDOR WIDTH, ELEVATOR CAPACITY, ETC. TO
ENSURE EASE OF DELIVERY.
COMPENSATION FOR ADDITIONAL CLEARANCES DUE TO THE
RIGGING EQUIPMENT AND HARDWARE TO BE USED MAY BE
REQUIRED, CONSULT WITH THE RIGGING CONTRACTOR WELL
IN ADVANCE OF DELIVERY TO ENSURE THERE IS ADEQUATE
CLEARANCES AND MANEUVERABILITY SPACE.
SITE USAGE, DELIVERY AND INSTALLATION
1)THE MINIMUM STANDARD DELIVERY REQUIREMENTS FOR THE EQUIPMENT ARE AS FOLLOWS:
* THE ENTIRE DELIVERY PATH MUST BE LARGE ENOUGH TO ACCOMMODATE ALL SYSTEM COMPONENTS AND BE FREE AND CLEAR
OF ALL OBSTRUCTIONS.
* MINIMUM CEILING HEIGHTS ARE TO BE AS NOTED IN THESE PLANS OR AS STATED IN THE O.E.M. REFERENCE DRAWINGS.
* FLOORS AND ENTIRE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING THE WEIGHT OF ALL SYSTEM COMPONENTS.
2)FOR STANDARD RIGGING AND DELIVERY ANY OPENINGS, DOORWAYS, OR TURNS IN THE DELIVERY PATH MUST BE ABLE TO
ACCOMMODATE THE GANTRY AND TABLE AS AN ASSEMBLY. SEE DETAILS HEREIN.
3)ADEQUATE STORAGE/STAGING SPACE IS REQUIRED FOR THE DELIVERY OF THE EQUIPMENT. THE STORAGE SPACE MUST BE
INSIDE THE BUILDING WITHIN CLOSE PROXIMITY TO THE WORK AREA, CLIMATE CONTROLLED, AND BE ABLE TO BE SECURED
AFTER HOURS.
4)THE CUSTOMER SHALL PROVIDE AT NO COST TO THE EQUIPMENT SUPPLIER A CONTAINER, (DUMPSTER, ROLL-OFF, ETC.), FOR
THE DISPOSAL OF ALL PACKAGING AND SHIPPING DEBRIS, GENERALLY CONSISTING OF PALLETS, CRATING, BOXES AND
PROTECTIVE PLASTIC WRAP.
5)THE ANTICIPATED TIME FRAME FOR INSTALLATION OF THE MEDICAL EQUIPMENT IS TYPICALLY THREE (3) WORKING DAYS FOR CT
SYSTEMS
6)THE INSTALLATION CREW RESERVES THE RIGHT TO HAVE ACCESS TO THE SITE AND PERFORM WORK OUTSIDE OF NORMAL
WORKING HOURS, (7:30AM - 5:00PM), WITH ADVANCED NOTICE TO THE OWNER.
7)THE PROTECTION OF FINISHED SURFACES IS THE RESPONSIBILITY OF THE CUSTOMER AND THE CUSTOMER SHOULD EXPECT
MINOR TOUCHUP OF FINISHED SURFACES AND PERFORM A FINAL CLEANING AFTER THE COMPLETION OF EQUIPMENT
INSTALLATION, (AT NO COST TO EQUIPMENT PROVIDER).
9'-6"
7'-0"
3'-7"
8'-0"
5'-0"
AQUILION DELIVERY REQUIREMENTS
PRIOR TO GANTRY DELIVERY, CHECK THE DELIVERY
ENTRANCE, CORRIDOR WIDTH, ELEVATOR CAPACITY, ETC. TO
ENSURE EASE OF DELIVERY.
COMPENSATION FOR ADDITIONAL CLEARANCES DUE TO THE
RIGGING EQUIPMENT AND HARDWARE TO BE USED MAY BE
REQUIRED, CONSULT WITH THE RIGGING CONTRACTOR WELL
IN ADVANCE OF DELIVERY TO ENSURE THERE IS ADEQUATE
CLEARANCES AND MANEUVERABILITY SPACE.
MINIMUM DELIVERY REQUIREMENTS
ALL FLOORS ALONG THE DELIVERY PATHWAY MUST BE CAPABLE OF SUPPORTING THE EQUIPMENT AND RIGGING HARDWARE.
ALL DOORWAYS ALONG THE DELIVERY PATH MUST BE AT LEAST 6'-9" CLEAR HEIGHT AND 43" IN CLEAR WIDTH.
THE MINIMUM CLEAR WIDTH OF ANY CORRIDOR LEADING INTO THE CT SCAN ROOM MUST BE AT LEAST 6'-7" WIDE.
SITE USAGE, DELIVERY AND INSTALLATION
1)THE MINIMUM STANDARD DELIVERY REQUIREMENTS FOR THE EQUIPMENT ARE AS FOLLOWS:
* THE ENTIRE DELIVERY PATH MUST BE LARGE ENOUGH TO ACCOMMODATE ALL SYSTEM COMPONENTS AND BE FREE AND CLEAR
OF ALL OBSTRUCTIONS.
* MINIMUM CEILING HEIGHTS ARE TO BE AS NOTED IN THESE PLANS OR AS STATED IN THE O.E.M. REFERENCE DRAWINGS.
* FLOORS AND ENTIRE DELIVERY PATH MUST BE CAPABLE OF SUPPORTING THE WEIGHT OF ALL SYSTEM COMPONENTS.
2)FOR STANDARD RIGGING AND DELIVERY ANY OPENINGS, DOORWAYS, OR TURNS IN THE DELIVERY PATH MUST BE ABLE TO
ACCOMMODATE THE GANTRY AND TABLE AS AN ASSEMBLY. SEE DETAILS HEREIN.
3)ADEQUATE STORAGE/STAGING SPACE IS REQUIRED FOR THE DELIVERY OF THE EQUIPMENT. THE STORAGE SPACE MUST BE
INSIDE THE BUILDING WITHIN CLOSE PROXIMITY TO THE WORK AREA, CLIMATE CONTROLLED, AND BE ABLE TO BE SECURED
AFTER HOURS.
4)THE CUSTOMER SHALL PROVIDE AT NO COST TO THE EQUIPMENT SUPPLIER A CONTAINER, (DUMPSTER, ROLL-OFF, ETC.), FOR
THE DISPOSAL OF ALL PACKAGING AND SHIPPING DEBRIS, GENERALLY CONSISTING OF PALLETS, CRATING, BOXES AND
PROTECTIVE PLASTIC WRAP.
5)THE ANTICIPATED TIME FRAME FOR INSTALLATION OF THE MEDICAL EQUIPMENT IS TYPICALLY THREE (3) WORKING DAYS FOR CT
SYSTEMS
6)THE INSTALLATION CREW RESERVES THE RIGHT TO HAVE ACCESS TO THE SITE AND PERFORM WORK OUTSIDE OF NORMAL
WORKING HOURS, (7:30AM - 5:00PM), WITH ADVANCED NOTICE TO THE OWNER.
7)THE PROTECTION OF FINISHED SURFACES IS THE RESPONSIBILITY OF THE CUSTOMER AND THE CUSTOMER SHOULD EXPECT
MINOR TOUCHUP OF FINISHED SURFACES AND PERFORM A FINAL CLEANING AFTER THE COMPLETION OF EQUIPMENT
INSTALLATION, (AT NO COST TO EQUIPMENT PROVIDER).
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
S-1 STRUCTURAL AND ANCHORINGC.G. WT. = 1,343 LBS.
(INCLUDES 350 LB. PATIENT)
30"
6" MIN.
C.G. WT. = 3,859 LBS.
FRONT ELEVATION
6" (MIN.)
STRUCTURAL NOTES
1.THE ANCHORING OF THE EQUIPMENT SHALL BE PERFORMED BY THE
EQUIPMENT PROVIDER IN ACCORDANCE WITH THE DESIGN AND
SPECIFICATIONS SET FORTH BY THE CUSTOMER'S STRUCTURAL
ENGINEER OF RECORD.
2.DUE TO THE DYNAMIC NATURE OF THE LOAD, BOTH HORIZONTAL AND
VERTICAL ACCELERATIONS SHOULD BE INCLUDED IN THE DESIGN
CALCULATIONS FOR THE SUPPORT STRUCTURE AS WELL AS
ANCHORING AND THRU-BOLTING FOR THE EQUIPMENT.
3.IF THE FLOOR THE EQUIPMENT IS TO BE PLACED ON IS ANYTHING
OTHER THAN A MINIMUM 6" THICK CONCRETE SLAB THE EQUIPMENT
PROVIDER MUST BE IMMEDIATELY NOTIFIED.
C.G. WT. = 1,343 LBS.
(INCLUDES 350 LB. PATIENT)
30"
6" MIN.
M12 ANCHORS
C.G. WT. = 3,859 LBS.
SEE DETAIL BELOW
FRONT ELEVATION
6" (MIN.)M16 ANCHORS
FLOOR SURFACE REQUIREMENTS
1.THE GANTRY AND PATIENT TABLE MUST BE INSTALLED ON THE
SAME PLANE. IT IS THE CUSTOMER'S RESPONSIBILITY TO MEET
FLOOR LEVELNESS SPECIFICATIONS AS OUTLINED IN THIS DETAIL
AND IN THE APPENDIX OF THE SITE PLANNING MANUAL, (3
16" MAX
OVER ENTIRE AREA).
2.TO ACHIEVE THE FLOOR LEVELNESS SPECIFICATIONS, AN EPOXY
PAD IS TO BE INSTALLED BY THE EQUIPMENT PROVIDER. THE EPOXY
PAD MUST BE INSTALLED DIRECTLY OVER THE CONCRETE FLOOR,
(NOT OVER ANY FINISH FLOORING). SEE APPENDIX SECTION OF THE
SITE PLANNING MANUAL FOR FULL DETAILS AND SPECIFICATIONS.
3.THE CURING OF THE PAD IS TEMPERATURE SENSITIVE. IN ORDER TO
PROPERLY CURE, THE SLAB MUST BE AT A MINIMUM OF 60°F (48 HR.
CURE TIME) OR 77°F (24 HR. CURE TIME). THIS SHOULD BE
COORDINATED WITH THE EQUIPMENT PROJECT MANAGER AND
MUST BE COMPLETED PRIOR TO DELIVERY OF THE SYSTEM.
4.THE EQUIPMENT MAY BE SET ON TOP OF FINISHED FLOORING ONLY
IF THE FINISH FLOORING PRODUCT WILL NOT COMPRESS AND CAN
SUPPORT THE EQUIPMENT WITHOUT COMPRESSING OVER TIME.
BOLTING REQUIREMENTS
1.A MINIMUM OF A 6" THICK CONCRETE SLAB IS REQUIRED FOR
ANCHORING.
2.THE WEIGHT CAPACITY OF THE FLOOR MUST BE EVALUATED BY THE
CUSTOMER'S STRUCTURAL ENGINEER. THE EQUIPMENT PROVIDER
RESERVES THE RING, IN THE INTEREST OF SAFETY, TO DELAY
INSTALLATION UNTIL UNTIL STRUCTURAL DESIGN DRAWINGS
STAMPED BY THE CUSTOMERS STRUCTURAL ENGINEER ARE
REVIEWED BY THE EQUIPMENT PROVIDER.
3.THE PATIENT TABLE AND GANTRY MUST ALWAYS BE BOLTED TO THE
FLOOR THROUGH THE DESIGNATED ATTACHMENT POINTS IN THE
GANTRY AND TABLE PEDESTAL.
4.ANCHORS TO MUST PROVIDE A MINIMUM EXTRACTION FORCE OF 621
POUNDS PER ATTACHMENT POINT IS REQUIRED.
5.IT IS THE RESPONSIBILITY OF THE CUSTOMER'S STRUCTURAL
ENGINEER TO DETERMINE THE ANCHORING DEPTH AND CONCRETE
STRENGTH NEEDED TO INSTALL THE GANTRY AND TABLE BASE.
CENTER OF GRAVITY
ISO-CENTER
3'-4 7
16"
1'-7 11
16"1'-11 7
16"
4 1
2"
10 5
8"3'-5 13
16"
STANDARD ANCHOR LOCATIONS SHOWN
SHOWN, IF FACILITY REQUIRES SEISMIC
ANCHORING, CONTACT YOUR PROJECT
PLANNER.
COORDINATE ANCHOR HOLE SIZES AND
LOCATION WITH INSTALLERS
INSTALLERS.
GANTRY AND TABLE BASE DETAIL
REFER TO APPENDIX, PAGE 77 OF SITE PLANNING MANUAL
FOR FULL REQUIREMENT, DIMENSIONS, SPECIFICATIONS
AND EPOXY LEVELING PAD INSTRUCTIONS
3'-10 7
16"
13"
2 3
4"
3 1
8"
3'-1 3
8"
7 1
16"
215
16"
7 7
8"14 15
16"
6'-0 7
16"
17 11
16"
7 7
8"
20 7
16"
12 3
16"
THE AREA BEHIND THE GANTRY
MUST BE KEPT FLAT AND LEVEL
EPOXY IN THIS AREA MUST BE REMOVED
TO USE MAIN DETECTOR REPLACEMENT
JIG. FLOOR TO BE FLAT
SHADED AREA INDICATES
LEVELING EPOXY, MUST BE
INSTALLED OVER CLEAN-DRY
CONCRETE.
13
4"
1 3
16"
10'-4 13
16"
3'-4 151
16 "
3'-5 3
8"
3'-10 7
16"
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
M-1 ENVIRONMENTAL SPECIFICATIONSCT SUITE HVAC REQUIREMENTS
1)AMBIENT TEMPERATURE SHOULD BE 68F TO 74F, HUMIDITY RANGE OF 40%
TO 70% NON CONDENSING. STATIC AMBIENT TEMPERATURE IS TO BE
PROVIDED AND MAINTAINED AS SPECIFIED. ALL CALCULATIONS ARE TO
UTILIZE PROVIDED HEAT OUTPUT SPECIFICATIONS OF EQUIPMENT.
2)A MINIMUM OF 10 AIR CHANGES PER HOUR IS SUGGESTED. CONSULT
LOCAL CODE.
3)EQUIPMENT IN ENCLOSED SPACES SUCH AS EQUIPMENT ROOMS,
CLOSETS AND COMPUTER ROOMS MUST BE PROVIDED WITH ADEQUATE
VENTILATION.
4)A/C SUPPLY OUTLET TO BE PROVIDED BY CUSTOMER AT FLOOR LEVEL OR
AS CLOSE AS POSSIBLE TO CONTROL ROOM DESK.
5)DUE TO HEAT GENERATED BY THE "CPU" AND "REC" UNITS, ADDITIONAL
VENTILATION IN THE CONTROL AREA IS REQUIRED. CUSTOMER OR
CONTRACTOR PROVIDED FANS MAY BE NECESSARY BELOW THE DESKTOP
FOR TECHNICIAN COMFORT. THE "CPU/REC" UNITS SHOULD NOT BE BE
ENCLOSED IN CASEWORK.
6)AIR SUPPLY DUCTS SHOULD NOT BE PLACED DIRECTLY OVER
EXAMINATION TABLES OR GANTRY FOR PATIENT COMFORT
CONSIDERATIONS.
7)THE AIRFLOW THROUGH TOSHIBA EQUIPMENT CABINETS IS FROM BOTTOM
TO TOP.
8)WHERE POSSIBLE A/C SUPPLY OUTLETS SHOULD BE LOCATED AT FLOOR
LEVEL. NO A/C OUTLET SHOULD BE WITHIN THE EXCLUSION ZONE SHOWN
BELOW AND AT NO TIME SHOULD THE CT SYSTEM BE EXPOSED TO DIRECT
AIRFLOW.
TIMER ON A/C ON
1.5
HRS
SCAN START TIME
TOLERANCE:
±3°F
RECOMMENDED TEMPERATURE CONDITIONS
IN CT SCAN ROOM
NOTES:
1)BTU / HR SHOWN FOR GANTRY AND PATIENT COUCH ONLY.
2)FUTURE GROWTH OF FACILITY MUST BE CONSIDERED WHEN FORECASTING
PATIENT NUMBERS FOR A/C REQUIREMENTS.
3'-3"
2'-9"
4'-6"4'-6"
A/C SUPPLY AIR
EXCLUSION ZONE
*TOSHIBA EQUIPMENT BTU / HOUR:
(INCLUDING GANTRY & COUCH ONLY)
EXAM ROOM STAND-BY:14,672 BTU/HR
EXAM ROOM SCANNING 2 PATIENTS:20,712 BTU/HR
EXAM ROOM SCANNING 3 PATIENTS:21,991 BTU/HR
EXAM ROOM SCANNING 4 PATIENTS:23,271 BTU/HR
EXAM ROOM SCANNING 5 PATIENTS:24,550 BTU/HR
EXAM ROOM SCANNING MAXIMUM:25,830 BTU/HR
NOTE: FUTURE GROWTH OF FACILITY MUST BE CONSIDERED
WHEN FORECASTING PATIENT NUMBERS FOR A/C
REQUIREMENTS.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-1 CT ELECTRICAL NOTESELECTRICAL NOTES
1.THESE SITE PLANS ARE INTENDED TO DEPICT THE CONCEPT OF THE ELECTRICAL REQUIREMENTS FOR THE
EQUIPMENT. THE DESIGN OF ALL ELECTRICAL ELEMENTS MUST BE SPECIFIED BY A LICENSED ELECTRICAL ENGINEER
IN ACCORDANCE WITH THE EQUIPMENT SPECIFICATIONS AND ALL APPLICABLE CODES.
2.IN ACCORDANCE WITH N.E.C. ARTICLE 517-72(B), THE EQUIPMENT CIRCUIT BREAKER(S) DISCONNECT MUST BE
LOCATED SO THAT THEY SHALL BE OPERABLE FROM A LOCATION READILY ACCESSIBLE FROM THE CONTROL AREA. IF
THIS IS IMPOSSIBLE OR IMPRACTICAL, THE USE OF A SHUNT TRIP TYPE BREAKER WILL BE NECESSARY TO SATISFY
THIS REQUIREMENT. THE EMERGENCY POWER OFF BUTTON FOR THE SHUNT TRIP SHOULD BE LOCATED IN THE
CONTROL AREA.
3.THE CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL CIRCUIT BREAKERS, CONDUITS, WIRES, JUNCTION
BOXES, DUCTS, TROUGHS, ETC.
4.THESE PLANS DO NOT SPECIFY ELECTRICAL REQUIREMENTS FOR EQUIPMENT NOT PROVIDED WITH THE CT SYSTEM.
5.INTERCONNECTING CABLES FOR THE CT EQUIPMENT ARE FURNISHED WITH THE CT SYSTEM AND INSTALLED BY THE
CT INSTALLERS.
6.EXCEPT FOR THEIR USE IN POWER LINE CONNECTIONS TO EQUIPMENT CABINETS, FLEXIBLE CONDUIT SHALL NOT BE
USED IN THIS INSTALLATION. ONLY FACTORY CONDUIT ELBOWS SHALL BE USED. DUCT WORK SHALL BE PROVIDED
WITH SWEEP ELBOWS.
7.ALL JUNCTION BOXES AND DUCTS THAT PENETRATE THE FLOOR SHALL BE WATERPROOF TYPE AND PROVIDED WITH
GASKETED WATERPROOF COVERS. ALL FLOOR JUNCTION BOXES AND DUCT COVERS SHALL BE CAPABLE OF
SUPPORTING 200 LBS OF CONCENTRATED LOAD.
8.GROMMETED OPENINGS ARE SHOWN FOR REFERENCE PURPOSES ONLY. VERIFY SIZE AND LOCATION WITH
INSTALLATION REPRESENTATIVE. ALL GROMMETED OPENINGS SHALL HAVE NO SHARP EDGES.
9.ALL CHASE & GROMMETED OPENINGS SHALL HAVE PLASTIC BUSHINGS.
10.ALL DUCT WORK SHALL HAVE A MINIMUM OF THREE COMPARTMENTS WITH TRANSITIONS FROM HORIZONTAL TO
VERTICAL WALL DUCT OR JUNCTION BOXES. LOCAL CODES MAY REQUIRE THE USE OF CROSS-OVER TUNNELS OR
OTHER SUCH DEVICES TO MAINTAIN CABLE SEPARATION.
11.ALL DUCTS, TROUGHS AND CONDUITS MUST BE ELECTRICALLY BONDED AS A GROUNDING PATH IN ACCORDANCE WITH
N.E.C. ARTICLE 517-13(B).
12.THE ELECTRICAL CONTRACTOR SHALL SUPPLY AND INSTALL NYLON MEASURING PULL STRING OR EQUIVALENT IN ALL
CONDUITS AND CLOSED DUCT WORK.
13.CONDUIT RUNS SHOWN ARE FOR REFERENCE ONLY. ALL CONDUIT RUNS MUST TAKE THE SHORTEST MOST DIRECT
ROUTE POSSIBLE. CONDUIT RUNS MAY HAVE A MAXIMUM OF (3) 90 DEGREE BENDS.
14.110 VAC GROUNDED OUTLETS SHALL BE PROVIDED ON WALLS NEAR EQUIPMENT FOR USE DURING EQUIPMENT
SERVICE.
15.THE ELECTRICAL CONTRACTOR MUST SUPPLY AND INSTALL ALL INCOMING POWER CABLES FROM CIRCUIT BREAKER(S)
TO THE EQUIPMENT.
16.CABLE TYPE MUST BE MTW MULTI-STRAND COPPER - NO ALUMINUM IS PERMITTED. CABLE SIZE MUST BE IN
ACCORDANCE WITH POWER QUALITY REQUIREMENTS SPECIFICATIONS.
17.CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL NECESSARY HARDWARE TO ENCLOSE INCOMING POWER
CABLES IN WATER TIGHT CONDUIT FROM CIRCUIT BREAKER(S) TO EQUIPMENT CABINET(S).
18.ANY CHANGES IN THE LOCATION OR TYPE OF CONDUIT, DUCT WORK, JUNCTION BOXES, ETC. MUST BE SUBMITTED IN
WRITING TO THE INSTALLATION PROJECT MANAGER FOR APPROVAL.
19.ALL DUCT WORK MAKING A 90 DEGREE ANGLE MUST BE CHAMFERED FOR CABLE ACCESS.
20.JUNCTION BOX SIZES MAY BE INCREASED AS NEEDED (WITH EXCEPTION OF THE "PCDU" / "VUC" / "PCDU" / OR "PDU"
JUNCTION BOX IF APPLICABLE.).
21.FIBER OPTIC CABLES FROM "REC" TO "GANT" & "REC" TO "CPU" REQUIRE A MINIMUM RADIUS OF 4 1/2". DUCT WORK
DESIGN MUST ACCOMMODATE THIS REQUIREMENT.
22.EMERGENCY POWER OFF (EPO) BUTTONS REQUIRED IN CONTROL AREA, EXAMINATION ROOM AND EQUIPMENT AREA..
23.DOOR (SAFETY) SWITCH REQUIRED ON ALL DOORS ACCESSING THE EXAMINATION ROOM IN ACCORDANCE WITH LOCAL
CODES.
ELECTRICAL NOTES
1.THESE SITE PLANS ARE INTENDED TO DEPICT THE CONCEPT OF THE ELECTRICAL REQUIREMENTS FOR THE
EQUIPMENT. THE DESIGN OF ALL ELECTRICAL ELEMENTS MUST BE SPECIFIED BY A LICENSED ELECTRICAL ENGINEER
IN ACCORDANCE WITH THE EQUIPMENT SPECIFICATIONS AND ALL APPLICABLE CODES.
2.IN ACCORDANCE WITH N.E.C. ARTICLE 517-72(B), THE EQUIPMENT CIRCUIT BREAKER(S) DISCONNECT MUST BE
LOCATED SO THAT THEY SHALL BE OPERABLE FROM A LOCATION READILY ACCESSIBLE FROM THE CONTROL AREA. IF
THIS IS IMPOSSIBLE OR IMPRACTICAL, THE USE OF A SHUNT TRIP TYPE BREAKER WILL BE NECESSARY TO SATISFY
THIS REQUIREMENT. THE EMERGENCY POWER OFF BUTTON FOR THE SHUNT TRIP SHOULD BE LOCATED IN THE
CONTROL AREA.
3.THE CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL CIRCUIT BREAKERS, CONDUITS, WIRES, JUNCTION
BOXES, DUCTS, TROUGHS, ETC.
4.THESE PLANS DO NOT SPECIFY ELECTRICAL REQUIREMENTS FOR EQUIPMENT NOT PROVIDED WITH THE CT SYSTEM.
5.INTERCONNECTING CABLES FOR THE CT EQUIPMENT ARE FURNISHED WITH THE CT SYSTEM AND INSTALLED BY THE
CT INSTALLERS.
6.EXCEPT FOR THEIR USE IN POWER LINE CONNECTIONS TO EQUIPMENT CABINETS, FLEXIBLE CONDUIT SHALL NOT BE
USED IN THIS INSTALLATION. ONLY FACTORY CONDUIT ELBOWS SHALL BE USED. DUCT WORK SHALL BE PROVIDED
WITH SWEEP ELBOWS.
7.ALL JUNCTION BOXES AND DUCTS THAT PENETRATE THE FLOOR SHALL BE WATERPROOF TYPE AND PROVIDED WITH
GASKETED WATERPROOF COVERS. ALL FLOOR JUNCTION BOXES AND DUCT COVERS SHALL BE CAPABLE OF
SUPPORTING 200 LBS OF CONCENTRATED LOAD.
8.GROMMETED OPENINGS ARE SHOWN FOR REFERENCE PURPOSES ONLY. VERIFY SIZE AND LOCATION WITH
INSTALLATION REPRESENTATIVE. ALL GROMMETED OPENINGS SHALL HAVE NO SHARP EDGES.
9.ALL CHASE & GROMMETED OPENINGS SHALL HAVE PLASTIC BUSHINGS.
10.ALL DUCT WORK SHALL HAVE A MINIMUM OF THREE COMPARTMENTS WITH TRANSITIONS FROM HORIZONTAL TO
VERTICAL WALL DUCT OR JUNCTION BOXES. LOCAL CODES MAY REQUIRE THE USE OF CROSS-OVER TUNNELS OR
OTHER SUCH DEVICES TO MAINTAIN CABLE SEPARATION.
11.ALL DUCTS, TROUGHS AND CONDUITS MUST BE ELECTRICALLY BONDED AS A GROUNDING PATH IN ACCORDANCE WITH
N.E.C. ARTICLE 517-13(B).
12.THE ELECTRICAL CONTRACTOR SHALL SUPPLY AND INSTALL NYLON MEASURING PULL STRING OR EQUIVALENT IN ALL
CONDUITS AND CLOSED DUCT WORK.
13.CONDUIT RUNS SHOWN ARE FOR REFERENCE ONLY. ALL CONDUIT RUNS MUST TAKE THE SHORTEST MOST DIRECT
ROUTE POSSIBLE. CONDUIT RUNS MAY HAVE A MAXIMUM OF (3) 90 DEGREE BENDS.
14.110 VAC GROUNDED OUTLETS SHALL BE PROVIDED ON WALLS NEAR EQUIPMENT FOR USE DURING EQUIPMENT
SERVICE.
15.THE ELECTRICAL CONTRACTOR MUST SUPPLY AND INSTALL ALL INCOMING POWER CABLES FROM CIRCUIT BREAKER(S)
TO THE EQUIPMENT.
16.CABLE TYPE MUST BE MTW MULTI-STRAND COPPER - NO ALUMINUM IS PERMITTED. CABLE SIZE MUST BE IN
ACCORDANCE WITH POWER QUALITY REQUIREMENTS SPECIFICATIONS.
17.CUSTOMER/CONTRACTOR IS TO SUPPLY AND INSTALL ALL NECESSARY HARDWARE TO ENCLOSE INCOMING POWER
CABLES IN WATER TIGHT CONDUIT FROM CIRCUIT BREAKER(S) TO EQUIPMENT CABINET(S).
18.ANY CHANGES IN THE LOCATION OR TYPE OF CONDUIT, DUCT WORK, JUNCTION BOXES, ETC. MUST BE SUBMITTED IN
WRITING TO THE INSTALLATION PROJECT MANAGER FOR APPROVAL.
19.ALL DUCT WORK MAKING A 90 DEGREE ANGLE MUST BE CHAMFERED FOR CABLE ACCESS.
20.JUNCTION BOX SIZES MAY BE INCREASED AS NEEDED (WITH EXCEPTION OF THE "PCDU" / "VUC" / "PCDU" / OR "PDU"
JUNCTION BOX IF APPLICABLE.).
21.FIBER OPTIC CABLES FROM "REC" TO "GANT" & "REC" TO "CPU" REQUIRE A MINIMUM RADIUS OF 4 1/2". DUCT WORK
DESIGN MUST ACCOMMODATE THIS REQUIREMENT.
22.EMERGENCY POWER OFF (EPO) BUTTONS REQUIRED IN CONTROL AREA, EXAMINATION ROOM AND EQUIPMENT AREA..
23.DOOR (SAFETY) SWITCH REQUIRED ON ALL DOORS ACCESSING THE EXAMINATION ROOM IN ACCORDANCE WITH LOCAL
CODES.
AQUILION ONE CT SCANNER POWER REQUIREMENTS
SUPPLY CONFIGURATION: 3 PHASE, 3 WIRE POWER, AND GROUND DELTA OR WYE (SEE NOTE A)
NOMINAL LINE VOLTAGE: 480 VAC 60 HZ (SEE NOTE B)
LINE VOLTAGE VARIATION:±10 % STEADY-STATE INCLUDING SAGS AND SURGES
LINE VOLTAGE BALANCE: 2 % MAXIMUM OF NOMINAL VOLTAGE BETWEEN PHASES
FREQUENCY VARIATIONS: ±1 HZ
HARMONIC DISTORTION: 3% STEADY-STATE, 5% FOR SHORT PERIODS (1 MINUTE OR LESS)
GROUND CONDUCTOR IMPEDANCE: 0.1 OHMS @ 60HZ, TO NEUTRAL-GROUND BONDING POINT (SEE NOTE D)
STANDARD TRANSFORMER CAPACITY:150 KVA
MAXIMUM SYSTEM DEMAND:125 KVA (IMAGING)
CONDUCTOR SIZES (SEE NOTE E)
FOR 1.5% IMPEDANCE OF BRANCH CONDUCTORS (20°C
CONDUCTOR SIZE 480 VAC BREAKER FRAME SIZE
1/0 AWG 377 FT.200 A
2/0 AWG 444 FT.200 A
CIRCUIT BREAKER SIZE:150 A (SEE NOTE F)
MOMENTARY MAX CURRENT:163 A 150 A
MAXIMUM VOLTAGE DROP:24.0 V
REGULATION %:5.0%
STANDARD POWER QUALITY NOTES:
1.A GROUNDED NEUTRAL POWER SOURCE IS REQUIRED TO ASSURE RELIABLE EQUIPMENT OPERATION. THE
NEUTRAL CONDUCTOR MAY NOT BE USED FOR A PARTICULAR SYSTEM.
2.IN CASES WHERE MULTIPLE VOLTAGES ARE PERMITTED, THE PREFERRED SYSTEM VOLTAGE IS SPECIFIED.
3.DUE TO THE HIGH INSTANTANEOUS POWER OF MEDICAL IMAGING SYSTEMS, USE THE HIGHEST AVAILABLE
VOLTAGE SOURCE. ENSURE THAT LOWER VOLTAGE SOURCES ARE DERIVED DIRECTLY FROM THE SERVICE
ENTRANCE OF THE FACILITY.
4.GROUND CONDUCTORS ARE REQUIRED TO BE THE SAME SIZE AS THE PHASE CONDUCTORS UNLESS A LARGER SIZE
IS REQUIRED BY CODE.
5.ALL FEEDER AND BRANCH CIRCUIT CONDUCTORS MUST BE COPPER - ALUMINUM IS NOT PERMITTED.
6.IF THE EQUIPMENT CIRCUIT BREAKER IS NOT LOCATED IN THE CONTROL AREA, A SHUNT TRIP BREAKER MUST BE
USED IN ORDER TO COMPLY WITH N.E.C. 517-72(B). A PUSH-BUTTON TO OPERATE THE SHUNT TRIP MUST BE
LOCATED IN THE CONTROL AREA.
7.A SEPARATE CIRCUIT, FED FROM THE FACILITY RADIOLOGY PANEL OR A MAIN SERVICE PANEL IS REQUIRED. USE
OF A SUB PANEL WITH LOADS SUCH AS ELEVATORS, HVAC, MOTORS, ETC., IS NOT PERMITTED.
8.DEVICES SUCH AS UNINTERRUPTIBLE POWER SUPPLIES, POWER CONDITIONERS, VOLTAGE REGULATORS, AND
FILTERS MAY BE INCOMPATIBLE WITH THIS IMAGING EQUIPMENT. CONSULT YOUR TOSHIBA SERVICE
REPRESENTATIVE PRIOR TO PURCHASING OR INSTALLING THESE DEVICES.
9.IF THE MAINS POWER GROUND CONDUCTOR IS TO BE RUN WITH THE POWER PHASE CONDUCTORS, THE GROUNDS
TO BUILDING STEEL OR EARTH GROUND ARE NOT TO BE RUN WITH THESE PHASE CONDUCTORS.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-2 ELECTRICAL NOTESGENERAL ELECTRICAL NOTES
1.ALL CONDUITS, TROUGHS, JUNCTION BOXES, BREAKERS, WIRES, DEVICES AND ACCESSORIES ARE FURNISHED AND INSTALLED BY CUSTOMER'S ELECTRICAL CONTRACTOR.
2.CONDUIT RUN IS NOT TO BE RUN IN SUCH A MANNER THAT WILL ALLOW CABLE POINT TO POINT LENGTHS TO BE EXCEEDED AS SHOWN IN CONDUIT LEGEND ON SHEET E-7.
3.ALL ELECTRICAL WORK SHALL BE IN COMPLIANCE WITH THE LATEST EDITION OF THE NATIONAL ELECTRICAL CODE (NFPA-70), O.S.H.A. REGULATIONS, AS WELL AS APPLICABLE REGULATIONS OF CITY, COUNTY, STATE AND FEDERAL AGENCIES.
PROVIDE MATERIALS AND EQUIPMENT THAT COMPLY TO ANSI, IEEE AND NEMA STANDARDS. WHERE APPLICABLE, PROVIDE ONLY MATERIALS AND PRODUCTS THAT ARE U.L. LISTED AND LABELED. CUSTOMER'S/CONTRACTOR'S WORK SHALL
COMPLY WITH THE LATEST EDITION OF NECA STANDARD OF INSTALLATION.
4.THE CONTRACTOR SHALL VERIFY EXISTING CONDITIONS IN THE FIELD TO INSURE THAT THE NEW WORK WILL FIT TO THE EXISTING STRUCTURE AS SHOWN ON THE DRAWINGS. SHOULD ANY CONDITIONS EXIST OR BE DISCOVERED THAT
PREVENT THE INSTALLATION OF WORK AS SHOWN, THE CONTRACTOR SHALL NOTIFY THE OWNER'S REPRESENTATIVE PRIOR TO FABRICATION OF EQUIPMENT, OR THE PERFORMANCE OF ANY WORK THAT MAY BE AFFECTED. DO NOT ALTER
DRAWINGS, DIMENSIONS, OR SPECIFICATIONS IN ANY WAY WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM THE SITE PLANNER. ALL DIMENSIONS ARE FROM FINISHED SURFACES. CONDUIT AND PULL BOXES TO BE
INSTALLED BY THE CUSTOMER/CONTRACTOR WITH LOCATIONS BEING FIELD VERIFIED WITH THE EQUIPMENT INSTALLERS.
5.POWER SUPPLIES FOR THE EQUIPMENT SHALL BE DEDICATED SERVICES KEPT ENTIRELY FREE AND INDEPENDENT OF ALL OTHER BUILDING WIRING AND EQUIPMENT, SUCH AS: ELEVATORS, GENERATORS, PUMPS, HVAC SYSTEMS, ETC. THE
CONTRACTOR SHALL COORDINATE THIS WORK WITH THE CUSTOMER/UTILITY COMPANY FIELD REPRESENTATIVE.
6.WORK NOT PROVIDED BY THE EQUIPMENT INSTALLATION TEAM BUT SHOWN ON DRAWINGS TO BE FURNISHED AND INSTALLED BY CUSTOMER/CONTRACTOR INCLUDES, BUT IS NOT LIMITED TO, THE FOLLOWING UNLESS NOTED OTHERWISE:
ELECTRICAL RACEWAYS AND DUCTS, WIRING TROUGHS, PULL BOXES, CONDUITS, CIRCUIT BREAKERS AND EMERGENCY OFF BUTTONS, DOOR SWITCHES AND WARNING LIGHTS, WIRING, WIRING DEVICES, CONNECTORS, LIGHT FIXTURES AND
GROUNDING
7.RACEWAY SHALL BE ELECTRIC METALLIC TUBING (EMT) FOR RIGID CONDUIT WORK, OR WHERE SHORT OFFSET CONNECTIONS ARE REQUIRED LIQUID TIGHT FLEXIBLE METAL CONDUIT SHALL BE USED.
8.FIELD BENDS SHALL NOT BE LESS THAN AS SHOWN IN THE NATIONAL ELECTRICAL CODE.
9.PROVIDE A JETLINE "SUPER TRUE TAPE", OR EQUIVALENT CONDUIT MEASURING TAPE FISH LINE IN ALL RACEWAYS AND CONDUITS.
10.CONDUIT BODIES SHALL NOT BE USED. WHERE A CONDUIT ENTERS A BOX, FITTING, OR OTHER ENCLOSURE, AN INSULATED THROAT CONNECTOR SHALL BE PROVIDED TO PROTECT THE WIRE FROM ABRASION. CONNECTORS SHALL BE DOUBLE
SET SCREW TYPE, STEEL CONCRETE TIGHT. KEEP RACEWAYS AT LEAST 6 INCHES AWAY FROM PARALLEL RUNS OF FLUES OR STEAM AND HOT WATER PIPES. INSTALL RACEWAY RUNS ABOVE WATER AND STEAM PIPES PROVIDED THAT CABLE
RUN DISTANCES ARE MAINTAINED. USE TEMPORARY CLOSURES TO PREVENT FOREIGN MATTER FROM ENTERING RACEWAY.
11.CONDUIT RUNS ARE SHOWN SCHEMATICALLY. INSTALL CONDUIT WITH A MINIMUM OF BENDS IN THE SHORTEST PRACTICAL DISTANCE CONSIDERING THE BUILDING CONSTRUCTION AND OBSTRUCTIONS, EXCEPT AS OTHERWISE INDICATED. THE
CONTRACTOR SHALL MAKE CERTAIN THAT ANY CONDUIT/RACEWAY RUNS CONTAINING MEDICAL EQUIPMENT SYSTEMS CABLES DO NOT EXCEED THE SPECIFIED MAXIMUM DISTANCES AS SHOWN ON THE ELECTRICAL DETAILS.
12.PROVIDE ENCLOSED METAL RACEWAY SYSTEM (WIRE DUCT) WHERE SHOWN ON DRAWINGS WITH DIVIDERS TO SEPARATE THE DUCT (FOR POWER AND MEDICAL EQUIPMENT CABLING). DIVIDERS AND CROSSOVER PIECES TO BE PROVIDED AS
NECESSARY.
13.FOR UL CERTIFIED SYSTEMS, THE CABLE TO CABLE AS WELL AS THE CIRCUIT TO CIRCUIT SEPARATION REQUIREMENT WAS EVALUATED DURING THE UL SYSTEM INVESTIGATION OF THIS EQUIPMENT. ADDITIONAL SEPARATION OF THE SYSTEM
CABLE ASSEMBLIES INTO SEPARATE OR PARTITIONED RACEWAYS, UNLESS OTHERWISE NOTED, IS NOT NECESSARY TO INSURE SEPARATION OF CIRCUITS, AS THEY CAN BE IN THE SAME RACEWAY.
14.PROVIDE WIRE DUCT/RACEWAY WITH ACCESSIBLE REMOVABLE COVERS. LOCATIONS OF OPENINGS (I.E. ACCESS PANELS) TO BE CUT IN FIELD ARE TO BE COORDINATED WITH THE EQUIPMENT INSTALLATION TEAM. ELECTRICAL PULL BOXES AND
RACEWAY COVERS SHALL BE INSTALLED IN A MANNER TO ALLOW ACCESSIBILITY FOR INSTALLATION AND MAINTENANCE.
15.IN-FLOOR TRENCH DUCT AND FLUSH FLOOR BOXES SHALL BE PROVIDED WITH FULLY CASKETED REMOVABLE COVERS.
16.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED HIGHER THAN 14 FEET ABOVE FINISHED FLOOR, THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL EQUIPMENT
CABLES AT THE CUSTOMER EXPENSE.
17.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED ABOVE A HARD CEILING (I.E. SHEET ROCK), A 24" x 24" ACCESS PANEL IS REQUIRED AT EACH JUNCTION BOX AND WITHIN 2 FEET OF EACH 90 DEGREE ELBOW OR TEE IN WIRE
DUCT/RACEWAY. THERE MUST BE FREE AND CLEAR ACCESS TO JUNCTION BOXES AND WIRE DUCT/RACEWAY. WHEN ACCESS PANELS ARE LOCATED MORE THAN 3 FEET FROM JUNCTION BOXES AND WIRE DUCT/RACEWAY THE ELECTRICAL
CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL SUPPLIED SYSTEM CABLES AT THE CUSTOMER EXPENSE.
18.WIRING SHALL BE INSTALLED IN METAL RACEWAY, 600 VOLT CLASS, STRANDED TYPE THHN-THWN, SINGLE CONDUCTOR ANNEALED COPPER FOR A MAXIMUM OPERATING TEMPERATURE OF 75 DEGREES C (165F). SIZED AS INDICATED. THE
CUSTOMER/CONTRACTOR SHALL LEAVE MINIMUM 10 FT. WIRE TAILS AT ALL OUTLET POINTS WITH WIRE IDENTIFICATION TAGGED AT BOTH ENDS FOR FINAL CONNECTION BY THE EQUIPMENT INSTALL TEAM.
19.IN ADDITION TO THE CIRCUIT BREAKER LOAD CURRENT RATING, CONSIDERATION MUST ALSO BE GIVEN TO SELECTING CIRCUIT BREAKERS THAT HAVE A HIGH ENOUGH SHORT CIRCUIT CURRENT WITHSTAND RATING TO SAFELY COORDINATE
WITH THE POWER SYSTEM AVAILABLE SHORT CIRCUIT CURRENT. GENERALLY, WHEN THE 480 VOLT, 3 PHASE, EQUIPMENT IS SERVED FROM A POWER SUPPLY SYSTEM THAT IS PROVIDED WITH A 500 kVA OR SMALLER TRANSFORMER, A
STANDARD 14,000 RMS AMPERE WITHSTAND RATED CIRCUIT BREAKER WILL BE ADEQUATE. HOWEVER, IF THE POWER SUPPLY SYSTEM TRANSFORMER IS LARGER THAN 500 kVA, THEN THE CIRCUIT BREAKERS HAVING A SHORT CIRCUIT
WITHSTAND RATING GREATER THAN 14,000 RMS AMPERES MAY BE REQUIRED.
20.THE ELECTRICAL CONTRACTOR MUST ADJUST CONDUIT RUN LENGTHS FOR SYSTEM CABLES BASED ON THE MAXIMUM LENGTHS LISTED IN THE SYSTEM CABLE DIAGRAM.
21.THE EPO'S MUST BE INSTALLED BY A QUALIFIED ELECTRICAL CONTRACTOR ACCORDING TO NATIONAL ELECTRICAL CODE, STATE AND LOCAL REGULATIONS. MEASURES SHOULD BE TAKEN TO DESIGN THE CIRCUIT IN SUCH A WAY THAT IT WILL
ALWAYS WORK WHEN THE MEDICAL EQUIPMENT IS POWERED. THE CUSTOMER IS SOLELY RESPONSIBLE FOR THE IMPLEMENTATION OF THE EPO's AND THEIR ASSOCIATED CIRCUITS AND MUST MAKE THE FINAL DETERMINATION CONSIDERING
ALL SITE CONDITIONS AND REGULATORY FACTORS. THE UPS WITH THE IMS SYSTEM, LOCATED IN THE CONTROL ROOM, SUPPLIES IT'S OWN 24V DC CONTROL VOLTAGE AND NEEDS A NORMALLY CLOSED CONTACT FOR EACH EPO. THIS 24V DC
CONTROL VOLTAGE IS TO BE USED FOR THIS UPS ONLY.
GENERAL ELECTRICAL NOTES
1.ALL CONDUITS, TROUGHS, JUNCTION BOXES, BREAKERS, WIRES, DEVICES AND ACCESSORIES ARE FURNISHED AND INSTALLED BY CUSTOMER'S ELECTRICAL CONTRACTOR.
2.CONDUIT RUN IS NOT TO BE RUN IN SUCH A MANNER THAT WILL ALLOW CABLE POINT TO POINT LENGTHS TO BE EXCEEDED AS SHOWN IN CONDUIT LEGEND ON SHEET E-7.
3.ALL ELECTRICAL WORK SHALL BE IN COMPLIANCE WITH THE LATEST EDITION OF THE NATIONAL ELECTRICAL CODE (NFPA-70), O.S.H.A. REGULATIONS, AS WELL AS APPLICABLE REGULATIONS OF CITY, COUNTY, STATE AND FEDERAL AGENCIES.
PROVIDE MATERIALS AND EQUIPMENT THAT COMPLY TO ANSI, IEEE AND NEMA STANDARDS. WHERE APPLICABLE, PROVIDE ONLY MATERIALS AND PRODUCTS THAT ARE U.L. LISTED AND LABELED. CUSTOMER'S/CONTRACTOR'S WORK SHALL
COMPLY WITH THE LATEST EDITION OF NECA STANDARD OF INSTALLATION.
4.THE CONTRACTOR SHALL VERIFY EXISTING CONDITIONS IN THE FIELD TO INSURE THAT THE NEW WORK WILL FIT TO THE EXISTING STRUCTURE AS SHOWN ON THE DRAWINGS. SHOULD ANY CONDITIONS EXIST OR BE DISCOVERED THAT
PREVENT THE INSTALLATION OF WORK AS SHOWN, THE CONTRACTOR SHALL NOTIFY THE OWNER'S REPRESENTATIVE PRIOR TO FABRICATION OF EQUIPMENT, OR THE PERFORMANCE OF ANY WORK THAT MAY BE AFFECTED. DO NOT ALTER
DRAWINGS, DIMENSIONS, OR SPECIFICATIONS IN ANY WAY WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM THE SITE PLANNER. ALL DIMENSIONS ARE FROM FINISHED SURFACES. CONDUIT AND PULL BOXES TO BE
INSTALLED BY THE CUSTOMER/CONTRACTOR WITH LOCATIONS BEING FIELD VERIFIED WITH THE EQUIPMENT INSTALLERS.
5.POWER SUPPLIES FOR THE EQUIPMENT SHALL BE DEDICATED SERVICES KEPT ENTIRELY FREE AND INDEPENDENT OF ALL OTHER BUILDING WIRING AND EQUIPMENT, SUCH AS: ELEVATORS, GENERATORS, PUMPS, HVAC SYSTEMS, ETC. THE
CONTRACTOR SHALL COORDINATE THIS WORK WITH THE CUSTOMER/UTILITY COMPANY FIELD REPRESENTATIVE.
6.WORK NOT PROVIDED BY THE EQUIPMENT INSTALLATION TEAM BUT SHOWN ON DRAWINGS TO BE FURNISHED AND INSTALLED BY CUSTOMER/CONTRACTOR INCLUDES, BUT IS NOT LIMITED TO, THE FOLLOWING UNLESS NOTED OTHERWISE:
ELECTRICAL RACEWAYS AND DUCTS, WIRING TROUGHS, PULL BOXES, CONDUITS, CIRCUIT BREAKERS AND EMERGENCY OFF BUTTONS, DOOR SWITCHES AND WARNING LIGHTS, WIRING, WIRING DEVICES, CONNECTORS, LIGHT FIXTURES AND
GROUNDING
7.RACEWAY SHALL BE ELECTRIC METALLIC TUBING (EMT) FOR RIGID CONDUIT WORK, OR WHERE SHORT OFFSET CONNECTIONS ARE REQUIRED LIQUID TIGHT FLEXIBLE METAL CONDUIT SHALL BE USED.
8.FIELD BENDS SHALL NOT BE LESS THAN AS SHOWN IN THE NATIONAL ELECTRICAL CODE.
9.PROVIDE A JETLINE "SUPER TRUE TAPE", OR EQUIVALENT CONDUIT MEASURING TAPE FISH LINE IN ALL RACEWAYS AND CONDUITS.
10.CONDUIT BODIES SHALL NOT BE USED. WHERE A CONDUIT ENTERS A BOX, FITTING, OR OTHER ENCLOSURE, AN INSULATED THROAT CONNECTOR SHALL BE PROVIDED TO PROTECT THE WIRE FROM ABRASION. CONNECTORS SHALL BE DOUBLE
SET SCREW TYPE, STEEL CONCRETE TIGHT. KEEP RACEWAYS AT LEAST 6 INCHES AWAY FROM PARALLEL RUNS OF FLUES OR STEAM AND HOT WATER PIPES. INSTALL RACEWAY RUNS ABOVE WATER AND STEAM PIPES PROVIDED THAT CABLE
RUN DISTANCES ARE MAINTAINED. USE TEMPORARY CLOSURES TO PREVENT FOREIGN MATTER FROM ENTERING RACEWAY.
11.CONDUIT RUNS ARE SHOWN SCHEMATICALLY. INSTALL CONDUIT WITH A MINIMUM OF BENDS IN THE SHORTEST PRACTICAL DISTANCE CONSIDERING THE BUILDING CONSTRUCTION AND OBSTRUCTIONS, EXCEPT AS OTHERWISE INDICATED. THE
CONTRACTOR SHALL MAKE CERTAIN THAT ANY CONDUIT/RACEWAY RUNS CONTAINING MEDICAL EQUIPMENT SYSTEMS CABLES DO NOT EXCEED THE SPECIFIED MAXIMUM DISTANCES AS SHOWN ON THE ELECTRICAL DETAILS.
12.PROVIDE ENCLOSED METAL RACEWAY SYSTEM (WIRE DUCT) WHERE SHOWN ON DRAWINGS WITH DIVIDERS TO SEPARATE THE DUCT (FOR POWER AND MEDICAL EQUIPMENT CABLING). DIVIDERS AND CROSSOVER PIECES TO BE PROVIDED AS
NECESSARY.
13.FOR UL CERTIFIED SYSTEMS, THE CABLE TO CABLE AS WELL AS THE CIRCUIT TO CIRCUIT SEPARATION REQUIREMENT WAS EVALUATED DURING THE UL SYSTEM INVESTIGATION OF THIS EQUIPMENT. ADDITIONAL SEPARATION OF THE SYSTEM
CABLE ASSEMBLIES INTO SEPARATE OR PARTITIONED RACEWAYS, UNLESS OTHERWISE NOTED, IS NOT NECESSARY TO INSURE SEPARATION OF CIRCUITS, AS THEY CAN BE IN THE SAME RACEWAY.
14.PROVIDE WIRE DUCT/RACEWAY WITH ACCESSIBLE REMOVABLE COVERS. LOCATIONS OF OPENINGS (I.E. ACCESS PANELS) TO BE CUT IN FIELD ARE TO BE COORDINATED WITH THE EQUIPMENT INSTALLATION TEAM. ELECTRICAL PULL BOXES AND
RACEWAY COVERS SHALL BE INSTALLED IN A MANNER TO ALLOW ACCESSIBILITY FOR INSTALLATION AND MAINTENANCE.
15.IN-FLOOR TRENCH DUCT AND FLUSH FLOOR BOXES SHALL BE PROVIDED WITH FULLY CASKETED REMOVABLE COVERS.
16.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED HIGHER THAN 14 FEET ABOVE FINISHED FLOOR, THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL EQUIPMENT
CABLES AT THE CUSTOMER EXPENSE.
17.WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED ABOVE A HARD CEILING (I.E. SHEET ROCK), A 24" x 24" ACCESS PANEL IS REQUIRED AT EACH JUNCTION BOX AND WITHIN 2 FEET OF EACH 90 DEGREE ELBOW OR TEE IN WIRE
DUCT/RACEWAY. THERE MUST BE FREE AND CLEAR ACCESS TO JUNCTION BOXES AND WIRE DUCT/RACEWAY. WHEN ACCESS PANELS ARE LOCATED MORE THAN 3 FEET FROM JUNCTION BOXES AND WIRE DUCT/RACEWAY THE ELECTRICAL
CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE EQUIPMENT INSTALL TEAM PULL SUPPLIED SYSTEM CABLES AT THE CUSTOMER EXPENSE.
18.WIRING SHALL BE INSTALLED IN METAL RACEWAY, 600 VOLT CLASS, STRANDED TYPE THHN-THWN, SINGLE CONDUCTOR ANNEALED COPPER FOR A MAXIMUM OPERATING TEMPERATURE OF 75 DEGREES C (165F). SIZED AS INDICATED. THE
CUSTOMER/CONTRACTOR SHALL LEAVE MINIMUM 10 FT. WIRE TAILS AT ALL OUTLET POINTS WITH WIRE IDENTIFICATION TAGGED AT BOTH ENDS FOR FINAL CONNECTION BY THE EQUIPMENT INSTALL TEAM.
19.IN ADDITION TO THE CIRCUIT BREAKER LOAD CURRENT RATING, CONSIDERATION MUST ALSO BE GIVEN TO SELECTING CIRCUIT BREAKERS THAT HAVE A HIGH ENOUGH SHORT CIRCUIT CURRENT WITHSTAND RATING TO SAFELY COORDINATE
WITH THE POWER SYSTEM AVAILABLE SHORT CIRCUIT CURRENT. GENERALLY, WHEN THE 480 VOLT, 3 PHASE, EQUIPMENT IS SERVED FROM A POWER SUPPLY SYSTEM THAT IS PROVIDED WITH A 500 kVA OR SMALLER TRANSFORMER, A
STANDARD 14,000 RMS AMPERE WITHSTAND RATED CIRCUIT BREAKER WILL BE ADEQUATE. HOWEVER, IF THE POWER SUPPLY SYSTEM TRANSFORMER IS LARGER THAN 500 kVA, THEN THE CIRCUIT BREAKERS HAVING A SHORT CIRCUIT
WITHSTAND RATING GREATER THAN 14,000 RMS AMPERES MAY BE REQUIRED.
20.THE ELECTRICAL CONTRACTOR MUST ADJUST CONDUIT RUN LENGTHS FOR SYSTEM CABLES BASED ON THE MAXIMUM LENGTHS LISTED IN THE SYSTEM CABLE DIAGRAM.
21.THE EPO'S MUST BE INSTALLED BY A QUALIFIED ELECTRICAL CONTRACTOR ACCORDING TO NATIONAL ELECTRICAL CODE, STATE AND LOCAL REGULATIONS. MEASURES SHOULD BE TAKEN TO DESIGN THE CIRCUIT IN SUCH A WAY THAT IT WILL
ALWAYS WORK WHEN THE MEDICAL EQUIPMENT IS POWERED. THE CUSTOMER IS SOLELY RESPONSIBLE FOR THE IMPLEMENTATION OF THE EPO's AND THEIR ASSOCIATED CIRCUITS AND MUST MAKE THE FINAL DETERMINATION CONSIDERING
ALL SITE CONDITIONS AND REGULATORY FACTORS. THE UPS WITH THE IMS SYSTEM, LOCATED IN THE CONTROL ROOM, SUPPLIES IT'S OWN 24V DC CONTROL VOLTAGE AND NEEDS A NORMALLY CLOSED CONTACT FOR EACH EPO. THIS 24V DC
CONTROL VOLTAGE IS TO BE USED FOR THIS UPS ONLY.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-3 CT ELECTRICAL LEGENDELEC
SYM.
ITEM DESCRIPTION
ALL SUPPLIED AND INSTALLED BY CUSTOMER / CONTRACTOR
UPS
JB1
SPK2
RECSPK1
CPU
GANT
EPO
WL
CB
MAIN
MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F.
TO BOTTOM OF BOX. 6' WHIP OF PROVIDE FLEXIBLE CONDUIT AND CONDUCTORS
FOR PHYSICAL CONNECTION TO "PCDU".
4" STANDARD J-BOX FOR "WARNING LIGHT" MOUNTED ABOVE PATIENT ENTRY
DOOR, SEE WIRING DETAIL, FIXTURE AND CONTROLLER BY CUSTOMER.
4" STANDARD J-BOX FOR "EMERGENCY POWER OFF" BUTTON. DPDT, NORMALLY
OPEN MUSHROOM HEAD PUSH BUTTON MOUNTED AT 60" A.F.F. WITH PROTECTIVE
COVER TIED INTO "CB".
4" SHALLOW J-BOX FOR SCAN ROOM SPEAKER, SURFACE MOUNTED ON WALL AT
50" A.F.F. TO BOTTOM OF BOX ATOP "WM".
MINIMUM 10"W X 10"H X 3-1/2"D PULL-BOX, FLUSH MOUNTED IN WALL @ 12" A.F.F.
TO BOTTOM OF BOX, NO COVER, OPENS INTO BACK OF "WD".
GROMMETED OPENING IN COVER OF WALL TROUGH, COORDINATE ACTUAL SIZE
AND LOCATION WITH INSTALLERS, (IF BEING PROVIDED).
GROMMETED OPENINGS IN TROUGHS FOR CABLING, COORDINATE ACTUAL
SIZE AND LOCATION WITH INSTALLERS.
MAIN 480V OR 208V, 3-PHASE, 3-WIRE ELECTRICAL SERVICE FOR CT EQUIPMENT.
WIRE SIZED PER REQUIREMENTS SHT. E-1, PROVIDE FULL SIZE GROUND WIRE.
MINIMUM 16"X16"X6" DEEP JUNCTION BOX SET FLUSH WITH FINISH FLOOR,
PROVIDE GROMMETED OPENING IN COVER, COORDINATE ACTUAL SIZE AND
LOCATION OF OPENING WITH INSTALLERS.
NOTES:
1.ALL CONDUITS, TROUGHS AND RACEWAYS ARE FURNISHED AND INSTALLED BY THE CUSTOMER'S ELECTRICAL
CONTRACTOR.
2.ALL CONDUITS, TROUGHS AND RACEWAYS MUST TAKE THE SHORTEST POSSIBLE ROUTE.
3.CONDUIT RUNS 7 & 11 ARE FOR FIBER OPTIC CABLES, MINIMUM 4.5" RADIUS BENDS.
4.JUNCTION BOX SIZES ARE APPROXIMATE, BOX SIZES MAY BE INCREASED AS NECESSARY TO ACCOMMODATE ALL
INCOMING CONDUITS.
5.CUSTOMER'S ELECTRICIAN MUST BE ON SITE WHEN THE EQUIPMENT IS DELIVERED TO VERIFY PHASE ROTATION,
(CLOCKWISE), TO LAND PRIMARY WIRES, AND TO COORDINATE ALL GROMMETED OPENINGS REQUIRED.OC
DEDICATED 120V ELECTRICAL OUTLETS (QUAD) FOR SYSTEM
EQUIPMENT AND SERVICE EQUIPMENT.
"EMERGENCY POWER OFF" BUTTON MOUNTED AT 60" A.F.F. WITH
PROTECTIVE COVER OR GUARD
WARNING LIGHT - REQUIRED AT ALL ENTRY POINTS
INTO CT SCAN ROOM, COMES ON WHEN SCANNING
ELECTRICAL JUNCTION BOX
WALL TROUGH
FLOOR TROUGH, SURFACE OR RECESSED
MINIMUM 10"W X 2-1/2"D SURFACE MOUNTED WALL DUCT WITH 2 EQUAL
COMPARTMENTS AND REMOVABLE COVERS ENTIRE LENGTH. MOUNT 12"
A.F.F. TO BOTTOM OF DUCT. PROVIDE GROMMETED OPENING INTO "JB1".
WT
WLS "WARNING LIGHT SWITCH" LOCATED IN SCAN ROOM FOR SERVICE PERSONNEL
TO TURN OFF LINE VOLTAGE TO CIRCUIT INSIDE CT GANTRY.
REC MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F.
TO BOTTOM OF BOX. GROMMETED OPENING IN BOTTOM OF BOX FOR SYSTEM
CABLES.
ELEC
SYM.
ITEM DESCRIPTION
ALL SUPPLIED AND INSTALLED BY CUSTOMER / CONTRACTOR
UPS
JB1
SPK2
RECSPK1
CPU
GANT
EPO
WL
PDU
CB
MAIN
MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F.
TO BOTTOM OF BOX. 6' WHIP OF PROVIDE FLEXIBLE CONDUIT AND CONDUCTORS
FOR PHYSICAL CONNECTION TO "PCDU".
4" STANDARD J-BOX FOR "WARNING LIGHT" MOUNTED ABOVE PATIENT ENTRY
DOOR, SEE WIRING DETAIL, FIXTURE AND CONTROLLER BY CUSTOMER.
4" STANDARD J-BOX FOR "EMERGENCY POWER OFF" BUTTON. DPDT, NORMALLY
OPEN MUSHROOM HEAD PUSH BUTTON MOUNTED AT 60" A.F.F. WITH PROTECTIVE
COVER TIED INTO "CB".
4" SHALLOW J-BOX FOR SCAN ROOM SPEAKER, SURFACE MOUNTED ON WALL AT
50" A.F.F. TO BOTTOM OF BOX ATOP "WM".
MINIMUM 10"W X 10"H X 3-1/2"D PULL-BOX, FLUSH MOUNTED IN WALL @ 12" A.F.F.
TO BOTTOM OF BOX, NO COVER, OPENS INTO BACK OF "WD".
GROMMETED OPENING IN COVER OF WALL TROUGH, COORDINATE ACTUAL SIZE
AND LOCATION WITH INSTALLERS, (IF BEING PROVIDED).
GROMMETED OPENINGS IN TROUGHS FOR CABLING, COORDINATE ACTUAL
SIZE AND LOCATION WITH INSTALLERS.
MAIN 480V OR 208V, 3-PHASE, 3-WIRE ELECTRICAL SERVICE FOR CT EQUIPMENT.
WIRE SIZED PER REQUIREMENTS SHT. E-1, PROVIDE FULL SIZE GROUND WIRE.
DEDICATED SHUNT-TRIP CIRCUIT BREAKER FOR CT EQUIPMENT PER ELECTRICAL
REQUIREMENTS. ACTUAL LOCATION TO BE DETERMINED BY CUSTOMER'S
ELECTRICAL ENGINEER. RUN CONDUIT AND WIRES AS REQUIRED INTO "PDU" BOX,
PROVIDE 6' LIQUID TIGHT WHIP AT "PDU" BOX TO CONNECT EQUIPMENT.
NOTES:
1.ALL CONDUITS, TROUGHS AND RACEWAYS ARE FURNISHED AND INSTALLED BY THE CUSTOMER'S ELECTRICAL
CONTRACTOR.
2.ALL CONDUITS, TROUGHS AND RACEWAYS MUST TAKE THE SHORTEST POSSIBLE ROUTE.
3.CONDUIT RUNS 7 & 11 ARE FOR FIBER OPTIC CABLES, MINIMUM 4.5" RADIUS BENDS.
4.JUNCTION BOX SIZES ARE APPROXIMATE, BOX SIZES MAY BE INCREASED AS NECESSARY TO ACCOMMODATE ALL
INCOMING CONDUITS.
5.CUSTOMER'S ELECTRICIAN MUST BE ON SITE WHEN THE EQUIPMENT IS DELIVERED TO VERIFY PHASE ROTATION,
(CLOCKWISE), TO LAND PRIMARY WIRES, AND TO COORDINATE ALL GROMMETED OPENINGS REQUIRED.
RUN CONDUIT CONDUIT CONDUIT CABLE CABLE LENGTH CABLE
NO.POINT TO POINT ROUTING DIAMETER POINT TO POINT USABLE LENGTH FURNISH/INSTALLED BY
1 CB PDU AS REQ'D PER NEC CB PCDU AS REQ'D + 6' WHIP CONTRACTOR
2 CB EPO AS REQ'D PER NEC CB EPO AS REQ'D CONTRACTOR
3 PDU GANT UNDER SLAB (2) 3"PCDU GANT PER OEM INSTALLERS
4 PDU JB1 UNDER SLAB (2) 3"PCDU JB1 PER OEM INSTALLERS
5 PDU EPO AS REQ'D 1
2"PCDU EPO AS REQ'D CONTRACTOR
6 REC1 GANT UNDER SLAB 3"REC GANT PER OEM INSTALLERS
7 REC2 JB1 UNDER SLAB (2) 3"REC JB1 PER OEM INSTALLERS
8 REC JB1 UNDER SLAB 1"REC JB1 PER OEM INSTALLERS
9 GANT JB1 UNDER SLAB (2) 3"GANT JB1 PER OEM INSTALLERS
10 WL GANT UNDER SLAB 1
2"WL GANT (2) #12 WIRES CONTRACTOR
11 SPK2 JB1 OVER CEILING 1
2"SPK2 CPU PER OEM INSTALLERS
OC
RJ45 CONNECTOR, NETWORK DATA CONNECTION TO CUSTOMER
NETWORK.
DEDICATED 120V ELECTRICAL OUTLETS (QUAD) FOR SYSTEM
EQUIPMENT AND SERVICE EQUIPMENT.
"EMERGENCY POWER OFF" BUTTON MOUNTED AT 60" A.F.F. WITH
PROTECTIVE COVER OR GUARD
WARNING LIGHT - REQUIRED AT ALL ENTRY POINTS
INTO CT SCAN ROOM, COMES ON WHEN SCANNING
ELECTRICAL JUNCTION BOX
WALL TROUGH
FLOOR TROUGH, SURFACE OR RECESSED
MINIMUM 10"W X 2-1/2"D SURFACE MOUNTED WALL DUCT WITH 2 EQUAL
COMPARTMENTS AND REMOVABLE COVERS ENTIRE LENGTH. MOUNT 12"
A.F.F. TO BOTTOM OF DUCT. PROVIDE GROMMETED OPENING INTO "JB1".
WT
WLS "WARNING LIGHT SWITCH" LOCATED IN SCAN ROOM FOR SERVICE PERSONNEL
TO TURN OFF LINE VOLTAGE TO CIRCUIT INSIDE CT GANTRY.
REC MINIMUM 12"X12"X10"DEEP JUNCTION BOX RECESSED 4" INTO WALL @ 28" A.F.F.
TO BOTTOM OF BOX. GROMMETED OPENING IN BOTTOM OF BOX FOR SYSTEM
CABLES.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-4 CT ELECTRICAL PLANNorth
EPO
EPO
2
PCDU
1'-8"
CL REMOTE
CONNECTIVITY
CONNECTION
W/110V RECEPT.
SPK2
WL
CB
10
120V POWER
SOURCE, SEE
DETAIL E-5
GANT
CABLE
ACCESS
AT GANTRY
EXISTING
ELECT.
TRANS.
EXIST
ELECT
PANEL
12'
CONDUIT, WIRES BY E.C.
EMPTY CONDUIT BY E.C.
JB1
WD
1
5
3
4
REC1 REC2
6 7
8
9
11
NOTES:
1.ALL CONDUITS ARE SHOWN SCHEMATICALLY, ALL
CONDUITS TO TAKE THE MOST DIRECT ROUTE.
2.J-BOXES MAY BE INCREASED IN SIZE AS
NECESSARY.
3.THE EPO CIRCUIT DESIGN IS BY THE ELECTRICAL
ENGINEER OF RECORDS.
OC INV SPK1 REC
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-5 ELECTRICAL DETAILSPCDU
BSCBEPOUPS
IN USE OR WARNING LIGHT DETAIL
SHUT OFF SWITCH, LOCATE IN CT
SCAN ROOM FOR SERVICE
X-RAY IN USE OR WARNING LIGHT
FURNISHED AND INSTALLED BY
CUSTOMER, LOCATED ABOVE ENTRY
DOOR. INTERLOCK WITH CT SYSTEM
SUPPLIED
TOSHIBA
SUPPLIED
CUSTOMER
NEUTRAL
120V
ONLY REQUIRED IF OPTIONAL UPS IS INSTALLED
GROUND SOURCE
(PER NEC SECT. 250)
CB
MAIN
BREAKER
SHUNT
TRIP
EPO
"Y" *
G
"X" *
1 Hr
ACCEPTABLE VOLTAGE RANGE
SYSTEM DAMAGE MAY OCCUR
(DISRUPTIVE IMPULSES)
SYSTEM DISRUPTION
MAY OCCUR
(DISRUPTIVE IMPULSES)
200% ABOVE
NOMINAL VOLTAGE
100% ABOVE
NOMINAL VOLTAGE
NOMINAL VOLTAGE
+10
-10% BELOW NOMINAL
SYSTEM DISRUPTION MAY OCCUR
(LOW VOLTAGE, SAGS, & OUTAGE)
1 Min1 Sec100 mS10 mS1 mS100 µS10 µS
SYSTEM ELECTRICAL REQUIREMENTS
AQUILION WITH "PCDU"
SUPPLY CONFIGURATION:3 PHASE DELTA OR WYE
CONNECTED, 150KVA SERVICE
SUPPLY VOLTAGE:480V, 60HZ, 100 AMP
OR
208V, 60HZ, 225 AMP
FOR CONDUCTOR SIZE USING
COMBINED LENGTHS FOR "X" AND
"Y" SEE CONDUCTOR SIZE TABLE.
NOTES:
1.UPS MAY, OR MAY NOT, APPLY.
2.ALL CONDUITS, CABLES, AND CONNECTIONS ARE BY
CUSTOMER'S ELECTRICAL CONTRACTOR.
MAIN BREAKER AND PCDU WIRING DIAGRAM POWER QUALITY REQUIREMENTS
CT GANTRY
SYSTEM DAMAGE IS LIKELY OVER OR
UNDER NOMINAL VOLTAGE INCLUDING
HIGH VOLTAGE, SWELLS, SURGES, &
SAGS.
FAILURES DUE TO INCOMING POWER
IS NOT COVERED BY WARRANTY OR
SERVICE AGREEMENT
PCDU
BSCBEPOUPS
SHUT OFF SWITCH, LOCATE IN CT
SCAN ROOM FOR SERVICE
X-RAY IN USE OR WARNING LIGHT
FURNISHED AND INSTALLED BY
CUSTOMER, LOCATED ABOVE ENTRY
DOOR. INTERLOCK WITH CT SYSTEM
SUPPLIED
TOSHIBA
SUPPLIED
CUSTOMER
NEUTRAL
120V
ONLY REQUIRED IF OPTIONAL UPS IS INSTALLED
PUSH BUTTON IN WITH NORMALLY OPEN
CONTACT IN CONTROL & SCAN ROOM
TO CONTROL SHUNT TRIP BREAKER
200V
150 amp
200V
30 amp
200V
50 amp
CONSOLE
GROUND SOURCE
(PER NEC SECT. 250)
SUPPLIED
WITH SYSTEM
SEE NOTE 1
NOTE 1 - TOSHIBA SUPPLIED POWER CABLES
BETWEEN "PDU" AND TOSHIBA EQUIPMENT ARE TO
USED IF THE LENGTHS PROVIDED ARE SUFFICIENT,
IN CASES WHERE EXTENDED CABLES ARE
REQUIRED THE CUSTOMER'S ELECTRICAL
CONTRACTOR MUST PROVIDE CABLES PER
LOCAL/NATIONAL CODES.
MAIN
SERVICE
480V
3-P
GANTRY
PDU
CB
MAIN
BREAKER
SHUNT
TRIP
EPO
"Y" *
G
"X" *
1 Hr
ACCEPTABLE VOLTAGE RANGE
SYSTEM DAMAGE MAY OCCUR
(DISRUPTIVE IMPULSES)
SYSTEM DISRUPTION
MAY OCCUR
(DISRUPTIVE IMPULSES)
200% ABOVE
NOMINAL VOLTAGE
100% ABOVE
NOMINAL VOLTAGE
NOMINAL VOLTAGE
-10% BELOW NOMINAL
SYSTEM DISRUPTION MAY OCCUR
(LOW VOLTAGE, SAGS, & OUTAGE)
1 Min1 Sec100 mS10 mS1 mS100 µS10 µS
SYSTEM ELECTRICAL REQUIREMENTS
AQUILION WITH "PCDU"
SUPPLY CONFIGURATION:3 PHASE DELTA OR WYE
CONNECTED, 150KVA SERVICE
SUPPLY VOLTAGE:480V, 60HZ, 100 AMP
OR
208V, 60HZ, 225 AMP
FOR CONDUCTOR SIZE USING
COMBINED LENGTHS FOR "X" AND
"Y" SEE CONDUCTOR SIZE TABLE.
NOTES:
1.UPS MAY, OR MAY NOT, APPLY.
2.ALL CONDUITS, CABLES, AND CONNECTIONS ARE BY
CUSTOMER'S ELECTRICAL CONTRACTOR.
MAIN BREAKER AND PCDU WIRING DIAGRAM POWER QUALITY REQUIREMENTS
LIGHT ON SIGNAL
FROM GANTRY
CT GANTRY
SYSTEM DAMAGE IS LIKELY OVER OR
UNDER NOMINAL VOLTAGE INCLUDING
HIGH VOLTAGE, SWELLS, SURGES, &
SAGS.
FAILURES DUE TO INCOMING POWER
IS NOT COVERED BY WARRANTY OR
SERVICE AGREEMENT!
% ABOVE NOMINAL
G
GROUND ELECTRODE PER NEC 250-30(A) BUILDING
STEEL, WATER PIPE OR GROUNDING ELECTRODE AS
CLOSE AS PHYSICALLY POSSIBLE TO THE "PDU".
GROUND CONDUCTORS MUST BE THE SAME SIZE AS
PHASE CONDUCTORS.
NOTES:
1."X-RAY ON" OR WARNING LIGHT TO BE
PROVIDED BY CUSTOMER/CONTRACTOR.
2.DO NOT USE FLUORESCENT FIXTURES, LED CAN
BE USED IF CUSTOMER/CONTRACTOR
PROVIDES REPLAY PER LAMP WATTAGE,
MINIMUM RATING 2 AMP.
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-6 ELECTRICAL DETAILSCT VIEW
WINDOW
TD
CT SCAN RM.
FLOOR
TYPICAL TRENCH AND WALL DUCT DETAILS
TYPICAL CONTROL ROOM CONFIGURATION
NOTE: VERIFY WITH EQUIPMENT VENDOR IF A CONSOLE
WAS INCLUDED IN YOUR ORDR, IF NOT THE CUSTOMER
MUST SUPPLY A COUNTERTOP OR TABLE.
TYPICAL SCAN ROOM ELEVATION
"JB" FLUSH
MOUNTED, OPEN
TO "WD"
ALL TROUGHS
TO HAVE
REMOVABLE
COVERS
ALL TROUGHS
TO BE 3
COMPARTMENTS
45 DEGREE CORNERS
SURFACE MOUNTED TROUGHS, 3
EQUAL COMPARTMENTS
TYPICAL PCDU AND J-BOX ARRANGEMENT
SHOWN USING WALL TROUGH
FINISHED WALL
PCDU
CONDUITS OR TROUGH FROM "PCDU" TO
"GANT" AS SITE CONDITIONS ALLOW
CT VIEW
WINDOW
CPU
WT
PDU
VT
PDU
TD
CT SCAN RM.
FLOOR
TYPICAL TRENCH AND WALL DUCT DETAILS
RECUPS
TYPICAL CONTROL ROOM CONFIGURATION
NOTE: VERIFY WITH EQUIPMENT VENDOR IF A CONSOLE
WAS INCLUDED IN YOUR ORDR, IF NOT THE CUSTOMER
MUST SUPPLY A COUNTERTOP OR TABLE.
TYPICAL SCAN ROOM ELEVATION
"JB" FLUSH
MOUNTED, OPEN
TO "WD"
ALL TROUGHS
TO HAVE
REMOVABLE
COVERS
ALL TROUGHS
TO BE 3
COMPARTMENTS
45 DEGREE CORNERS
SURFACE MOUNTED TROUGHS, 3
EQUAL COMPARTMENTS
TYPICAL PCDU AND J-BOX ARRANGEMENT
SHOWN USING WALL TROUGH
FINISHED WALL
PDU
CONDUITS OR TROUGH FROM "PCDU" TO
"GANT" AS SITE CONDITIONS ALLOW
CONDUITS FROM
GANT & PCDU
JB1
CONDUITS TO
JB1
10" X 10" X 4"D J-BOX
FLUSH IN WALL, OPEN
TO WALL DUCT.
WALL DUCT
FLEX CONDUIT TO "PDU" BY
ELECTRICAL CONTRACTOR
JOB NUMBER
SHEET
PROJECT DATE
REVISION HISTORY
1.
2.
3.
4.
5.
6.
2-14-24
2024-01EQUIPMENT PLANNING INFORMATION - AQUILION ONE CTINTEGRITY VETERINARY CENTER518 PLEASANT STNORTHAMPTON, MA 01060ISSUED FOR COORDINATION PURPOSES ONLY - NOT TO BE USED FOR PERMITTING OR CONSTRUCTION PURPOSES
5335 Avion Park Drive
Unit A
Highland Heights, Ohio
(888)505-0319
These drawings and all information
contained herein are the property of
PrizMED Imaging Solutions and are
to be treated as confidential.
These drawings are not to be
reproduced, copied, distributed,
disclosed or published directly or
indirectly without the expressed
written authorization of PrizMED
Imaging Solutions.
WORK ORDER NUMBER
W-4849
2-22-24 WILL IT FIT SKETCH
2-28-24 PRELIMS
3-10-24 FINALS ISSUED
E-7 SYSTEM CABLE DIAGRAMRUN
NO.
CABLE
(END TO END)
CABLE LENGTH
(MAX. USABLE)
39'-4"
59'-0"
55'-9"
75'-0"
18'-0"
52'-5"
39'-4"
52'-5"
26'-2"
52'-5"
75'-0"
39'-4"
52'-5"
18'-4"
18'-4"
18'-4"
"POWER SYSTEM" TO "GANT"
(39'-4") - (POWER & GROUND)
"POWER SYSTEM" TO "CPU" (59'-0") - (POWER)
"GANT" TO "CPU" (55'-9") - (POWER)
"CPU" TO "SPK1" (26'-2") - (SIGNAL)
"SPK2" TO "CPU" (52'-5") - (SIGNAL)
"INV" TO "GANT" (75'-0") - (SIGNAL)
"CPU" TO "INV" (75'-0") - (SIGNAL)
"REC" TO "CPU"
(52'-5" POWER)
"REC" TO "CPU"
(65'-7" FIBER & SIGNAL)
"REC" TO "GANT"
(36'-1" FIBER)
"GANT" TO "CPU" (52'-5") - (SIGNAL)
A
B
C
D
F
G
E
H
I
J
K
POWER
SYSTEM
POWER
SYSTEM
NOTES:
1)THIS CABLE DIAGRAM IS FOR REFERENCE ONLY AND MAY NOT CONTAIN THE EXACT EQUIPMENT AS SHOWN IN THE ATTACHED DRAWING PACKAGE.
2)THE ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR ENSURING THE CONDUIT RUNS DO NOT EXCEED THE CABLE LENGTHS SHOWN.
3)ALL CABLE CONNECTIONS SHOWN ARE THE MAXIMUM LENGTH AND CABLES CANNOT BE EXTENDED.
4)THIRD PARTY ITEM CONNECTIONS TO BE VERIFIED WITH VENDOR OF SUCH ITEMS.
5)POWER AND SIGNAL CABLES MUST BE RUN IN SEPARATE CONDUIT OR SEPARATE COMPARTMENTS IN TROUGHS.
SPK2 GANTREC CPU SPK1 INV
GANT
CPU
CPUGANT
GANT CPU
SPK1CPU
SPK2 CPU
INV
INV
REC
REC
REC GANT
CPU
CPU
CPU
GANT
A
B
C
D
E
F
G
H
I
J
K
PCDU
RUN
NO.
CABLE
(END TO END)
CABLE LENGTH
(MAX. USABLE)
39'-4"
59'-0"
55'-9"
75'-0"
18'-0"
52'-5"
39'-4"
52'-5"
26'-2"
52'-5"
75'-0"
39'-4"
52'-5"
18'-4"
18'-4"
18'-4"
"POWER SYSTEM" TO "GANT"
(39'-4") - (POWER & GROUND)
"POWER SYSTEM" TO "CPU" (59'-0") - (POWER)
"GANT" TO "CPU" (55'-9") - (POWER)
"CPU" TO "SPK1" (26'-2") - (SIGNAL)
"SPK2" TO "CPU" (52'-5") - (SIGNAL)
"INV" TO "GANT" (75'-0") - (SIGNAL)
"CPU" TO "INV" (75'-0") - (SIGNAL)
"REC" TO "CPU"
(52'-5" POWER)
"REC" TO "CPU"
(65'-7" FIBER & SIGNAL)
"REC" TO "GANT"
(36'-1" FIBER)
"GANT" TO "CPU" (52'-5") - (SIGNAL)
A
B
C
D
F
G
E
H
I
J
K
POWER
SYSTEM
POWER
SYSTEM
NOTES:
1)THIS CABLE DIAGRAM IS FOR REFERENCE ONLY AND MAY NOT CONTAIN THE EXACT EQUIPMENT AS SHOWN IN THE ATTACHED DRAWING PACKAGE.
2)THE ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR ENSURING THE CONDUIT RUNS DO NOT EXCEED THE CABLE LENGTHS SHOWN.
3)ALL CABLE CONNECTIONS SHOWN ARE THE MAXIMUM LENGTH AND CABLES CANNOT BE EXTENDED.
4)THIRD PARTY ITEM CONNECTIONS TO BE VERIFIED WITH VENDOR OF SUCH ITEMS.
5)POWER AND SIGNAL CABLES MUST BE RUN IN SEPARATE CONDUIT OR SEPARATE COMPARTMENTS IN TROUGHS.
SPK2 GANTREC CPU SPK1 INV
GANT
CPU
CPUGANT
GANT CPU
SPK1CPU
SPK2 CPU
INV
INV
REC
REC
REC GANT
CPU
CPU
CPU
GANT
A
B
C
D
E
F
G
H
I
J
K
AQUILION 6ONE CT SYSTEM CABLE LENGTHS
* THESE ARE THE ORIGINAL LENGTHS, PRE-OWNED SYTEMS MAY HAVE DIFFERENT CABLE SETS
PCDU